MS Table (All Models)
Installation Guide
MS Table Installation Guide Rev 1 PN IG-0012 21/09/16
Change History
Rev
Comment
Name
Date
How to use this Installation Guide
This document is intended to be used only as a Quick Start Guide, to facilitate safe and
correct installation and initial configuration of your Peak Scientific gas generator.
For more extensive user operation directions, we recommend you download the User
Manual from the website.
Please visit www.peakscientific.com/downloads to
download the full User Manual for your gas generator.
Thank you for selecting Peak Scientific to meet your laboratory gas generation needs,
and should you require any further assistance or support please do not hesitate to
contact Peak Scientific or Peak Partner from which you purchased your generator.
Page 2
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this document detail the
most known potential hazards, but by definition cannot be all inclusive. If the user
employs an operating procedure, item of equipment or a method of working which
is not specifically recommended by Peak Scientific, the user must ensure that the
equipment will not be damaged or become hazardous to persons or property.
Symbols
This document uses the following symbols to highlight specific areas important to the
safe and proper use of the Generator
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the Generator or the Application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the Generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
WARNING
Page 3
Fittings Kit
x1
1/4” - 6mm Push Fit Fitting
02-4600
x1
x1
8mm Hex Key 00-0007
US Mains Cable C13 6-15 04-1029
x6**
M5 Plain Washer 00-1040
x1
x3m*
6mm PE Tubing 00-1282
US Mains Cable C13 5-15 04-1039
x1
UK Mains Cable C13 04-1028
x6**
M5 Spring Washer 00-1044
x12m*
x6**
M5 Screw 00-5476
1/4” Teflon Tubing 00-1266
x1
x3
6mm Push In Plug 02-1268
1/4” Compression Fitting 02-4421
x1
EU Mains Cable C13 04-1038
Page 4
*MS Table 1N & 2N Hi-Flow only
**Mounting Fittings
1.
0°
0°
Unpack the MS Table from the shipping crate and position on a flat surface, in desired
area.
2.
To lock the levelling feet in place, firstly slacken off the 17mm top nut.
Page 5
3.
Then, using a 14mm spanner unscrew the levelling foot until it reaches the floor.
4.
Unscrew the foot a further 1 full turn before tightening the locking nut to lock the foot
in place.
For non-gas versions of the MS Table proceed to step 8, or for instructions on fitting
the optional water bottle mount, complete step 5 before moving on to step 8.
For gas versions, continue following all remaining steps.
Page 6
5.
00-1044
00-5476
00-1040
OPTIONAL: To attach the water bottle mount, locate the two mounting holes,
towards the back of the MS Table on the left side. Using the screws and washers
provided attach the water bottle mount in the order of: Water bottle mount, Plain
Washer (00-1044), Spring Washer (00-1044) then M5 Screw (00-5476). Do this for
both mounting holes.
6.
02-1268
Insert the 6mm Blanking Plug (02-1268) from the fittings kit into the Drain port at
the rear of the MS Table.
Page 7
7.
Connect an air supply to the compressed air inlet bulkhead, located at the rear
of the unit.
8.
04-1028/04-1029/
04-1038/04-1039
Select the appropriate mains cable from the fittings kit and plug the IEC 60320/C19
cable into the mains input at the rear of the generator.
Page 8
9.
UK
EU
US
110V
CAUTION
US
230V
110VAC 50/60Hz OR 230VAC 50/60Hz
Plug the mains cable into the corresponding 110VAC 50/60Hz OR 230VAC 50/60Hz
single phase power supply.
CAUTION ensure the correct cable has been selected for your electrical supply
10.
For non-gas versions of the MS Table proceed to step 15.
30
minutes
Leave the generator switched on for a period of 30 minutes to purge the system.
Page 9
11.
After 30 minutes of running, the generator can be disconnected from the air supply.
12.
The 6mm blanking plug can now be removed from the Drain port.
Page 10
13.
00-1282
Connect tubing to the Curtain, Source & Exhaust outputs on the generator and
attach to the corresponding inputs on your instrument.
14.
00-1282
Re-connect the compressed air supply to the Air inlet.
Connect 6mm PE Tubing 00-1282 to the Drain outlet.
Fit the other end of the tubing to a suitable drain connection or container. The
container MUST NOT have an airtight seal as water and air will be expelled
Page 11
periodically under a slight pressure.
15.
/
CONGRATULATIONS
Your Peak Scientific MS Table is now fully installed, operational and ready to supply
gas on demand to your instrument.
Page 12
General Notes
Electrical requirements for 230V generators is 110VAC 50/60Hz OR 230VAC
50/60Hz.
1.
The Teflon tube supplied in the fittings kit is to allow the generator to be installed
adjacent to the instrument it is supplying. If the generator is to be installed further
away from the instrument then attention needs to be paid to pressure drops that could
occur. The maximum distance away from the instrument the generator can be placed
still using 6mm (4mm internal diameter) or ¼” (3/16” internal diameter) tubing is 10m.
For further information relating to tubing lengths please refer to the user manual.
2.
Electrical Requirements
Generator
MS Table
Voltage
230
110
Frequency
50
60
Plug Type
UK, EU, US 230V,
US 110V
Please familiarise yourself with the full User Manual available at www.peakscientific.
com/downloads. This includes all other technical specifications, operating
instructions, service requirements, contact details and trouble shooting.
Page 13
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/register.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 14
Please visit www.peakscientific.com/downloads to
download the full User Manual for your gas generator.
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com.
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
MS Table Installation Guide Rev 1 PN IG-0012 21/09/16