NITROPAC - Manual  

Full screen preview
Resource tools

Original PDF File

2.5 MB

Download

View directly in browser

2.5 MB

View in browser
  •  Share
  •  Edit with AI
Resource details

Resource ID

4093

Access

Open

Contributed by

Derek Rennie

Updated date

02 April 21

Review date

01 May 24

Title

NITROPAC - Manual

Revision Number

1

Revision Details

First Upload - DL

Keywords

Precision, Hydrogen, SL, Data, Sheet, H2, GC, FID, GC-FID

Product Range

Precision

Language

US English

Company Brand

Peak Scientific

Extracted text

Operation Manual

GB

Nitrogen Generators

Nitropac N01x – N06x

IOM_DO_NITROPAC_GB_2017_07_03.docx

Operation Manual Nitropac N01x – N06x

Page 2

Operation Manual Nitropac N01x – N06x

Contents
How to use this manual

4

Introduction

5

Technical Description

5

Basic concept
General construction
Nitropac model number breakdown

5
6
6

Warranties and Liabilities

7

Safety Notices

8

Symbols
Safety Notice to Users

8
8

Product Overview

9

Pressure Switch Overview

10
10

Electrical Panel Layout

11

Technical Specification

12

Environment
Compressed Air Supply
Electrical Requirements
General

12
12
12
13

Installation

14

Generator Environment
Unit Controls
Inlet / Outlet Connections
Compressed Air Quality
Electrical Connection
Recommended Piping Layout
Commissioning / Safe Start-Up Procedure

15
16
16
17
18
18
19

Normal Operation

21

Cleaning

21

Service Requirements

22

Safe Isolation Process
Service Schedule

22
22

Service Counter Reset

23

PLC Settings

24

Troubleshooting

25

Pneumatic Schematic - Rev 2A

26

Electrical Schematic – Rev 3

27

Page 3

Operation Manual Nitropac N01x – N06x

How to use this manual
This manual is intended for end users and has been written so that it can either be read as a step
by step guide to installation and usage; or as a reference document where you can skip to the
relevant information. However, please note that this manual includes important safety information
for how to install and operate the system properly.
Please review each of the following sections carefully.
Thank you for selecting Donaldson to meet your Gas Generation needs, should you require any
further assistance or support please do not hesitate to contact Donaldson.

Page 4

Operation Manual Nitropac N01x – N06x

Introduction
The Donaldson Nitropac Generator is designed to cater for a wide variety of industrial and
scientific applications. Your generator will have been carefully selected to meet your specific
pressure, flow, and purity requirements, if you have any questions regarding the sizing of your
system please do not hesitate to contact Donaldson.

Technical Description
Basic concept
A modular range of Carbon Molecular Sieve (CMS) nitrogen gas generators. The generators
operate using Pressure Swing Adsorption (PSA). This is where the product contains two separate
containers or “beds” of granular carbon pellets.
The unit requires a compressed air supply to operate, ultimately it is akin to many standard air
filtration / drying products. The inlet compressed air is passed into the first “bed”, as the air
passes across the carbon bed the oxygen is adsorbed, allowing the nitrogen in the compressed
air to carry on through the bed for collection and further use. After a certain time the online bed of
carbon will become saturated with oxygen. At this point in time, the control system will operate
various valves to bring the second bed of carbon on line. As the second bed comes on line the
first bed is vented to atmosphere to release the adsorbed oxygen and regenerate the bed ready
to be re-used again. This cycle will continue to repeat until the user stops consuming nitrogen.
Should the demand for Nitrogen be less than the rated output flow, or indeed should the demand
stop the generator will automatically go into ECO Mode and the front panel LED will illuminate. In
ECO mode the changeover of the columns is suspended which will stop the consumption of inlet
compressed air. The control system will automatically detect when the demand for Nitrogen
resumes and the generator will start to produce Nitrogen again.

