Corona Compressed Air Site Preparation Guide
2014 © Peak Scientific - Rev. 0 – 07/01/14
Corona CA Site Prep Guide
Contents
Customer Responsibilities Change History Related Manuals Safety Notices Safety Notice to Users Technical Specifications Environment Generator Outlets Electrical Requirements General Site preparation Environmental control Space provision Electrical requirements Unpacking Space Required Unpacking Instructions Copper Tubing Stacking Units 3 4 4 5 5 6 6 6 6 6 7 7 8 9 12 12 12 13 14
Page 2
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Customer Responsibilities
To ensure a successful and timely installation of your Peak generator, please refer to this set of requirements. Correct site preparation is the first key step in ensuring that your generator and systems operate reliably over an extended lifetime. This document is an information guide and checklist that outlines the requirements for your site. It is essential your site meets the following specification prior to the installation date. For details, see specific sections within this document
The necessary spatial requirements are met. The correct environment is provided for the generator. Electrical outlet locations and quantities are planned. Adequate exhaust ventilation is provided. The correct tubing lengths and diameters are used in relation to the distance from the instrument.
Failure to meet the Site requirements of your Peak generator as stated above and detailed in this document could result in the unit underperforming and possible loss of warranty. Please sign below to confirm your compliance with the aforementioned requirements. Once complete please return to Peak Scientific, or whomever the unit was purchased from. Return addresses can be found on the back page of this document.
Name:
Company:
Signed:
Date:
Model:
Cust. Sales Order No.:
Page 3
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Change History
Rev. 1 2 3 4 Comment Name Date
Related Manuals
Document number
UM-CoronaCompressedAir IG-CoronaCompressedAir
Manual name
Corona Compressed Air User Manual Corona Compressed Air Installation Guide
Description
Describes the operation of the generator and all service requirements. Supplied with the generator. Details the installation process of the generator. Supplied with the generator.
Page 4
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Safety Notices
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the Generator:
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met. A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the Generator or the application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met.
Safety Notice to Users
This site preparation guide must be read thoroughly and understood before installation and operation of your Peak Corona Compressed Air Generator. Incorrect preparation or use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or preparing for installation, personnel must employ safe practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations.
Page 5
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Technical Specifications
Environment
Min/Max Operating Ambient Temperature Maximum Altitude Maximum Relative Humidity Min/Max Storage Temperature 5°C - 35°C / 41°F - 95°F 2000 metres 80% -20°C - 60°C / -4°F - 140°F
Generator Outlets
Max Flow Rate Max Pressure Particles Phthalates Suspended Liquids Gas Outlets Drain Outlets Pressure Gauges 32 l/min 110 psi / 7.58 bar <0.01µm None None 1 1 1
Electrical Requirements
Voltage Frequency Current Input Connection Type Power Cord Type Pollution Degree Insulation Category 110VAC 50/60 Hz 5.30 A C14 Plug C13 Socket to Local Connection 2 II 230VAC 50/60 Hz 2.45 A
General
Generator Dimensions cm (inches) Generator Weight Kg (lbs) Shipping crate dimensions cm(inches) Shipping Weight Kg (lbs) Noise Level 41.6 x 54.2 x 40.6 (16.3 x 21.3 x 15.9) 41.5 (91.5) 58.2 x 79.0 x 43.5 (22.9 x 31.1 x 17.1) 64 (141) 50 dB
Page 6
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Site preparation
Environmental control
If the generator is stored in an enclosed space the environment must be controlled via an air conditioner or extraction fan.
>5 C
o
o <35 C <80% RH
Max Temp; 35 C Max Relative Humidity; 80% Non-Condensing
o
Generator MUST NOT be stored or installed outside.
Page 7
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Space provision
The minimum space should be provided as follows; 75mm 75mm
406mm
566mm
75mm
617mm
Failing to provide adequate cooling space around the generator may cause damage to the Generator. This will reduce service life and invalidate warranty.
Page 8
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Electrical requirements
The below power cables are supplied with the generator.
