NM30LA - User Manual
Original PDF File3.0 MB |
Download |
View directly in browser3.0 MB |
View in browser |
Resource ID
102
Access
Open
Contributed by
Derek Rennie
Review date
01 May 24
Title
NM30LA - User Manual
Revision Number
1
Revision Details
Reviewed 31/12/20 - BP
Keywords
User Manual,
Company Brand
Peak Scientific
Reviewed on 3rd may 2017 - LC
Extracted text
NM30LA-MS GAS GENERATOR USER MANUAL NM30LA-MS-230v (19”RACK SYSTEM)
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Contents
Page
Contents Document Change History Warranty Statement Declaration of Conformity Technical Specification Introduction Unpacking and Installation Electrical connection Unpacking and Installation and connection Installation – General Controls Installation – Generator Environment Installation - General Dimensions Operation - Principle of Operation Operation - Controls Maintenance - Routine Maintenance Maintenance – Routine Maintenance Maintenance – Routine Maintenance Pneumatic Diagram Maintenance Log Notes - Additional Notes 2 3 4 5 6 7 8 9 10 11 11 12 13 14 15 16 17 18 19
2
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
History
Rev. No.
1 2 3 4 Document Created
Changed
Initials
GL TG TG SK
Date
09/11/06 08/05/07 25/09/02 06/02/09
Waste Bottle Note Added Notes Added for –Indoor use only & Altitude Updated to New Format
3
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Warranties & Liabilities
Warranties and Liabilities
1. The Company warrants that it has title to the Goods. 2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory. 3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act 1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause. 4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must:4.1. reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts; and 4.2.allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company. 5. Subject to the Company being notified of any defect as is referred to in sub-clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof). 6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof. 7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
SAFETY NOTICE TO USERS These instructions must be read thoroughly and understood before installation and operation of your Peak Scientific NM30LA. Use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations. WARNING: Nitrogen is not a poisonous gas, but if the concentration in the inhaled air becomes too high there will be a risk of asphyxiation.
4
UM-NM30LA-MS Gas Generator Declaration of Conformity
UM –NM30LA-MS
Declaration of Conformity
5
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Technical Specification
Generator Environment
Minimum Operating Ambient Temperature Maximum Operating Ambient Temperature Maximum Relative Humidity Maximum Altitude Outlet Gas Maximum Gas Outlet Pressure Maximum Gas Outlet Flow Start up time for Purity Particles Electrical Requirements @230V±10% ac (50/60Hz) Compressor Circuit Breaker Control Circuit Breaker Electrical Connection General Dimensions W x D x H Weight Shipping Weight cm (inches) Kg (lbs) Kg (lbs)
5 oC (41 oF) 30 oC (86 oF) 70% 2000 Meters
100 psig (6.8 barg) 30 Litres /min (ATP) 60 minutes 0.01um
5.0 Amps 8.0 Amps 2.0 Amps Single Phase Power Cord
60 x 67 x 94 (24 x 27 x 37) 112.5 (248) 169 (372.58)
6
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Introduction Introduction
Welcome to the User Manual for the Peak Scientific Instruments NM30LA-MS Laboratory Gas Generator. Enclosed in this manual you will find the information required to ensure that your generator is operated & serviced according to our recommended guide lines which will prepare you for long and trouble free gas generation. Please review each of the following sections carefully and ensure that the maintenance log at the rear of this manual is updated for future reference. Thank you for selecting Peak Scientific Instruments (PSI) to meet your Gas Generation needs, and should you require any further assistance or support please do not hesitate to contact us at the addresses on the front cover of this manual.
7
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Installation Unpacking and Installation
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any signs of transit damage.
ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS PURCHASED.
Following the un-packing instructions posted on the side of the crate. It will require 2 people to lift the crate clear and to maneuver the generator onto the floor. Thereafter the generator can be moved to its final location on the castors provided.
NOTE: - Included with the generator is a pack containing keys, manuals, service filters & fittings. Be careful not to discard these with the packing. Please save the product packaging for storage or future shipment of the generator.
Following installation of the unit, care should be taken to ensure that the waste bottle is NOT a sealed container, as this may cause possible pressure build-ups, explosion of the bottle, and consequential damage to the unit. Any bottle used should have a form of exhaust hole, before being fitted to this unit.
Electrical Connection
This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
Important Electrical Notice
EARTH/GROUND (E):LIVE (L):Neutral (N):-
Green & Yellow Brown Blue
OR
Green Black White
Connect the generator to a single-phase supply using the power cord provided.
8
UM –NM30LA-MS
Rear View.
1 off 1/4" x 1/4" Compression Fitting (02-4421) with White Teflon 1/4" Tubing (00-1266)
Depending on what connection your Instrument has at the inlet. You will require Either. 1 off 1/4" x 6mm Push-Fit (02-4331) with White Teflon 6mm Tubing (00-1302)
UM-NM30LA-MS Gas Generator
OUTLET
1 off 1/8" x 6mm Push-Fit (02-4330) with Yellow Nylon 6mm Tubing (00-1306)
1 off 1/8" x 6mm Push-Fit (02-4330) with Yellow Nylon 6mm Tubing (00-1306)
SUPPLY DRAIN DRAIN
Power Cord (04-1038-EU) (04-1029-US) (04-1028-UK) 230Vac
Connections
9
UM –NM30LA-MS
Front View.
Nitrogen Pressure Gauge.
Air Outlet Fans.
Nitrogen Generator
UM-NM30LA-MS Gas Generator
Nitrogen Pressur e
Nitrogen Outlet. Elapsed Time Meter.
0
OUTLET
1
Serial Plate. On/Off Switch. Mains Power IEC. Socket.
Drain Connection. Drain Connection.
SUPPLY DRAIN DRAIN
Air Inlet Fans.