Page 5

Operation Manual Nitropac N01x – N06x

General construction
The range will consist of 6 different sizes of the product. A 3D Cad image can be seen below:

BASE MODEL

6th SIZE IN RANGE

User Interface

CMS “Beds” ranging
from 1 to 6 sets
depending on model

Front End

The front end cabinet, control system and valves are consistent across the range. To increase the
outlet capacity as you go up the range of generators additional banks of CMS beds will be added.
(Note: for 5 to 6 column sets additional vent valves and silencers are required and are fitted to the
rear of the CMS columns)

Nitropac model number breakdown
N

DONALDSON RANGE

3
4
5

BASIC UNIT
NITROPAC WITH % O2 ANALYSER
NITROPAC WITH PPM O2 ANALYSER

Page 6

Operation Manual Nitropac N01x – N06x

Warranties and Liabilities
Claims for warranty and liability concerning personal injury or material damage become void if
caused by one or more of the following reasons:










Inappropriate use or application of the unit.
Incorrect installation, start-up, operation and maintenance of the unit.
Operation of a noticeably defective unit
Failure to follow operation manual procedures regarding concerning transport, storage,
installation start-up and maintenance of the unit.
Implementing constructional modifications to the unit.
Inadequate monitoring of parts on the unit that are subject to wear
Improper repairs
Use of non-original parts

Please also note the general warranty conditions for products of the manufacturer.

Page 7

Operation Manual Nitropac N01x – N06x

Safety Notices
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and
proper use of the Generator.

A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.

A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the Generator or the Application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.

Caution, risk of electric shock. Ensure power to the Generator has been removed
before proceeding.

Table 1 - Safety Symbols

Safety Notice to Users
These instructions must be read thoroughly and understood before installation and operation of
your Donaldson Nitropac. Use of the Generator in a manner not specified by Donaldson MAY
impair the SAFETY provided by the equipment.
When handling, operating or carrying out any maintenance, personnel must employ safe
engineering practices and observe all relevant local health and safety requirements and
regulations.

Nitrogen is not a poisonous gas, but if the concentration in the inhaled air
becomes too high there will be a risk of asphyxiation.
Ensure that adequate ventilation is provided for the surrounding area,
(depending on the operating specification and model selected the
generator can produce up to a maximum flow of 2,556 L/min)

Page 8

Operation Manual Nitropac N01x – N06x

Product Overview

PLC
Interface

Equalization Valve

Power Switch

Electrical Panel

Pressure Gauges 4x
Inlet Pressure Regulator

Equalization Valve

Changeover Valves 2x

DF Prefilter
DF Afterfilter

Outlet Flow
Controller

Exhaust silencers
2x

Outlet Pressure
Regulator

Exhaust Valves
2x

Page 9

Operation Manual Nitropac N01x – N06x

Pressure Switch Overview
There are two possible versions of pressure switch, pre-2016 units will have Black
FESTO pressure switches as below:

Output B indicator

Output A indicator

Pressure display

Bar graph display

Edit button

The pressure switch has two channels:
Channel A is used to trigger ECO mode
Channel B is used to trigger low pressure alarm
Setting Channel A (ECO mode)
Press edit once
Press up arrow to select channel A
Press edit
Press edit again to show SP
Using up or down arrows adjust to 0.05bar less than inlet pressure.
Press edit again to show HY (hysteresis)
Using up or down arrows adjust to 0.2bar
Press edit
Using up and down arrows set to NO (Normally open)
Press edit to save and exit
NOTE: The pressure switch is for indication only. Calibration is not required.
Pressure switch is pre-set at the factory, no adjustment should be required. The above is
for information only.

Page 10

Operation Manual Nitropac N01x – N06x

Post 2016 units will have the new White SMC Pressure switch, both perform the same
function and are fully interchangeable, setup instructions for SMC are as follows:

Again, the pressure switch has two channels:
Channel F1 is used to trigger ECO mode
Channel F2 is used to trigger low pressure alarm
F0 is used to set the switch pressure unit.

