Precision Nitrogen (All Models)
User Manual

Copyright © Peak Scientific Instruments Ltd 2018 — Precision Nitrogen User Manual Rev 6 RSID 226 EN —24/04/2018

Contents
Change History
3
How to use this Manual
3
Introduction 4
Warranties and Liabilities
5
Safety Notices
6
Symbols 6
Safety Notice to Users
6
Declaration of Conformity
7
Environmental Declaration
8
Technical Specification
9
Unpacking 11
Fittings Kit Contents
12
Unpacking Instructions
12
Installation 13
Generator Environment
13
Generator Overview
14
Nitrogen General Dimensions
14
Nitrogen Trace General Dimensions
15
Nitrogen Rear Connections
16
Nitrogen Trace Rear Connections
17
Electrical Connection
18
Air Purity
19
Class 1 Particulate
19
Class 4 Water
19
Class 1 Oil
19
Start-Up Sequence
20
On Demand Gas
20
Unusual Operation
20
Connecting to the application
21
Tubing Lengths
21
Service Requirements
22
Service Schedule
22
Peak Protected
23
Cleaning 24
Troubleshooting 25

Page 2

Change History
Rev

Comment

Name

Date

1

Release

Liam Couttie

14/12/2016

2

Technical Specification Update

Liam Couttie

08/02/2016

3

Nitrogen Headspace Added

Liam Couttie

06/09/2016

4

Product Overview Correction

Liam Couttie

21/12/2017

5

Cleaning Information Update

Liam Couttie

28/02/2018

6

Declarations Update

Liam Couttie

24/04/2018

How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.

Page 3

Introduction
Precision is a modular system with differing flow rates that allows you to add multiple
generators as your gas and analysis demands change. The Precision Series will keep
your gas flowing with consummate accuracy and reliability, whatever your needs, while
fitting neatly into any lab space.
The Precision Nitrogen, Headspace and Nitrogen Trace Generators provide a constant
and consistent source of Nitrogen for detector make-up gas for GC applications, as
well as headspace vial pressurization, purge and trap, tube conditioning for thermal
desorption and sample blow down. In addition to these applications, the Precision
Nitrogen Trace model can be used for carrier gas in GC applications as it also removes
hydrocarbons by means of catalytic oxidation.
Other features include:
• Generates Nitrogen on demand from compressed air.
• Regenerative PSA Carbon Molecular Sieve to ensure Oxygen and Moisture removal.
• Minimum maintenance with only a filter change per annum.
• Small and Stackable.
With the Precision Nitrogen based on proven technology, it removes moisture and
oxygen to leave Ultra High Purity Nitrogen. The Nitrogen Trace further removes
Hydrocarbons (as Methane) to To ensure these generator models meet our high expectations with regards to reliability
and performance, we have tested them extensively at our manufacturing plant and with
end users around the world to ensure reliability and longevity of the system.

Page 4

Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods
shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free
from defects in material and workmanship for the lesser of a period of twelve
months from the date of delivery or thirteen months from the date of dispatch
from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person
dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common
law are hereby expressly excluded save to the extent they may not be lawfully
excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of
this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of
clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality
of the foregoing that any defect is not the direct or indirect result of lack of repair
and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or
incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant
packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub-
clause 2 hereof within a reasonable time of it becoming apparent and subject always
to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or
repair the defective Goods or refund a proportionate part of the Price. The Company
shall have no further liability to the Customer (save as mentioned in sub-clause 6
hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim
for death or personal injury to any person in so far as such is attributable to the
negligence or breach of duty of the Company or any failure by the Company to
comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect
of any claim by the Customer for costs, damages, loss or expenses (whether direct,
indirect, consequential or otherwise) or indemnity in any respect howsoever arising
including, but not by way of limitation, liability arising in negligence (other than
pursuant to clause 6 above) that may be suffered by the Customer or any third
party.

Page 5

Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.

Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the generator.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.

Safety Notice to Users

WARNING

WARNING

These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.

WARNING

Page 6

Declaration of Conformity
We

Peak Scientific Instruments Ltd.

