Collision Gas Add-On Data Sheet
Original PDF File |
233 KB | Download |
View directly in browser |
233 KB | View in browser |
Resource ID
4679
Access
Open
Contributed by
Cleo Denholm
Updated date
10 May 23
Review date
10 May 25
Title
Collision Gas Add-On Data Sheet
Revision Number
5
Revision Details
fax number change - CD
Keywords
Genius Xe, Xe, Genius, Data, Collision, Addon, Add, On, Collision Gas, Add on,
Product Range
*None
Language
UK English
Company Brand
Peak Scientific
Public / Private
public
Extracted text
Collision Gas AddOn
Part number: 3304886
Description
Designed to work exclusively with Genius XE 35 and Genius XE 70,
our Collision Gas AddOn supplies Ultra High Purity Nitrogen to the
collision cell at a flow of up to 200cc/min.
With a set purity of 99.999%, the Collision Gas AddOn produces a
high quality flow with no reduction to the output of the Genius XE
generator. The Collision Gas AddOn allows labs to draw nitrogen
from their Genius XE 35 or Genius XE 70 which removes the need for
gas cylinders for UHP nitrogen.
Application Types
Generator Details
• Thermo Orbitrap
•
•
Gas types: Nitrogen
Integrated compressor: No
• Agilent QTOF
• Agilent 6560
• Waters Xevo TQ Absolute
Key Features
•
•
•
•
•
•
Can be used to supply Ultra High Purity Nitrogen for the collision
cell 99.999% nitrogen at 200cc/min*
Designed to be used alongside a Genius XE 35 or XE 70, creating
a dual purity system
HMI Screen – monitor the performance of the Collision Gas Add
On
Maintenance interval every year, down time minimized as service
can be performed at same time as Genius XE
No risk of running out of gas in the middle of analysis
2 year comprehensive manufacturer’s warranty
*Nitrogen generator providing main gas source to an instrument.
(Genius XE 35 or XE 70 systems)
Technical Specifications
Collision Gas AddOn
Technical Specifications
Collision Gas AddOn
Gas Type
Nitrogen
Max Gas Flow:
200cc/min
Max Output Pressure:
80psi/5.5bar
Max Output Pressure:
80psi/5.5bar
Max Purity:
99.999%
Gas Outlets Fitting:
1 x 1/4” BSPP
Min Air Inlet Flow:
4 L/min
Power Consumption:
20W @ 100VAC / 48W @ 240VAC
Voltage:
100v 240v
Frequency:
50/60 Hz
Current:
0.2A
Max Operating Temp:
35°C / 95°F
Particles:
<0.01µm
Noise Level:
Silent in operation
Size (HxWxD) mm
165 x 360 x 425 mm
Size (HxWxD) Inches
6.5 x 14.2 x 16.7 inches
Generator Weight
14kg / 30.8lbs
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the world,
and come backed with a 12 month warranty. Beyond this period however you can ensure that your investment continues to be
[Protected] by our comprehensive generator care cover.
Our worldclass aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the
reassurance of instant access to worldwide technical support and priority onsite response in the untimely event of a breakdown.
For ordering parts visit: www.peakscientific.com/ordering/
For service plans visit: www.peakscientific.com/service/serviceplans/
Peak Scientific UK
Tel: +44 (0)141 812 8100
Fax: +44 (0)141 812 8200
Peak Scientific’s Quality Management System conforms to: ISO:9001
Peak Scientific North America
Tel: +18007676532
Fax: +19786089503
Peak Scientific China
Tel: +86 21 5079 1190
Fax: +86 21 5079 1191
For a full list of our worldwide
office locations, please visit:
Web: www.peakscientific.com
Email: discover@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2023 Collision Gas AddOn Datasheet engb 18/05/2023