Page 11

Operation Manual Nitropac N01x – N06x

Electrical Panel Layout
Circuit
Breaker

Power
Supply

PLC

Expansion
Module

Triple Deck

Pressure
Switch

Control
Relays

Valve
Manifold

% Variant

Percentage
Oxygen Sensor

PPM Variant

O2 Analyser

PPM Oxygen
Sensor

Page 12

O2 Analyser

Operation Manual Nitropac N01x – N06x

Technical Specification
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature*
Maximum Relative Humidity
Maximum Altitude

+5°C (41°F)
+50°C (122°F)
80% (non-condensing)
2000m

When taken out of storage, the Generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.

Compressed Air Supply
Depending on your specific application, the pressure and flow required from the compressed air supply to
the generator will vary. For your specific requirements please refer to your quotation documents, or contact
Donaldson for further information. However all installations must meet the following conditions:

Minimum Air Quality
Minimum Inlet Air Pressure
Maximum Inlet Air Pressure
Minimum Inlet Air Temperature
Maximum Inlet Air Temperature

IS0 8573-1:2010 class 2.2.1
6 barg (87 psig)
10 barg (145 psig)
+5oC (41°F)
+35oC (95°F)

If you are in any doubt over the quality of your inlet compressed air DO NOT CONNECT to the generator,
and contact Donaldson. Donaldson can offer a full range of compressors and air preparation equipment if
required, which can be backed up by our global service support network.

It is the User / Installer’s responsibility to ensure the generator is connected
to a suitably rated air supply, the air supply must also provide suitable
protection to prevent over pressurization of the Nitrogen Gas Generator.

Electrical Requirements
Voltage

100 - 240 VAC ±10%

Frequency

50/60 Hz

Current

2.0 – 1.0 A

Input connection

C20 Plug

Power cord (Supplied)

C19 socket to local connection

Pollution degree

2

Installation category

Class I Protection

Transient Overvoltage

Category II

Page 13

Operation Manual Nitropac N01x – N06x
General
Model
Width mm(in)
Height mm(in)
Depth mm(in)
Weight kg(lbs)
Shipping weight kg(lbs)
Noise level
Model
Width mm(in)
Height mm(in)
Depth mm(in)
Weight kg(lbs)
Shipping weight kg(lbs)
Noise level

N01*

N02*

760(29.92)
197(433)
277(609)

920(36.22)
282(620)
364(801)

N06*
500 (19.68)
1738 (68.42)
1560(61.42)
622(1368)
712(1566)
59 dBA @ 1m

Page 14

N03*
500 (19.68)
1738 (68.42)
1080(42.52)
367(807)
451(992)
59 dBA @ 1m

N04*

N05*

1240(48.82)
452(994)
538(1184)

1400(55.12)
537(1181)
625(1375)

Operation Manual Nitropac N01x – N06x

Installation
It is the user/installer’s responsibility to ensure that the generator is located and
protected against any external influences such as vibration, shock, wind, snow
loading, earthquake or fire. The installation should conform to all local regulations
and should be leak tight and completed by technically competent personnel.

Final Location of the generator should be carefully considered, the largest model in the range will
weight 622 kg / 1368 lbs. Equipment is only to be installed on floor with a weight rating of min
1000kg/m2 or 200 lbs per Sq.ft.
Once in position the foot plate of the generator should be secured to the floor with fixings suitable
to the materials of construction of the floor. 13mm diameter holes are provided adjacent to the
levelling bolts to allow the unit to be fixed to the floor. Depending on the generator model you will
have 4 or 8 fixing positions. Typically 10mm or 3/8” floor fixings approx. 75mm or 3” in length will
be suitable on most concrete floors.

Page 15

Operation Manual Nitropac N01x – N06x
Generator Environment
The Generator is designed for indoor use only. It should be installed adjacent to the
application it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Consideration should be given to the location of the generator to ensure it is
protected from extreme fluctuations in ambient temperature. Ensure that adequate
ventilation is provided for the surrounding area, (depending on the operating
specification and model selected the generator can produce up to a maximum flow
of 2,556 L/min). Installation in a confined space or poorly ventilated space is not
recommended, however if you choose to do so ambient oxygen monitoring
equipment is recommended.
Please refer to the drawing below for the general dimensions of the unit.