General Controls
10
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Generator Environment
Generator Environment
The Generator is designed for indoor use only. The unit should be installed adjacent to the Mass Spectrometer it is supplying. If this is not convenient then the generator can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe. Performance of the generator (like all sophisticated equipment) is affected by ambient conditions. Continuous operation in ambient temperatures exceeding 25ºC will lead to a reduction in capacity and prolonged operation in temperatures exceeding 30ºC will shorten the life of the unit. Note should also be taken to the proximity of Air Conditioning outlets. These can sometimes give rise to ‘pockets’ of air with high relative humidity. Operation of the generator within such a pocket could adversely affect its performance. Consideration should also be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”) should be maintained down both sides and across the top of the unit. Please refer to the drawing below for the general dimensions of the generator. MAXIMUM AMBIENT CONDITIONS: 30ºC (dry bulb) 70%RH (Max.)
General Dimensions
Gas Generator API Systems
Nitroge n Pressure
0 1
11
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Principle of Operation
Peak Scientific Instruments NM30LA-MS The Nitrogen generator employs “Hollow Fibre Membrane” Technology to efficiently separate Nitrogen from other gases present in ambient air. An overview of this process can be seen below.
9.
Flow Diagram
2. 6.
7.
8.
4. 5. 3.
10.
11.
1.
Pneumatic Diagram
Air is drawn into the system by the Compressor (1) and passed through the Cooling Coil (2) and the AFD – 3000 Micro Mist Filter (3) into the Membrane (5). After the Membrane (5) the gas is passed through a Reverse Acting Carbon Filter (6) to remove any remaining impurities and via the Non-Return Valve (Check Valve) (7) into the Receiver (10). The stored gas is regulated (11) to give the required output pressure and flow. Receiver pressure is measured (9) to allow the generator to un-load (4) and shut-down should demand cease. A Safety Relief Valve (8) is fitted to protect the system against over pressure. Gas is delivered out of the machine via the Outlet Port (12). Any moisture collected by the AFD – 3000 Micro Mist Filter (3) is expelled via the Drain Port (13).
12
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Commissioning
Operator Training The NM30LA-MS Laboratory Gas Generator is designed specifically to minimize operator involvement. Given that the generator is installed as described in earlier sections and is serviced in accordance with the following maintenance recommendations then it should simply be a matter of turning the generator on. The generator will automatically produce the factory default flow and pressure noted below. NITROGEN OUTLET 30 litres per min (ATP) 100 PSI.G (680kPa)
Should the demand for gas be less than the rated flow then the generator will automatically regulate the internal pressure and cycle of the compressors as required. This should be all the input your generator requires from you, the only additional tasks required are, If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact your local representative or the factory as soon as possible. Please ensure that both the drain ports at the rear of the generator are led to a suitable connection or container. It should be noted that the generator will liberate considerable amounts of water from these ports (approx. 1-2 litres weekly). If a container is used it should be emptied at regular intervals. Ensure that the generator is serviced in accordance with the following maintenance recommendations.
13
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Maintenance Schedule
Routine Maintenance Servicing and/or repair of the Generator should only be undertaken by a TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated condition. SAFELY ISOLATED CONDITION Definition: The unit is in a Safely Isolated Condition when it is disconnected from its application, fully de-pressurised and isolated from the Electrical Supply. Directions for isolating the generator are shown below. Isolating the Generator: a)Switch off the unit. b)Disconnect the unit from the mains supply. c)Ensure the internal pressure gauges (page 12) reads zero. (If gauge does not fall to zero, loosen outlet fitting slightly to allow trapped gas to escape). d)Disconnect from the application. Maintenance Schedule
SERVICE INTERVAL 1 YEAR EVERY 6 MONTHS ANNUAL SERVICE KIT COMPRESSOR (THE LESSER) UNIT 6000 HRS OR 18MONTHS *ALTERNATIVE TO COMPRESSOR ASSY. COMPONENT AFD3000 FILTER ELEMENT (AFD) x 2 RAC FILTER ELEMENT x 1 INLET FILTER (x4) ALL PARTS LISTED ABOVE 2750 COMPRESSOR 2688 COMPRESSOR COMPRESSOR TRAY ASSEMBLY (x1) 2750 COMPRESSOR RE-FIT KIT (X1) 2688 COMPRESSOR RE-FIT KIT (X1) PART NO. 02-4335 00-4425 02-4187 08-4435 08-1425 08-1420 08-4649 06-5529 06-5542
on Request
Note: *Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 times, this is a more cost effective solution, however a degree of technical expertise is required and can be time consuming. Service kits are available for all routine maintenance. Please contact Peak Scientific for further details.
14
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Maintenance
RAC Filter
The Element is as shown, and should be changed at 12-month intervals. The Power supply to the Generator MUST be turned OFF and the Generator MUST be depressurised prior to attempting to remove ANY filter bowl. Failure to do this may cause injury.
RAC FILTER Element (00-4425)
The filter housing is removed by un-screwing in a counter clockwise direction
AFD3000 Filter
AFD 3000 Element (02-4335)
The filter housing is removed by pulling the black clip and turning the Bowl ¼ Turn either direction
15
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Maintenance
Compressor Intake Filter The Element is as shown, and should be changed at 6-month intervals. The Power supply to the Generator MUST be turned OFF prior to attempting to remove ANY filter. Failure to do this may cause injury.
Compressor 2750 Intake filter
Filter Element PAD 02-4187
Compressor 2688 Intake filter
16
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Pneumatic
Pneumatic Diagram
17
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Maintenance log
Maintenance Log
Model- NM30LA
Serial number
Work Done
Remarks
Date
Name
18
UM-NM30LA-MS Gas Generator
UM –NM30LA-MS
Notes
19