Genius 3010/13/20/22
User Manual
Copyright © Peak Scientific Instruments Ltd 2018 — Genius3010-3013-3020-3022 User Manual Rev 2 RSID 120 PN UM-0008 EN —01/02/2018
Contents
Change History
3
How to use this Manual
3
Introduction 4
Warranties and Liabilities
5
Safety Notices
6
Symbols 6
Safety Notice to Users
6
Declaration of Conformity
7
Environmental Declaration
8
Technical Specification
9
Unpacking 13
Fittings Kit Contents
14
Installation 15
Generator Environment
15
3010/3013 General Dimensions
16
3010/3013 Rear Connections
17
3020/3022 General Dimensions
18
3020/3022 Rear Connections
19
Electrical Connection
20
Start-Up Sequence
21
Normal Operation
22
On Demand Gas
22
Generator Cycling
22
Unusual Operation
22
Connecting to the application
23
Tubing Lengths
23
Service Requirements
24
Service Schedule
24
Peak Protected
26
Cleaning 27
Troubleshooting 28
Page 2
Change History
Rev
Comment
Name
Date
2
3010 Tech Spec Update
Liam Couttie
01/02/2018
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The Genius 3010, 3013, 3020 & 3022 have been developed to provide a supply of
nitrogen gas for your LC/MS instrument. We’ve built them with internal air compressors
as standard, removing the reliance upon an in-house air supply.
These model provide a source of nitrogen gas with other features including:
• Quiet in operation – Insulated compressor compartment for minimum disruption
• Cool – Minimal heat build-up in the lab
• Intelligent – High duty service indicator
• Mobile – Flexibility to position the generator where required
• Simple installation – Generator designed as a plug and play system
• Economical – More cost effective than any other gas supply method
• Convenient – Gas on demand, no hazards, no need to worry about running out of
gas
The Genius 3010, 3013, 3020 & 3022 are based on proven technology, and selectively
remove oxygen, moisture and other gases to leave clean, dry, phthalate free nitrogen.
Internal air compressors make these units independent from in-house air supplies and
fitted castors allow the user to easily position the units in the lab.
To ensure these Generator models meet our high expectations with regards to reliability
and performance, we have tested them extensively at our manufacturing plant and with
end users around the world to ensure reliability and longevity of the system.
Page 4
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods
shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free
from defects in material and workmanship for the lesser of a period of twelve
months from the date of delivery or thirteen months from the date of dispatch
from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person
dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common
law are hereby expressly excluded save to the extent they may not be lawfully
excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of
this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of
clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality
of the foregoing that any defect is not the direct or indirect result of lack of repair
and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or
incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant
packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub-
clause 2 hereof within a reasonable time of it becoming apparent and subject always
to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or
repair the defective Goods or refund a proportionate part of the Price. The Company
shall have no further liability to the Customer (save as mentioned in sub-clause 6
hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim
for death or personal injury to any person in so far as such is attributable to the
negligence or breach of duty of the Company or any failure by the Company to
comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect
of any claim by the Customer for costs, damages, loss or expenses (whether direct,
indirect, consequential or otherwise) or indemnity in any respect howsoever arising
including, but not by way of limitation, liability arising in negligence (other than
pursuant to clause 6 above) that may be suffered by the Customer or any third
party.
Page 5
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING
Page 6
Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment:
Nitrogen Gas Generator
Models:
Genius 3010, 3013, 3020 & 3022
To which this declaration relates, is in conformity with the applicable EC directives,
harmonized standards, and other normative requirements.
• Low Voltage Directive 2015/35/EU
EN 61010-1: 2010 Electrical Equipment for measurement, control and laboratory use.
• Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for measurement, control and laboratory use.
• FCC47 CFR Part 15 Class B
Unintentional radiators; Conducted and Radiated emissions limits.
All evaluation, testing and certification issued by:
Nemko Canada Inc.
303 River Road
Ottawa
Ontario
Canada
K1V 1H2
Signed: Name: Chris Pugh
Date:
Page 7
03/02/2017
Position:
Chief Technology Officer
Environmental Declaration
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Declare that:
Equipment:
Nitrogen Gas Generator
Models:
Genius 3010, 3013, 3020 & 3022
Is fully compliant with the following Directives
• 2012/19/EU
WEEE (Waste of Electrical and Electronic Equipment)
• 2011/65/EU
RoHS 2 (Restriction of Hazardous Substance)
Peak Scientific Instruments Ltd fully complies with its obligations towards the European
WEEE (Waste of Electrical and Electronic Equipment) Directive 2012/19/EU. These
obligations are being met within the B2B compliance group.
Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to
ensure that our products comply with the principles and requirements of the European
recast RoHS (Restriction of Hazardous Substances) Directive 2011/65/EU. Similar
directives in the United States and China, for example, have also been captured within
this program.
Where a specific certificate of compliance is required, this can be requested, on
a product serial number basis, directly from Peak Scientific Instruments Ltd, by
contacting us through our website on www.peakscientific.com
Signed: Name: Chris Pugh
Date:
Page 8
10/06/2015
Position:
Engineering & Operations Director
Technical Specification
Genius 3010
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
35°C (95°F)
2000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-Up Time
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlets
Pressure Gauges
6.9 bar (100 psi)
64 L/min (2.26 cfm)
30 mins
<0.01µm
NONE
NONE
1 x ¼” BSPP
1 x ¼” BSPP
1
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
230 VAC ± 10%
50/60 Hz
12 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 9
60 x 85 x 132.2 (23.6 x 33.5 x 52.0)
168.5 kg (371 lb)
249 kg (548 lb)
Technical Specification
Genius 3013
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
35°C (95°F)
2000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Chip Cube Mode
Start-Up Time
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlets
Pressure Gauges
6.9 bar (100 psi)
64 L/min (2.26 cfm)
increased O2 content to 10%
30 mins
<0.01µm
NONE
NONE
1 x ¼” BSPP
1 x ¼” BSPP
1
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
230 VAC ± 10%
50/60 Hz
12 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 10
60 x 85 x 132.2 (23.6 x 33.5 x 52.0)
189 kg (417lb)
265.5 kg (585lb)
Technical Specification
Genius 3020
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
35°C (95°F)
2000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-Up Time
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlets
Pressure Gauges
2 x independant outlet of 6.9 bar (100 psi)
2 x independant outlet of 32 L/min (1.13 cfm)
30 mins
<0.01µm
NONE
NONE
2 x ¼” BSPP
1 x ¼” BSPP
2
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
230 VAC ± 10%
50/60 Hz
12 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 11
60 x 85 x 132.2 (23.6 x 33.5 x 52.0)
189 kg (417lb)
265.5 kg (585lb)
Technical Specification
Genius 3022
Environment
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
35°C (95°F)
2000 m
80% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Generator Outlets
Maximum Gas Output Pressure
Maximum Outlet Flow Rate
Start-Up Time
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlets
Pressure Gauges
2 x independant outlet of 8.1 bar (116 psi)
2 x independant outlet of 32 L/min (1.13 cfm)
30 mins
<0.01µm
NONE
NONE
2 x ¼” BSPP
1 x ¼” BSPP
2
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
230 VAC ± 10%
50/60 Hz
12 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 12
60 x 85 x 132.2 (23.6 x 33.5 x 52.0)
189 kg (417lb)
265.5 kg (585lb)
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator to
the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful
not to discard these with the packaging.
Page 13
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. 6mm Teflon Tubing x 3m*
2. ¼” Teflon Tubing x 3m*
3. 6mm PE Tubing x 3m
4. ¼” Compression Fitting x 1**
5. 6mm Push Fit Fitting
x 2**
6. Flow Control Silencer x 1***
7. UK Mains Power Cable x 1
8. EU Mains Power Cable x 1
9. US Mains Power Cable 230v
x1
10. 8mm Hex Key x 1
*
x6m in Genius 3020/22
**
x3 in Genius 3020/22
***
x2 in Genius 3020/22
All of the generators output ports are located on the output panel at the rear of the
unit.
Page 14
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation
of the unit within such a pocket could adversely affect its performance. Consideration
should also be given to the air flow around the unit. It is recommended that an air gap
of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please
refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
35 °C (95 °F)
Page 15
Generator Overview
General Dimensions
600 mm / 23.6”
850 mm / 33.5”
1322 mm / 52"
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING
Page 16
3010/3013 Rear Connections
6mm Push-Fit Fitting
¼" Compression Fitting
Mains Cable
Page 17
3020/3022 Rear Connections
6mm Push-Fit Fitting
¼" Compression Fitting
Mains Cable
Page 18
Electrical Connection
Connect the generator to an appropriate 110 or 230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.
WARNING
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements for 230V generators is 195V - 253V 50/60Hz. For 195V - 219V a
06-3210 Dual Tap Transformer is required.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING
Page 19
Start-Up Sequence
CAUTION
Before the Generator is connected to the application, the Generator
should be operated in isolation (i.e. not connected to the application)
for thirty minutes. This is to ensure any impurities present are purged
from the system. Failure to do this may harm the application.