Maximum Ambient Conditions: +50°C (dry bulb) 80%RH (Max) Non-Condensing

General Dimensions
500

760

1738

Note: Each additional tank
assembly = 160mm

1560

The Generator must always be placed on a level surface. Failure to do so
will affect the stability of the Generator.

Page 16

Operation Manual Nitropac N01x – N06x
Unit Controls
Column 1
Pressure Gauge

Column 2
Pressure Gauge

Control
Screen

Eco Mode
LED

Power Switch

Service Tank Pressure Outlet Pressure
Gauge
Gauge
LED

Inlet / Outlet Connections
Power Inlet

Nitrogen Output
¾” BSP Female

Air Inlet
1” BSP Female

Tank Return
¾” BSP Female

Page 17

Tank Supply
¾” BSP Female

Operation Manual Nitropac N01x – N06x
All of the Generator output ports are located on the output panel on the Left hand side of the
Generator. See below for recommended piping layout

Compressed Air Quality
The Nitropac Generator is an air purification system; it does not generate any gas pressure. Gas
pressure is created and supplied to the generator by the user’s compressed air system. It is the
user / installer’s responsibility to ensure that all components connected to the gas
generator comply with local health and safety regulations and that the compressed
air system is suitably protected from over pressure, including appropriately sized
safety relief devices for both the compressed air and nitrogen process tanks.

No pressure greater than 10 barg should be applied to the inlet of the Nitropac product.
The Nitropac Generator will typically be the last stage in a complete air compression and filtration
system, the inlet air quality must meet a minimum of IS0 8573-1:2010 class 2.2.1. Class 2.2.1. is
further defined as:
Class 2 – Particulate
Per cubic meter of air the particulate count should not exceed 400,000 particles in the 0.1 to 0.5
micron range, 6000 particles in the 0.5 to 1 micron range and 100 particles in the 1 to 5 micron
range.
Class 2 – Water
A minimum pressure dewpoint (PDP) of -40odegC (-40odegF) is required, no liquid water is
permitted.
Class 1 – Oil
Per cubic meter of air the maximum permissible oil content is 0.01mg, total level for liquid,
aerosol and vapor.
A typical installation will consist of the following items:













Air Compressor (Oil-Free or Oil Lubricated)
Compressed Air Storage Tank (with automatic condensate drain)
Oil/Water Separator
Pre-filter
Active Carbon Filter
Line Air Regulator
Desiccant Air Dryer
Dust Filter
Active Carbon Tower
Dust Filter
Nitropac Nitrogen Generator
Nitrogen Process Gas Tank

Sizing of all components in this line will have a critical effect on the performance of the Nitrogen
Gas Generator, in particular the Nitrogen process Gas Tank. Please refer to your quotation
documents for details of the items we have recommended to meet your specific requirements.

Page 18

Operation Manual Nitropac N01x – N06x
Should you require any further assistance or support please do not hesitate to contact
Donaldson. A full installation and commissioning service can be provided through the Donaldson.

Electrical Connection
Connect the Generator to a single-phase AC voltage supply using the power cord
provided. The generator is fitted with an internal transformer that can accept any supply
from 100 to 240 volts AC. If the appropriate power cord is not supplied; a new plug,
rated to at least 5 amps, can be fitted by a qualified electrician.

DO NOT USE inadequately rated detachable Mains cords.

This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED. Before
connecting the unit to the mains supply, please check the information on the serial plate.
The mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):- Green & Yellow

or

Green

LIVE (L):-

Brown

or

Black

Neutral (N):-

Blue

or

White

Recommended Piping Layout
To allow proper operation and commissioning of the generator it is important to include bypass
valves at the generator inlet / nitrogen outlet of the system. It is also recommended that a Nonreturn valve be fitted on the exit of the system to prevent downstream pressure from returning to
the system and damaging the generator. Please see below for recommended piping layout.