Of

Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE

Declare that:
Equipment: Nitrogen Generator


Models:



Precision Nitrogen 250cc, 600cc, 1000cc, Headspace
Precision Nitrogen Trace 250cc, 600cc, 1000cc


To which this declaration relates, is in conformity with the applicable EC directives,
harmonized standards, and other normative requirements.
• Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Electrical Equipment for measurement, control and laboratory use.
• Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for measurement, control and laboratory use.
• FCC47 CFR Part 15 Class B
Unintentional radiators; Conducted and Radiated emissions limits.
All evaluation, testing and certification issued by:


Nemko Canada Inc.

York EMC Services LTD





303 River Road, Ottawa

Grangemouth Technology Park,



Ontario, Canada

Earls Road, Grangemouth, UK

K1V 1H2 FK3 8UZ

Signed: Name: Chris Pugh
Date:

Page 7

26/10/2016

Position:

Chief Technology Officer

Environmental Declaration
We

Peak Scientific Instruments Ltd.

Of

Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE

Declare that:
Equipment: Nitrogen Generator


Models:



Precision Nitrogen 250cc, 600cc, 1000cc, Headspace
Precision Nitrogen Trace 250cc, 600cc, 1000cc

Is fully compliant with the following Directives
• 2012/19/EU

WEEE (Waste of Electrical and Electronic Equipment)

• 2011/65/EU

RoHS 2 (Restriction of Hazardous Substance)

Peak Scientific Instruments Ltd fully complies with its obligations towards the European
WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These
obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to
ensure that our products comply with the principles and requirements of the European
recast RoHS (Restriction of Hazardous Substances) Directive 2011/65/EU. Similar
directives in the United States and China, for example, have also been captured within
this program.
Where a specific certificate of compliance is required, this can be requested, on
a product serial number basis, directly from Peak Scientific Instruments Ltd, by
contacting us through our website on www.peakscientific.com
Signed: Name: Chris Pugh
Date:

Page 8

10/06/2015

Position:

Engineering & Operations Director

Technical Specification
Precision Nitrogen
Environment
250cc

Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*

600cc

1000cc

5°C (41°F)
35°C (95°F)
2000 m
70% Non-Condensing
-20°C (-4°F)
60°C (140°F)

Inlet Conditions
Inlet Air Pressure Min-Max
Minimum Air Inlet Flow
Minimum Air Inlet Quality
Particles

6.9-8.3 bar (100-120 psi)
35 l/min
ISO 8573-1:2010 Class[1.4.1]

Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-up time
Particles
Phthalates
Suspended Liquids
Gas Outlets

250cc

5.5 bar (80 psi)
600cc
1.5 Hours
NONE
NONE
1 x ¼” BSPP

1000cc

For optimum performance from your Precision Nitrogen generator, the inlet gas supply pressure should be
30PSI greater than the required output pressure.

Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category

100-240v
50/60 Hz
0.4-0.2A
C14 Plug
C13 Socket to Local Connection
2
II

General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 9

38.0 (14.9) x 54.0 (21.2) x 25.6 (10.0)
21 (46.2)
26 (57.3)
24 (53)
29 (64)

Technical Specification
Precision Nitrogen Headspace
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*

5°C (41°F)
35°C (95°F)
2000 m
70% Non-Condensing
-20°C (-4°F)
60°C (140°F)

Inlet Conditions
Inlet Air Pressure Min-Max
Minimum Air Inlet Flow
Minimum Air Inlet Quality
Particles

8.3-8.6 bar (120-125 psi)
18 l/min
ISO 8573-1:2010 Class[1.4.1]

Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-up time
Particles
Phthalates
Suspended Liquids
Gas Outlets

6.9 bar (100 psi)
250cc
3 - 3.5 Hours
NONE
NONE
1 x ¼” BSPP

For optimum performance from your Precision Nitrogen generator, the inlet gas supply pressure should be 30PSI greater than the
required output pressure.
Though the Precision Headspace Generator delivers nitrogen at a purity suitable for most non- trace applications, we would
recommend the installation of oxygen and hydrocarbon traps for analysis at lower detection limits (
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category

110-230v ±10%
50/60 Hz
0.37 -0.17 A
C14 Plug
C13 Socket to Local Connection
2
II

General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 10

38.0 (14.9) x 54.0 (21.2) x 25.6 (10.0)
21 (46.2)
24 (53)

Technical Specification
Precision Nitrogen Trace
Environment
250cc

Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*

600cc

1000cc

5°C (41°F)
35°C (95°F)
2000 m
70% Non-Condensing
-20°C (-4°F)
60°C (140°F)

