Account Menu
  • Log in
  • Contact us
  •  Dash
  • Search results
  •  Featured collections
  •  Recent
  •  Knowledge Base
 All resources
Product Range
By date

 Geographic search

 Advanced search


Log in

Username
Password


 Click here to apply for an account

 Click here if you have forgotten your password

 Use SSO to log in

View all results

DOC696 - Infinity XE 5020 IQ/OQ 

Full screen preview

Resource tools

Original PDF File

459 KB Download
  •  Share
Resource details

Resource ID

3486

Access

Open

Contributed by

Iain Kean

Updated date

26 February 20

Review date

26 February 22

Title

DOC696 - Infinity XE 5020 IQ/OQ

Revision Number

1

Revision Details

Initial release

Keywords

DOC696, Infinity, XE, 5020, installation, qualification, operation, operational, IQ/OQ, IQ, OQ, service, certificate, certification, compliance, protected
IQOQ, IQ, OQ, IQ/OQ, Installation Qualification Operational Qualification

Product Range

Infinity

Language

UK English

Company Brand

Peak Scientific

Public / Private

public

Extracted text

INSTALLATION QUALIFICATION
Instrument Origin and Identification
Instrument Series:

Infinity XE 50 Series

Model:

5020

Instrument Description:

Nitrogen Gas Generator

Manufacturer:

Peak Scientific Instruments Ltd
Fountain Crescent
Inchinnan Business Park
Inchinnan
Renfrew PA4 9RE
Scotland, UK

Service Contact:

Peak Scientific Instruments Ltd
Tel.
+44 (0)141-812-8100
Email support@peakscientific.com

Serial Number:
Condition when Received:

New
Used
Reconditioned

Date Received:
Installed Location:
Purchase Order Number:
Identification / Asset Number:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 1 of 8

INSTALLATION QUALIFICATION
Instrument Specification
Instrument Parameter

Specified Range

Acceptable

Output
Nitrogen Pressure

Refer to generator specific data

Nitrogen Flow Rate

Refer to generator specific data

System Purge Time

1 minute

Environment and Siting
Has the instrument been adequately acclimatized since transport or storage?
(Requires 3 hours before operation)
Yes

No

Operating Parameter

Specified Range

Ambient Temperature

5-30 ºC (41-86 ºF)

Maximum Relative Humidity

80% Non-Condensing

Maximum Altitude

2000 m (6500 ft)

Supply Voltage & Frequency

100/230V ±10%, 50/60Hz
Refer to User Manual for acceptable range

Siting – In-House / Large Space

Minimum Clearance around Machine – 75mm

Siting – In-House / Confined Space
Mains Power Outlet

Acceptable

Air Conditioner/Extraction Fan
Within 2m Radius to Generator

Comments:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 2 of 8

INSTALLATION QUALIFICATION
Instrument Delivery and Documentation
Unpack the instrument carefully. Are any items missing against the Packing List?
Yes

No

If yes, state missing items:
Is there any damage to the instrument?
Yes

No

Yes

No

If yes, describe damage:

Manufacturer informed:

Is all standard documentation included?
Present

Missing

Declaration of Conformity
Warranty Registration Card

Installation
Installed By:

User

Authorized Engineer

Comments:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 3 of 8

INSTALLATION QUALIFICATION
Summary Report
Instrument: INFINITY XE 5020

Serial Number:

Assessment of Complete Installation Qualification:
No Deviations
Deviation

Deviations

Impact on Operation

Justification for Acceptance

Successful completion of the preceding activities and checks indicate that this instrument has been
satisfactorily delivered and installed. The instrument has passed the Installation Qualification
procedure and may now be submitted for Operational Qualification.
IQ Completed By:

Date:

Deviations Approved By:

Date:

IQ Approved By:

Date:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 4 of 8

OPERATIONAL QUALIFICATION
Pre-Run Checks
Pre-Run Check

OK

Discrepancy

N/A

Confirm inlet air pressure meets requirement.
Refer to generator specific data
Confirm inlet air flow meets requirement. Refer
to generator specific data
Power-off / Isolate Generator from Mains Power
/ Allow pressure to drop to Zero
Outlet pressure gauge showing 0 psig
Drain line connected to suitable outlet
Nitrogen Outlet disconnected
Comments:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 5 of 8

Functional Tests
Note: A flow controller should be temporarily connected to the Nitrogen outlet to run these
tests.
Functional Test

Acceptance Criteria

Pass

Apply Mains Power / Switch On

On-Off Switch illuminates

Open Compressed Air inlet
valve

Outlet Pressure Gauge shows
increasing pressure

Fail

Refer to generator specific data
to set outlet pressure
Allow to run for 10 minutes.
(Ensure Side panel is fitted)
Shut off outlet flow

Pressure builds to set outlet
pressure
Confirm “Eco-Mode” LED
illuminates and the pressure
settles at pressure switch
setting (default 90Psi)

After functional test 1. Power-off / Isolate Generator from Mains Power / Allow pressure to drop to Zero
2. Remove test flow controller
3. Connect Gas line to instrument
4. Apply Mains Power / Switch On
5. Allow to run for 10 minutes
Output pressure

Stable pressure shown on
output pressure gauge

Output Connection Leak Tight

No Leaks

Comments:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 6 of 8

OPERATIONAL QUALIFICATION
Calibration
This instrument is supplied fully calibrated by the manufacturer using equipment traceable to national
standards. Calibration directly following installation is not usually required.
Calibration Step

Pass

Fail

Calibrate output flow(s)
Serial No. of Calibrated Flow Meter -:
Calibration Due Date of Flow Meter -:
Calibrate output pressure(s)
Serial No. of Calibrated Pressure Gauge -:
Calibration Due Date of Pressure Gauge -:
Comments:

Training
Date Instrument Released for Use:
Trained Operators

Read User
Manual

Practical
Training

Authorized Signature

Comments:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 7 of 8

OPERATIONAL QUALIFICATION
Summary Report
Instrument: INFINITY XE 5020

Serial Number:

Assessment of Complete Operational Qualification:
No Deviations
Deviation

Deviations

Impact on Operation

Justification for Acceptance

Successful completion of the preceding activities and checks indicate that this instrument is operating
satisfactorily following delivery and installation. The instrument has passed the Operational
Qualification procedure and may now be released for use.
OQ Completed By:

Date:

Deviations Approved By:

Date:

OQ Approved By:

Date:

Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK

T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com

Doc. Ref.: DOC696
Revision: 1
Page 8 of 8

Search for similar resources