Page 19

Operation Manual Nitropac N01x – N06x

Commissioning / Safe Start-Up Procedure
With reference to the diagram above in “Recommended Piping Layout” section.
Once the generator has been installed, DO NOT Switch on the power immediately. It is
recommended that the inlet compressed air supply valve (1) remains closed, and that the inlet
bypass valve (2) is opened slightly to sufficiently allow the inlet compressed air to vent and purge
any remaining moisture from the newly installed pipework (if this is the first operation of the
compressed air supply, it is recommended that the air be vented for a minimum of 2hrs to allow
the air filtration and dryer to stabilise). Once the inlet pipework has been purged, the inlet bypass
valve (2) can then be closed and the pressure gradually introduced to the generator by slowly
opening the compressed air inlet (1) valve. The outlet bypass valve (3) should now be open to
atmosphere, and the delivery valve (4) to the customer application should be closed. Ensure that
both Process Tank Isolation valves (5) are both fully open. The power can then be turned on. On
first start up air will be introduced to Column 1 and the front panel gauge will slowly start to rise.
The front panel display will show a message “COLUMN 1” and a timer counting up. After approx.
30 to 80 seconds (timing will vary depending on your specific performance requirements) the front
panel display will briefly show an “EQUALIZING” message and the pressure on Column 1 and 2
gauges should level out. Column1 will then vent rapidly to zero and column 2 will continue to rise
slowly. During this process the TANK pressure gauge should continue to slowly rise. Allow the
Nitrogen Process Gas Tank to reach pressure, then continue to vent the outlet gas through the
bypass valve (3) for a minimum of 4 hours (ideally this should be done overnight to fully purge all
the remaining oxygen/moisture in the system). The Generator has been pre-set in the factory to
give the specified output flow-rate and pressure. Failure to achieve the factory specification after
maintenance may be as a result of an incorrect service procedure, please review any
maintenance carried out. If unable to achieve specification contact Donaldson for further
assistance. After this time the outlet bypass valve (3) can be closed, and the delivery valve (4) to
the customer application should be slowly opened to pressurise the line.

Page 20

Operation Manual Nitropac N01x – N06x
The design of the generator is that it will deliver up to your specific output pressure and flow of
Nitrogen. Should the demand for Nitrogen be less than the rated output flow, or indeed should the
demand stop the generator will automatically go into ECO Mode and the front panel LED will
illuminate. In ECO mode the changeover of the columns is suspended which will stop the
consumption of inlet compressed air. The control system will automatically detect when the
demand for Nitrogen resumes and the generator will start to produce Nitrogen again.

Page 21

Operation Manual Nitropac N01x – N06x

Normal Operation
The Nitropac Gas Generator is designed specifically to minimize operator involvement. Given that
the system is installed as described in earlier sections and is serviced in accordance with the
specified maintenance recommendations (see Service Requirements), then it should simply be a
matter of turning the Generator on. Note: the generator will only produce Nitrogen on demand. It
should typically be left in a powered state, with compressed air supply live. If the system is
shutdown, or suffers a power failure, it will restart automatically. However, if left without inlet air
pressure for a prolonged period and the Nitrogen Process tank is allowed to vent all pressure,
then the commissioning process above should be repeated to purge the oxygen from the system
before good quality gas can be produced again.
The Generator will automatically produce the factory set flow and pressure.
Note: The side panels of the generator should not be removed during operation
unless you have received training and are technically competent to manage the
potential risks present. Located inside the cabinet are the Vent silencers and
Safety Relief valve, these will periodically release gas at pressure, and could
cause injury.

On the rear and top of the generator are located inspection plugs. These plugs
must NOT be removed whilst the system is under pressure. If removed, it is likely
that pressure will be released, violently and cause injury.