Inlet Conditions
Inlet Air Pressure Min-Max
Minimum Air Inlet Flow
Minimum Air Inlet Quality
Particles

8.3-10 bar (120-145 psi)
18 l/min
22 l/min
ISO 8573-1:2010 Class[1.4.1]

Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-up time
Particles
Phthalates
Suspended Liquids
Gas Outlets

250cc

5.5 bar (80 psi)
600cc
1.5 Hours
NONE
NONE
1 x ¼” BSPP

1000cc

Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category

115 VAC ±10%
60 Hz
4.4A

230VAC ±10%
50 Hz
2.1A

C14 Plug
C13 Socket to Local Connection
2
II

General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)

Page 11

38.0 (14.9) x 54.0 (21.2) x 40.6 (15.8)
38 (83.6)
38 (83.6)
36(79.2)
42 (92.4)
42 (92.4)
40(88)

Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking.

Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator
onto the bench.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful
not to discard these with the packaging. 

Page 12

Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. Teflon Tubing x 3m
2. ¼” Compression Fitting x 1
3. 1⁄8” Compression Fitting



x 1

4. UK Mains Power Cable x 1
5. EU Mains Power Cable x 1
6. US Mains Power Cable 110v

x1

7. US Mains Power Cable 230v

x1

All of the generators output ports are located on the output panel at the rear of the
unit.

Unpacking Instructions
The unit weight constitutes a two person lift and as such, safe lifting practices should
be employed; do not attempt to lift on your own, as you will significantly increase the
chance of injury & damage to yourself and others around you.
1. Remove the screws encircled in red around the bottom of the crate lid and lift
upwards.
2. Next, with someone on either side of the unit, position your hands underneath the
unit ready to lift.
3. Ensuring your knees are bent and your back is straight, lift the unit to the desired
location.

Page 13

Installation
Generator Environment
The generator is designed for indoor use only. Storing of corrosive substances near
the generator or compressed air supply is not recommended. Saline salt solution
spillages around the generator or their compressed air supply should be cleaned up
immediately. It should be installed adjacent to the application(s) it is supplying. If this
is not convenient then the unit can be sited elsewhere, however, consideration should
be made of the lengths of pipe runs as pressure drops can result from extended runs of
pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation
of the unit within such a pocket could adversely affect its performance. Consideration
should also be given to the air flow around the unit. It is recommended that an air gap
of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please
refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:

5 °C (41 °F)

Maximum Operating Ambient Temperature:

35 °C (95 °F)

Page 14

Generator Overview
Nitrogen General Dimensions

380 mm / 14.9”

256 mm / 10"

540 mm / 21.2”

The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING

Page 15

Nitrogen Trace General Dimensions

380 mm / 14.9”

406mm / 15.9”

540 mm / 21.2”

The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING

Page 16

Nitrogen/Headspace Rear Connections
Power Inlet

¼" Compression Fitting

⁄8” Compression Fitting

1

Mains Cable

Page 17

Nitrogen Trace Rear Connections
Power Inlet

¼" Compression Fitting

Mains Cable

⁄8” Compression Fitting

1

Page 18

Electrical Connection
Connect the generator to an appropriate 110 or 230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.

WARNING

This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.

EARTH/GROUND (E):-

Green & Yellow or

Green

LIVE (L):-

Brown

or

Black

Neutral (N):-

Blue

or

White



Electrical requirements are 110 or 230VAC nominal +/- 10% depending on chosen
model. However, running continuously at voltages outwith this is not recommended.
Extended periods at extremes can have a detrimental effect on the operation and life of
the generator.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING

Page 19

Air Purity
The Precision Nitogen Series generator should be connected to an air supply that, as a
minimum, meets ISO8753-1:2010 Class 1.4.1

Class 1 Particulate
In each cubic metre of compressed air, the particulate count should not exceed 20,000
particles in the 0.1 - 0.5 micron size range, 400 particles in the 0.5 - 1 micron range and
10 particles in the 1 - 5 micron size range.

Class 4 Water
A pressure dew point (PDP) of +3°C or better is required and no liquid water is allowed.

Class 1 Oil
In each cubic metre of compressed air, not more than 0.01mg of oil is allowed. This is a
total level for liquid oil, oil aerosol and oil vapour.

