IQ OQ - PRECISION HYDROGEN SL
Original PDF File210 KB |
Download |
View directly in browser210 KB |
View in browser |
Resource ID
3582
Access
Open
Contributed by
Derek Rennie
Updated date
01 July 20
Review date
01 May 24
Title
IQ OQ - PRECISION HYDROGEN SL
Revision Number
1
Revision Details
Intital upload
Language
UK English
Company Brand
Peak Scientific
Extracted text
INSTALLATION QUALIFICATION
Instrument Origin and Identification
Instrument Series:
PRECISION HYDROGEN
Model:
Precision Hydrogen SL
Instrument Description:
Hydrogen Gas Generator
Flow Variant:
100cc
200cc
Water Load Variant:
Manual Fill
Pumped Fill
Pressure Fill
Colour Variant:
Black
White
Manufacturer:
Peak Scientific Instruments Ltd
Fountain Crescent
Inchinnan Business Park
Inchinnan
Renfrew PA4 9RE
Scotland, UK
Service Contact:
Peak Scientific Instruments Ltd
Tel.
+44 (0)141-812-8100
Email support@peakscientific.com
Serial Number:
Condition when Received:
New
Used
Reconditioned
Date Received:
Installed Location:
Purchase Order Number:
Identification / Asset Number:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 1 of 8
INSTALLATION QUALIFICATION
Instrument Specification
Instrument Parameter
Specified Range
Acceptable
Output
Pressure
100 psi (6.9 bar)
Hydrogen Flow Rate
100cc/min (ATP)
(Select as appropriate)
200cc/min (ATP)
Environment and Siting
Has the instrument been adequately acclimatized since transport or storage?
(Requires 3 hours before operation)
Yes
No
Operating Parameter
Ambient Temperature
Maximum Relative Humidity
Specified Range
10-35 ºC (50-95 ºF)
90% Non-Condensing
Maximum Altitude
Supply Voltage & Frequency
Acceptable
3000 m (9800 ft)
100-240V ±10%, 50/60Hz
Refer to User Manual for acceptable range(s)
Siting – Generator Placement
Sit Generator on a Flat Surface
Minimum Clearance around Machine – 75mm
Siting – General Lab Conditions
Mains Power Outlet
Water Purity Requirement
Extraction Fan/Fume Hood (Recommended)
Within 2m Radius to Generator
<1.0µ Siemens/cm OR >1 Mohm-cm
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 2 of 8
INSTALLATION QUALIFICATION
Instrument Delivery and Documentation
Unpack the instrument carefully. Are any items missing against the Fitting kit list as per
Installation Guide?
Yes
No
If yes, state missing items:
Is there any damage to the instrument?
Yes
No
Yes
No
If yes, describe damage:
Manufacturer informed:
Fittings Kit
Present
Missing
Present
Missing
Fittings Kit included
Is all standard documentation included?
User Manual (Available via download from Peak Extranet)
Warranty Registration Card
Installation Guide
Site Preparation Guide
(Available via Download from Peak Extranet)
Installation
Installed By:
User
Authorized Engineer
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 3 of 8
INSTALLATION QUALIFICATION
Summary Report
Instrument:
Precision Hydrogen SL
Serial Number:
Assessment of Complete Installation Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument has been
satisfactorily delivered and installed. The instrument has passed the Installation Qualification
procedure and may now be submitted for Operational Qualification.
IQ Completed By:
Date:
Deviations Approved By:
Date:
IQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 4 of 8
OPERATIONAL QUALIFICATION
Pre-Run Checks
Pre-Run Check
OK
Discrepancy
N/A
Check water purity
Deioniser column installed
Connect delivery connection to instrument
Vents to fume hood
Pumped and Valved Fill generators only*
Position of water bottle
Water inlet pressure 1 ≤ P ≤ 3 bar
Connect inline water filter
Connect water supply
Comments:
*Refer to Installation Guide and User Manual for details
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 5 of 8
Functional Tests
Functional Test
Acceptance Criteria
Pass
Fail
Fan Running
Apply Mains Power / Switch On
Bubbles produced in water tank
Water tank LED pulsing green
Water tank LED solid dim green
Allow to run for 10 minutes
Bubbles in water tank cease
Water tank LED turns solid
bright green
Press front button
Output pressure 100 psi*
No leaks between generator
and instrument
Output connections leak tight†
Notes:
* If lower pressures are required, an external pressure regulator should be fitted to the output port of
the generator.
Liquid leak detectors should not be used on the Precision Hydrogen. If a liquid solution must be
used, then it should be IPA based, or a portable Hydrogen detector should be used.
Serial number of calibrated flow meter:
Calibration due date of flow meter:
Serial number of calibrated pressure gauge:
Calibration due date of pressure gauge:
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 6 of 8
OPERATIONAL QUALIFICATION
Training
Date Instrument Released for Use:
Trained Operators
Read User
Manual
Practical
Training
Authorized Signature
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 7 of 8
OPERATIONAL QUALIFICATION
Summary Report
Instrument:
Precision Hydrogen SL
Serial Number:
Assessment of Complete Operational Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument is
operating satisfactorily following delivery and installation. The instrument has passed the Operational
Qualification procedure and may now be released for use.
OQ Completed By:
Date:
Deviations Approved By:
Date:
OQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-983
Revision: 1
Page 8 of 8