Cleaning
Clean the outside of the Generator only using warm soapy water and a clean damp cloth. Ensure
the cloth is thoroughly rung out to remove excess fluid prior to use. Do not use decontamination
or cleaning agents that could cause a HAZARD as a result of a reaction with parts of the
Generator or material contained within it. If there is any doubt about the compatibility of
decontamination or cleaning agents please contact your Donaldson representative.
Cleaning should only be undertaken with the power switched off and the
power cord removed from the Generator.

Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to the
Generator.

Care should be taken with Leak Detections Liquids.

Page 22

Operation Manual Nitropac N01x – N06x

Service Requirements
Servicing and/or repair of the Generator should only be undertaken by a
TECHNICALLY COMPETENT PERSON with the Generator in a safely
isolated condition.

Safe Isolation Process
To shut the system down, close the inlet air supply valve (1), BUT leave the electrical power on to
the generator. Close the Nitrogen outlet valve (4) to the application, and slowly open the bypass /
commissioning valve (3) to allow the Nitrogen product gas to vent to atmosphere. NOTE: do not
isolate the Process Tank valves (5) as this will trap pressure in the tank. For the product to be
completely safe to work on ALL pressure must be fully dissipated. Ensure ALL front panel gauges
read zero before turning off the power on the front panel and removing the mains cord from the
left hand side of the generator before proceeding. Note: Due to the nature of the Carbon Sieve it
could take considerable time for the generator to release all the trapped oxygen. The bypass
valve (3) should remain open at all times, as the CMS can release oxygen and pressure can build
again if closed.
Once service operations have been completed the generator can be re-connected to the mains
supply and the Commissioning process described in earlier sections should be repeated.
Due to the simplicity of the design and the small number of moving parts the Industrial Nitropac
Series Nitrogen Generators will have a long and trouble free life. However as with all technical
equipment it should be regularly inspected and serviced as below.

Service Schedule

Service
interval
12 months
48 months
48 months
12 months
12 months

Component
Vent Silencer Element
Pneumatic Valve Exhaust Silencer
N2 Supply NRV
O2 Analyser % (#)
O2 Analyser PPM (#)
DF Pre-filter
DF After-filter

Qty.
N01x to N05x
2
5
2
1
1
1
1

Note # - Analyser service is dependent on selected model.

Page 23

N06x
4
7
2
1
1
1
1

Operation Manual Nitropac N01x – N06x

Service Counter Reset
The generators service hour counter counts the remaining time till service is due, starting at
8760hrs counting down to Zero. Once zero is reached the Service LED will illuminate, the
generator will continue to function, but should not be operated for prolonged periods without
carrying out the required maintenance. Once the required maintenance has been carried out the
counter can be reset through the front panel PLC Interface by following the procedure below.

In order to access the reset function, the engineer must follow the instructions below:
1. Press the ▼ button 3 or 4 times (depends if startup timer is active) until time date screen
is displayed, and then press ESC.
2. Press ▼ to highlight LOGO! Settings, then press ENTER.
3. Press ▼ to highlight Program, then press ENTER.
4. Press ▼ to highlight Set Parameter, then press ENTER, parameter list will be displayed.
5. Press ▼ until S_HRS-RESET is reached, then press ENTER. S-HRS RES is displayed
and switch should be highlighted as off.
6. Press ENTER and off will flash, press ▼ to change off to on, then press ENTER
7. Switch will flash back to off again, automatically.
8. Press the ESC button 5 times till time date screen is displayed.
9. Press ▲until Service Due screen is displayed, and confirm service due is now shown as
8760 HRS.
The service counter has now been successfully reset.

Page 24

Operation Manual Nitropac N01x – N06x

PLC Settings
In the Set Parameters menu of the PLC program there are 19 adjustable parameters. These will
be set during manufacture; however they can be changed to meet site and customer
requirements. The following explains the function of each parameter:
To gain access to set parameters from the main screen,
1.
2.
3.
4.

