Extracted text
Infinity XE 60
Nitrogen generator supporting multiple instruments
Part Number : See Reverse
Service Kit : See Reverse
Your local gas generation partner
Description
Producing between 10 and 500 L/min of high purity nitrogen gas, the Infinity
XE 60 series nitrogen membrane gas generators can comfortably supply
multiple laboratory instruments with instrument grade nitrogen.
Offering labs a cost-effective alternative to cylinders in a compact, space saving
nitrogen solution, the Infinity XE 60’s Intelligent Membrane Management (IMM)
uniquely optimizes air consumption depending on set purity, inlet pressure/
flow and nitrogen demand. With variable N2 purities (95- 99.5%), XE 60 can
meet the demands of numerous applications such as ELSD, NMR, FT/MS,
Sample Evaporators, as well as LC-MS and should your lab demands grow
over time, Infinity XE 60 can have its nitrogen output increased on-site up to a
maximum of 500 L/min all within the same original footprint.
To operate, the Infinity series of generators require an external source of
compressed air, which meets a minimum quality grade of ISO 8573-1:2010 Class
1.4.1. Where required, Peak can provide assistance on external compressors &
pre/post filtration as part of delivering a complete solution for your needs.
Applications
)
LC-MS
)
NMR
)
ELSD
)
FT/MS
)
Sample Evaporators
Key Features
405.00mm
940.00mm
995.00mm
INFINITY XE 60Dimensions
)
Flow rates ranging from 10-500 L/min*
)
N2 purity up to 99.5%
)
Versatile solution with varying Nitrogen flow, pressure &
purity capabilities
)
In-field expansion of nitrogen capacity, for models without
dry air outlet, to meet growing lab demands
)
Optional second outlet for Dry Air
)
Intelligent Membrane Management (IMM) optimizes
compressed air requirement based on gas consumption
)
IMM reduces overall compressor run time and energy
consumption, reducing carbon footprint
)
Fully comprehensive 2-Year Warranty**
* Dependent on inlet pressure, inlet/outlet flow & Nitrogen purity setting
** subject to meeting 2 year warranty conditions
www.peakscientific.com/warranty-statement/
Copyright 2023 Peak Scientific - Rev - Infinity XE 60 Data sheet Rev 6 RSID 4130 EN - 05/07/23
Technical Specifications
Flow Rate Min - Max†
(Nitrogen L/min)
6010
6020
6030
6040
up to 130 L/min
up to 260 L/min
up to 390 L/min
up to 500 L/min
Purity N2 Min - Max
up to 99.5%
N2 Dewpoint
up to -50°C
Flow Rate Min - Max†
up to 210 L/min
(Dry Air L/min)
up to 210 L/min
Air Dewpoint
up to 210 L/min
up to 210 L/min
-20°C
Max Outlet Pressure
135Psi (Depending on Inlet Pressure)
Inlet Air Quality Requirement
ISO 8573-1:2010 (1:4:1)
Inlet Air Pressure Requirement
60 - 145Psi (Depending on Outlet Pressure requirements)
Inlet Air Pressure Flow Rate††
262 L/min
Voltage
100 - 230 VAC +/- 10%
Frequency
50 / 60 Hz
Current
0.5A @ 110V / 0.25A @ 230V
Input Connection
C20 Plug
Power Cord
C19 Socket to local connection
Power Consumption
0.1 kW
Operating Temperature
5°C - 30°C / 41°F - 86°F
Pollution Degree
2
Installation Category
II
Dimensions mm (inches) W x D x H
405 (15.9) x 940 (37) x 995 (39.2)
Weight Kg (Lbs)
75Kg (165.4lbs)
78.5Kg (173.1bs)
80Kg (176.4bs)
82.5Kg (181.9bs)
Shipping Weight Kg (Lbs)
92.5Kg (204bs)
95Kg (209.5bs)
97.5Kg (215bs)
100Kg (220.5.1bs)
†Dependant on inlet pressure, inlet/outlet flow and Inlet Air Quality
†† For Infinity XE users with the Dry Air option, please allow for an additional compressed air requirement of 262L/min to support the Dry Air output.
Ordering Information
Part Number (Nitrogen Only)
Part Number (Nitrogen & Dry Air)
3303147
3303148
3303149
3303150
3303147-AD
3303148-AD
3303149-AD
3303150-AD
Annual Service Kit
Fixed Price Preventative Maintenance
Visit: www.peakscientific.com/ordering
Complete/Standard Maintenance Plan
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 2 year warranty**. Beyond this period however you can ensure that you’re investment continues to
be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the
reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Peak Scientific UK
Tel: +44 (0)141 812 8100
Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649
Fax: +1 978 608 9503
Product Certifications
Peak Scientific’s Quality
Management System
conforms to: ISO:9001:2008
For a full list of our worldwide
office locations, please visit:
Web: www.peakscientific.com
Email: marketing@peakscientific.com
Copyright 2023 Peak Scientific - Rev - Infinity XE 60 Data sheet Rev 6 RSID 4130 EN - 05/07/23