For cables pertaining to countries not displayed above, it is the responsibility of the end user to provide an appropriate power cable which meets the requirements defined in the Technical Specification section on page 6.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
If the mains power at your location is less than 110V/220V then a suitable transformer should be sourced, to boost the mains power to an acceptable level.
Page 9
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
The power cable supplied with the generator is 2m long. A mains socket providing the power should be located adjacent to the generator and within a 1.5m radius to the generator.
Page 10
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
On-Site Transit
When moving the generator in its shipping crate, doorways and other openings such as elevators must fit with the sizes in the figure below.
>850mm
>650mm 435mm
41.5kg
Page 11
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Unpacking
Space Required
The image below shows the minimum space required to unpack the generator from its shipping crate. 2m
1m
Top of Shipping Crate
1m
2m
Tools
Required : Recommended :
Pozidrive #2 Screwdriver
Drill with Pozidrive #2 Bit
Unpacking Instructions
Remove all screws encircled in red, use Drill or Pozidrive #2 screwdriver. Once all 8 screws have been removed, lift lid upwards. Next with someone on either side of the unit, position your hands underneath the unit ready to lift. Ensuring your knees are bent and your back is straight, lift the unit to the desired location.
Page 12
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Copper Tubing
Soldering may lead to contamination. Compression fittings recommended.
Page 13
2014 © Peak Scientific - Rev. 0 - 07/01/14
Corona CA Site Prep Guide
Stacking Units
The units in the Peak GC Range can be stacked on top of one another, allowing for better use of space in laboratories. When stacked on top of a work bench, the maximum height a combination of generators can reach is 1 metre. When stacked from the floor, the maximum height a combination of generators can reach is 1.7 metres. Example Combinations have been illustrated in the image below.
>1 metre
>1.7 metres
Page 14
2014 © Peak Scientific - Rev. 0 - 07/01/14
Peak Scientific UK Fountain Crescent Inchinnan Business Park Inchinnan PA4 9RE Scotland, UK Tel: +44 (0)141 812 8100 Fax: +44 (0)141 812 8200 Peak Scientific Germany Herriotstrasse 1 60528 Frankfurt
Peak Scientific North America 19 Sterling Road Suite #1 Billerica, MA 01862 USA Tel: +1 866 647 1649 Fax: +1 978 608 9503 Peak Scientific China Room 606 Building 1 Lane 2277 Zuchongzhi Road Pudong New Area Shanghai 201203, China Tel: +86 21 5079 1190 Fax: +86 21 5079 1191 Peak Scientific Brazil Peak Scientific Brasil Av. Quieroz Filho 1700 – torre Sunny-Conj. 602 São Paulo SP Brasil Tel: +1 866 647 1649
Peak Scientific India 202, Amsri Shamira Old Lancer Line Opp. St. Mary’s Degree College S.D. Road Secunderabad 500 025, India Tel: +91 40 2780 0663 Fax: +91 40 2780 0663 Peak Scientific Mexico Solon 352 Col. Los Morales Polanco 11530 Mexico, D.F. Mexico
Peak Scientific Africa PO Box 478 Somerset Mall 7137 South Africa Tel: +27 (0)2185 16542 Fax: +27 (0)8654 64473 Peak Scientific Australia PO Box 65 Belmont Victoria 3216 Australia
Germany
Tel: +49 (0)69 677 33 490 Fax: +49 (0)69 677 33 200 Peak Scientific Taiwan 4F.-6, No.736 Zhongzheng Rd. Zhonghe Dist. New Taipei City
Tel: Fax:
+1 866 647 1649 +1 978 608 9503
Tel:
+61 1300 965 352
Peak Scientific Japan K.K. 2-7-56, 2F Fuji Building 28 Kita Aoyama, Minato-Ku Tokyo, Japan 107-0061
Peak Scientific Singapore 3 Science Park Drive #03-14 The Franklin Science Park Drive 1 Singapore, 118223
Tel: +886 2 8226 2383 Fax: +886 2 8226 9499
Tel: +81 3-6864-0468
Tel: +65-6777 1966 Fax: +65-6777 1978
2014 © Peak Scientific - Rev. 0 – 07/01/14