Press the ▼ button 3 or 4 times (depends if startup timer is active) until time date screen
is displayed, and then press ESC.
Press ▼ to highlight LOGO! Settings, then press ENTER.
Press ▼ to highlight Program, then press ENTER.
Press ▼ to highlight Set Parameter, then press ENTER, parameter list will be displayed.

The parameters will appear in the following order:
1. CYCLE TIMER - Main PSA change over timer do not alter without factory approval.
2. RE_START_EN – Remote Start Enable, only possible with optional factory configuration,
do not alter without factory approval.
3. START TIMER – From power switch on, the time in which the PLC will ignore High O2
content and low pressure alarms to allow the generator to reach required O2 content and
pressure.
4. TEST-ECO LED – Enables test of ECO mode indication LED.
5. TEST-SER LED – Enables test of Service Due indication LED.
6. RGEN-INTERVL – When the generator has been in ECO mode for set time the PLC will
ignore pressure switch and force the generator to run to maintain required O2 content.
7. RGEN BLOCK – Set time for the PLC to ignore pressure switch before returning to ECO
mode.
8. BPASS-Timer – If optional Bypass valve is fitted, this controls the time the bypass
feature can operate before a high O2 fault is triggered.
9. ECO-DELAY – Once pressure switch upper setting is reached time delay before
generator is switched into ECO mode.
10. ECO-BLOCK – Set to ON prevents ECO mode Function.
11. S_HRS-RESET – Service hour counter reset back to 8760 hours.
12. OV.LO-O.RIDE – If optional outlet valve is fitted, this allows you to override it and force it
open.
13. BPASS-O.RIDE - If optional bypass valve is fitted, this allows you to override it and force
it closed.
14. OP.OVALVE_EN – Set to on if outlet valve fitted.
15. OP.BYPASS_EN – Set to on if Bypass valve fitted.
16. ANALYSER-NIL – Set to on if no O2 Analyser is fitted.
17. ANALYSER-PPM – Set to on if PPM O2 Analyser is fitted.
18. ANALYSER-% – Set to on if percentage O2 Analyser is fitted.
19. SHUT-BLOCK – When an alarm condition happens with shutdown block switched off,
generator will shut down. When shutdown block is on, generator will continue to run with
an alarm condition (low pressure and high O2 content alarms).

Page 25

Operation Manual Nitropac N01x – N06x

Troubleshooting
Problem

Possible solution




The Generator will not switch on and
the power switch does not illuminate.








Contact Donaldson Technical Help.










Disconnect power cord from the rear of the Generator
and open LH panel:
Check that the circuit breaker is turned on (switch
in the up position). Reconnect power cord.
Check PLC for loose wires.
Check pressure switch for loose wires.
Connect power cord and power up Generator and open LH panel:
Check for 24V coming out of power supply.
Check pressure switch LED is lit.
Check PLC to ensure RUN light is illuminated.



Contact Donaldson Technical Help.





Check Generator has power ON.
Check inlet pressure regulator.
Check for leaks in system.



Contact Donaldson Technical Help.


The Generator will not switch on but the
power switch is illuminated.

Compressed air is on but pressure is
not building in system.

Ensure power cord is plugged into the Generator and
that the power socket is turned on.
Check the fuse in the power cord plug.
Check for any disconnected terminals at power
socket connected at inside rear of unit.
Check for loose connections at inside of power switch located on
front of Generator.
Check for 100-240VAC going into DC power supply.
Check for 24VDC coming from power supply.
Check circuit breakers are ON – switched up

Page 26

Operation Manual Nitropac N01x – N06x

Pneumatic Schematic - Rev 2A

Page 27

Operation Manual Nitropac N01x – N06x

Electrical Schematic – Rev 3

Page 28

Donaldson Filtration Deutschland GmbH
Industrial Filtration Solutions
Büssingstraße 1
D-42781 Haan, Deutschland

Tel. +49 (0) 21 29/5 69-0
Fax +49 (0) 21 29/5 69-1 00
Internet: www.donaldson.com
E-Mail CAP-de@donaldson.com

