DOC696 - Infinity XE 5020 IQ/OQ
Original PDF File459 KB |
Download |
View directly in browser459 KB |
View in browser |
Resource ID
3486
Access
Open
Contributed by
Derek Rennie
Updated date
09 December 21
Review date
01 May 24
Title
DOC696 - Infinity XE 5020 IQ/OQ
Revision Number
1
Revision Details
Initial release
Keywords
DOC696, Infinity, XE, 5020, installation, qualification, operation, operational, IQ/OQ, IQ, OQ, service, certificate, certification, compliance, protected
IQOQ, IQ, OQ, IQ/OQ, Installation Qualification Operational Qualification
Product Range
Infinity
Language
UK English
Company Brand
Peak Scientific
Extracted text
INSTALLATION QUALIFICATION
Instrument Origin and Identification
Instrument Series:
Infinity XE 50 Series
Model:
5020
Instrument Description:
Nitrogen Gas Generator
Manufacturer:
Peak Scientific Instruments Ltd
Fountain Crescent
Inchinnan Business Park
Inchinnan
Renfrew PA4 9RE
Scotland, UK
Service Contact:
Peak Scientific Instruments Ltd
Tel.
+44 (0)141-812-8100
Email support@peakscientific.com
Serial Number:
Condition when Received:
New
Used
Reconditioned
Date Received:
Installed Location:
Purchase Order Number:
Identification / Asset Number:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 1 of 8
INSTALLATION QUALIFICATION
Instrument Specification
Instrument Parameter
Specified Range
Acceptable
Output
Nitrogen Pressure
Refer to generator specific data
Nitrogen Flow Rate
Refer to generator specific data
System Purge Time
1 minute
Environment and Siting
Has the instrument been adequately acclimatized since transport or storage?
(Requires 3 hours before operation)
Yes
No
Operating Parameter
Specified Range
Ambient Temperature
5-30 ºC (41-86 ºF)
Maximum Relative Humidity
80% Non-Condensing
Maximum Altitude
2000 m (6500 ft)
Supply Voltage & Frequency
100/230V ±10%, 50/60Hz
Refer to User Manual for acceptable range
Siting – In-House / Large Space
Minimum Clearance around Machine – 75mm
Siting – In-House / Confined Space
Mains Power Outlet
Acceptable
Air Conditioner/Extraction Fan
Within 2m Radius to Generator
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 2 of 8
INSTALLATION QUALIFICATION
Instrument Delivery and Documentation
Unpack the instrument carefully. Are any items missing against the Packing List?
Yes
No
If yes, state missing items:
Is there any damage to the instrument?
Yes
No
Yes
No
If yes, describe damage:
Manufacturer informed:
Is all standard documentation included?
Present
Missing
Declaration of Conformity
Warranty Registration Card
Installation
Installed By:
User
Authorized Engineer
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 3 of 8
INSTALLATION QUALIFICATION
Summary Report
Instrument: INFINITY XE 5020
Serial Number:
Assessment of Complete Installation Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicate that this instrument has been
satisfactorily delivered and installed. The instrument has passed the Installation Qualification
procedure and may now be submitted for Operational Qualification.
IQ Completed By:
Date:
Deviations Approved By:
Date:
IQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 4 of 8
OPERATIONAL QUALIFICATION
Pre-Run Checks
Pre-Run Check
OK
Discrepancy
N/A
Confirm inlet air pressure meets requirement.
Refer to generator specific data
Confirm inlet air flow meets requirement. Refer
to generator specific data
Power-off / Isolate Generator from Mains Power
/ Allow pressure to drop to Zero
Outlet pressure gauge showing 0 psig
Drain line connected to suitable outlet
Nitrogen Outlet disconnected
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 5 of 8
Functional Tests
Note: A flow controller should be temporarily connected to the Nitrogen outlet to run these
tests.
Functional Test
Acceptance Criteria
Pass
Apply Mains Power / Switch On
On-Off Switch illuminates
Open Compressed Air inlet
valve
Outlet Pressure Gauge shows
increasing pressure
Fail
Refer to generator specific data
to set outlet pressure
Allow to run for 10 minutes.
(Ensure Side panel is fitted)
Shut off outlet flow
Pressure builds to set outlet
pressure
Confirm “Eco-Mode” LED
illuminates and the pressure
settles at pressure switch
setting (default 90Psi)
After functional test 1. Power-off / Isolate Generator from Mains Power / Allow pressure to drop to Zero
2. Remove test flow controller
3. Connect Gas line to instrument
4. Apply Mains Power / Switch On
5. Allow to run for 10 minutes
Output pressure
Stable pressure shown on
output pressure gauge
Output Connection Leak Tight
No Leaks
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 6 of 8
OPERATIONAL QUALIFICATION
Calibration
This instrument is supplied fully calibrated by the manufacturer using equipment traceable to national
standards. Calibration directly following installation is not usually required.
Calibration Step
Pass
Fail
Calibrate output flow(s)
Serial No. of Calibrated Flow Meter -:
Calibration Due Date of Flow Meter -:
Calibrate output pressure(s)
Serial No. of Calibrated Pressure Gauge -:
Calibration Due Date of Pressure Gauge -:
Comments:
Training
Date Instrument Released for Use:
Trained Operators
Read User
Manual
Practical
Training
Authorized Signature
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 7 of 8
OPERATIONAL QUALIFICATION
Summary Report
Instrument: INFINITY XE 5020
Serial Number:
Assessment of Complete Operational Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicate that this instrument is operating
satisfactorily following delivery and installation. The instrument has passed the Operational
Qualification procedure and may now be released for use.
OQ Completed By:
Date:
Deviations Approved By:
Date:
OQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: DOC696
Revision: 1
Page 8 of 8