Page 20

Start-Up Sequence
Switch the generator on using the power switch on the rear panel of the unit. Please
note: Air will only flow through the generator when the power is switched on. The LED’s
on the front panel will illuminate YELLOW for the start-up period.

YELLOW for 90 Minutes

The generator will purge impurities from its internal storage tank until Oxygen content
is less than 5PPM, approximately 1.5 hours (3-3.5 hours for Headspace) of uninterrupted
operation. On completion, the Fascia will illuminate GREEN or BLUE (70-**** Part
Numbers Only) and the generator is now ready to be connected to the application.

GREEN / BLUE



On Demand Gas
The generator will produce nitrogen on demand. If the application is operating
and requires a gas flow, the system it is connected to will supply this to suit the
requirements of the application. If the application requirement for gas stops, the system
will also stop, once it has reached its upper set limit in the internal storage tanks. If the
demand from the application starts again, the system will detect the demand for gas
and will automatically start again to suit the demand.

Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.
Page 21

Connecting to the application
Once the initial purge run of 1.5 hours (3-3.5 hours for Headspace) has completed the
generator is now ready to be connected to the application(s).
The pressure in the internal storage tanks must be allowed to dissipate
before connecting the generator to the application(s)
CAUTION

Attach the ¼” compression fitting to the outlet of the generator. Using the ¼” tubing
supplied, connect the outlet of the generator to the inlet on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.

CAUTION

Once the tubing is connected to the application, please ensure that it
is thoroughly checked for being leak-tight. Even the slightest leak in
the gas supply between the generator and the application can lead to
a reduction in efficiency.

Tubing Lengths

CAUTION



The diameter of the tubing which will be connected to the gas outlet is
important and is determined by the length of tubing required. Failure
to follow these recommendations could lead to an excessive pressure
drop between the generator and its application.


Use ¼”/3⁄16” (¼” O/D, 3⁄16” I/D) P.T.F.E. tubing.


> 10 - 40 meters: Use 3⁄8”/ 5⁄16” (3⁄8” O/D, 5⁄16” I/D). Tubing and fittings not supplied
in the fittings kit.

> 40 metres:


Please contact Peak Scientific with the relevant distance and
we will calculate the flow resistance and the tubing size required.

A combination of ¼”/3⁄16” and 3⁄8”/5⁄16” tubing may be used to ensure that there is no large
diameter tubing within the lab (i.e. for the first 20 meters from the generator use 3⁄8”/5⁄16”
and the final 10 meters to the application use ¼”/3⁄16” tubing). Keep the connections and
bends to a minimum.

Page 22

Service Requirements
Service Schedule
Purchase Interval

12 months

Component
Precision Nitrogen Annual Service Kit

www.peakscientific.com/
ordering
Precision Nitrogen Trace Annual Service Kit

Page 23

Visit

Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.

[Peak Protected] can provide…

Installation

A dedicated Peak engineer will visit
your lab to install and setup your
generator

IQ/OQ

Certified assurance for applications
requiring documented qualification

Complete plan

Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance

Spares

Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime

Premium Protected

Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance

Technical Support Hotline

Around the clock support by phone
or online with our global technical
helpdesk

To find out more about protecting your investment visit: www.peakscientific.com/protected

Page 24

Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the generator.

CAUTION

Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator. Chlorine-based beach cleaning solutions should not be
used on the generator or within their operating environment.
Care should be taken with Leak Detections Liquids.

CAUTION

Page 25

Troubleshooting
Problem

Possible Solution
• Ensure power cord is plugged into the
generator and that the power socket is
turned on.

The generator will not switch on and the
power switch does not illuminate.

• Check the fuse in the power cord plug.

The fascia LED’s have not illuminated but the
power switch is illuminated.

• Contact your service provider.

The instrument is reporting low pressure.

• Contact your service provider.

• Check for leaks between the generator
and instrument.
• Contact your service provider.

The fascia Yellow illumination does not
change to green/blue.

Page 26

• Contact your service provider.

Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.

Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.

Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?

Yes

What gas requirements do
you have in your lab?

Hydrogen

No

Nitrogen

Zero Air

Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only






Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 27

Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com

Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com

Copyright © Peak Scientific Instruments Ltd 2018 — Precision Nitrogen User Manual Rev 6 RSID 226 EN —24/04/2018