EU Declaration of Conformity
We hereby declare that the products indicated hereafter comply with the stipulations of the relevant guide lines
and technical standards. This declaration only refers to products in the condition in which they have been
placed into circulation. Parts which have not been installed by the manufacturer and / or modifications which
have been implemented subsequently remain unconsidered.
Product designation / series:
Type:

Serial number:

Nitropac
N013, N023, N033, N043, N053, N063
N014, N024, N034, N044, N054, N064
N015, N025, N035, N045, N055, N065
see type plate

Product description and function:

Pressure Swing Adsorption N2 Gas Generator

Low-Voltage Directive 2014/35/EU
Applied standards and technical
specifications:

: EN61010-1:2010 Safety requirements for electrical
equipment for measurement, control, and laboratory
use

The devices with a supply voltage of 24 V DC do not come under the scope of application of the Low-Voltage
Directive.

EMC Directive 2014/30/EU
Applied standards and technical
measurement, specifications:

EN 61326-1:2013 Electrical Equipment for
control and laboratory use.

Pressure Equipment Directive 2014/68/EU
Fluid group acc. to Art. 13:

Group 2

Applied rules, standards and technical specifications:

ASME Section VIII Division 1 and BS EN 12392:2000.
Conformity Assessment Procedure: B + D

Pressure Assembly Consisting of the following key parts:
Tag No. or
Line No. Piping Spec or Description
Front End Control Cabinet inc.
1
associated valves & piping
08-0700 CMS Assemblies
(1 to 10 Model dependent)
15

Output Pressure Relief Valve

Fluid

TS
Phase degC

PS
bar

Air / N2

Gas

50

10.00

Air / N2

Gas

50

Air / N2

Gas

50

PED Notified Body, responsible for monitoring
of Module D Quality Assurance System:

170628_Declaration of Conformity_EN_Nitropac.doc

PED Hazard Cat.
PS.DN
DN
V
mm litres PS.V Fluid Chart Cat
25

-

-

2G

2

10.00

-

50.96

509.6

2G

2

II

B+D

10.00

20

-

-

2G

2

IV

B+D

BSI Assurance UK Ltd
Kitemark Court
Knowlhill
Milton Keynes MK5 8PP
EC Number: 0086

SEP

Quality
Module
A

Donaldson Filtration Deutschland GmbH
Industrial Filtration Solutions
Büssingstraße 1
D-42781 Haan, Deutschland

Tel. +49 (0) 21 29/5 69-0
Fax +49 (0) 21 29/5 69-1 00
Internet: www.donaldson.com
E-Mail CAP-de@donaldson.com

Name and address of person, authorized to compile the technical file and documents:
Wolfgang Bongartz
Büssingstrasse 1
42781 Haan
Germany
This declaration of conformity is issued under the sole responsibility of the manufacturer.

Haan, 27.06.2017
Place and date of issue
Authorized Person
Wolfgang Bongartz
Engineering Manager Industrial

170628_Declaration of Conformity_EN_Nitropac.doc

Technical alterations reserved!
Technische Änderungen vorbehalten • Sous réserve de modifications technicques
• Technische gereserveerde wijzigingen • Alterazioni tecniche riservate
• Alteraciones técnicas reservadas • As alterações técnicas reservaram
• Ret til tekniske ændringer forbeholdes • Reserverade tekniska förändringar
• Oikeudet teknisiin muutoksiin pidätetään • Αλλαγές που διατηρούνται τεχνικές
• Technické změny vyhrazeny • Jätame endale õiguse teha tehnilisi muudatusi
• Műszaki módosítások joga fenntartva • Pasiliekame teisę daryti techninio pobūdžio pakeitimų
• Lespējamas tehniskas izmaiņas • Zmiany techniczne zastrzeżone • Tehnične spremembe pridržane •
Technické zmeny vyhradené • Ne rezervăm dreptul de a efectua modificări tehnice
• Запазваме си правото на технически промени • Teknik değişiklikler saklı tutulur