Extracted text
Precision Hydrogen Trace
Agilent Approved Hydrogen Generator
Part Number : 74-0250 / 74-0500
Service Kit : See Reverse
Your local gas generation partner
Description
The Precision Hydrogen Trace generator is designed primarily for
GC carrier gas use, and can also be used for detectors requiring
Hydrogen fuel gas such as FID and FPD. This model is approved
for use specifically with Agilent Technologies GC instruments,
with a single generator capable of supplying multiple devices
depending on model and flow requirements.
These generators offer a safe, reliable and convenient source
of Hydrogen for GC, generating very high purity Hydrogen by
utilising a Proton Exchange Membrane to create the gas from
deionized water, prior to Pressure Swing Absorption and
molecular sieve technology to remove moisture content down
to trace levels. Precision Hydrogen Trace generators come with
highly robust safety features as standard, providing peace of mind
in the laboratory and a far safer, dependable and more convenient
alternative to cylinder gas.
Applications
)
Carrier Gas
• GC-MS
• GC
)
540 mm
Detector Gas
• Fuel Gas
380 mm
406 mm
)
Suitable for flame gas and carrier gas at trace detection limit.
)
99.99999% purity
)
Approved for use with
Agilent Technologies GC
instruments
)
Internal leak detection with auto-shutdown safety features
)
Proven PEM technology to generate hydrogen reliably
)
Certified to latest revision CSA standards for peace of mind
and compliance
)
Regenerative PSA dryers to ensure high level of hydrogen purity
)
Minimal maintenance required (deionizer cartridge replacement)
)
Water bottle included, protecting against contamination
)
Combine multiple units for higher flow requirements
)
GC in-oven hydrogen leak detector available as an optional extra
)
Automatic loading pump as standard
)
Creates hydrogen on demand, minimal storage of hydrogen in
the system
)
Compact space-saving modular design
)
3 year cell warranty as standard
Precision Hydrogen Trace shown in a typical Precision stack featuring additional
Nitrogen and Zero Air modules
Copyright © Peak Scientific Instruments Ltd 2022 — Precision Hydrogen Trace Agilent Data Sheet RSID 359 Rev 9 EN —02/05/2022
Technical Specifications
Hydrogen Trace, 250cc
Hydrogen Trace, 500cc
250 cc/min
500 cc/min
Max Flow Rate*
Max Pressure
100 psi / 6.9 bar
Purity
99.99999%**
Dew Point
-75 °C / -103 °F
Gas Outlets
1 x 1/8” Swagelok compression fitting
Water Purity Requirement
<1.0μ Siemens/cm OR >1 Mohm-cm
Water Consumption
0.17 - 0.46 L/day
0.4 - 1.2 L/day
Operating Temperature
10°C - 35°C / 50°F - 95°F
Electrical Requirements
110-230V 50/60HZ 6A
Power Consumption
477 Watts
787 Watts
Heat Output
up to 4706 BTU/hr
Generator Dimensions (WxDxH)
406 x 380 x 540 mm / 16 x 15 x 21.3’’
Generator Weight
31 Kg (68.3 lbs)
Noise Level
Silent in operation
Water bottle
4 liter Precision Water Bottle included
* All flow rates are stated in sccm at 273.15K and 1.01bar
** Based on O2 content independently verified by National Physical Laboratory, UK
Ordering Information
Part Number
74-0250
74-0500
Annual Service
Standard Maintenance Plan
www.peakscientific.com/ordering
Complete Maintenance Plan
Accessories
Water Bottle 8L
Hydrogen Leak Detector (GC in-oven)
Part Number
10-9017
10-9010
NB: 8L Water Bottle is optional accessory, 4L Water Bottle is supplied with the generator.
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 12 month warranty. Beyond this period however you can ensure that you’re investment continues
to be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you the
reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a breakdown.
Peak Scientific UK
Tel: +44 (0)141 812 8100
Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649
Fax: +1 978 608 9503
Peak Scientific’s Quality
Management System
conforms to: ISO:9001:2008
For a full list of our worldwide
office locations, please visit:
Web: www.peakscientific.com
Email: marketing@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2022 — Precision Hydrogen Trace Agilent Data Sheet RSID 359 Rev 9 EN —02/05/2022