Before re-connecting the Generator to the mains and switching it on for the isolation
run, it is necessary to fit the silencers (supplied in the fittings kit) to the outlet ports of
the output panel as shown below.
Once this is done, the Generator can be re-connected to the mains and switched on.
Pressure will start to build in the internal storage tanks which can be monitored by
watching the output pressure gauges on the front panel. This will climb to the factory
set pressure as noted in the specifications.
Once this pressure is reached, the compressors will continue to run until the internal
tank upper pressure limit is reached and the compressor has run for a period of at least
two and a half minutes.
The compressors will then rest until the internal tank lower pressure limits is reached.
Once this limit is reached the compressors will switch back on again. This compressor
cycling is normal and will continue throughout the operation of the Generator.
When the system has been operated for a period of 30 minutes, all the internal pipework and storage tanks will have been purged with nitrogen.
Once purged, and the tubes can be connected at the rear of the unit. The generator
should be allowed to run for 1 hour to allow the gas to reach purity.
Page 20
Normal Operation
The Genius 3010/3013/3020 & 3022 Gas Generators are designed specifically to
minimize operator involvement. Given that the systems are installed as described
in earlier sections and is serviced in accordance with the specified maintenance
recommendations (see Service Requirements), then it should simply be a matter of
turning the Generator on when it is required.
The Generator will automatically produce the factory set flow and pressure as detailed
in the Technical Specifications.
On Demand Gas
The generator will produce nitrogen on demand. If the application is operating
and requires a gas flow, the system it is connected to will supply this to suit the
requirements of the application. If the application requirement for gas stops, the system
will also stop, once it has reached its upper set limit in the internal storage tanks. If the
demand from the application starts again, the system will detect the demand for gas
and will automatically start again to suit the demand.
Generator Cycling
The Generator is designed for the internal compressors to cycle. This cycling reduces
the duty (run time) on the compressors. The rate at which they cycle will be dependent
on the gas required to satisfy the demand of the application. If the application demands
the maximum gas flow of the Generator, the compressor duty will be higher, (the rest
period in the compressor cycle will be shorter). If the application demand is lower than
the maximum gas flow, then the duty on the compressors will decrease, (the rest period
in the compressor cycle will be longer). If the Generator is installed in an extreme
environment or is subjected to low supply voltage or high altitude the compressors may
undergo periods where they do not cycle.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.
Page 21
Connecting to the application
Once the initial purge run of 30 minutes has completed, and the generator has been
running for 1 hour, it is ready to be connected to the application(s).
The pressure in the internal storage tanks must be allowed to dissipate
before connecting the generator to the application(s)
CAUTION
Attach the appropriate fittings to the outlets of the generator. Using the tubing
supplied, connect the outlets of the generator to the inlets on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.
CAUTION
Once the tubing is connected to the application, please ensure that it
is thoroughly checked for being leak-tight. Even the slightest leak in
the gas supply between the generator and the application can lead to
a reduction in efficiency.
Tubing Lengths
CAUTION
The diameter of the tubing which will be connected to the gas outlet is
important and is determined by the length of tubing required. Failure
to follow these recommendations could lead to pressure between
generator and application.
< 10 meters:
Use 6/4 (6mm O/D, 4mm I/D) P.T.F.E. tubing.
> 10 - 40 meters: Use 10/8 (10mm O/D, 8mm I/D). Tubing and fittings not
supplied in the fittings kit.
> 40 metres:
Please contact Peak Scientific with the relevant distance andwe
will calculate the flow resistance and the tubing size required.
A combination of 6/4 and 10/8 tubing may be used to ensure that there is no large
diameter tubing within the lab (i.e. for the first 20 meters from the Generator use 10/8
and the final 10 meters to the application use 6/4 tubing). Keep the connections and
bends to a minimum.
The imperial equivalents are: 6/4 = 1/4” O/D, 3/16” I/D.
Page 22
10/8 = 3/8” O/D, 5/16” I/D.
Service Requirements
Service Schedule
Purchase Interval
Component
Visit
Genius 3010/13 Annual Maintenance Kit
(Pre March '14 Manufactured Generators)
Genius 3010/13 Annual Maintenance Kit
(Post March '14 Manufactured Generators)
12 Months
Genius 3020/22 Annual Maintenance Kit
(Pre March '14 Manufactured Generators)
www.peakscientific.com/ordering
Genius 3020/22 Annual Maintenance Kit
(Post March '14 Manufactured Generators)
Compressor
Assembly
Compressor Assembly
(Pre March '14 Manufactured Generators)
Compressor Assembly
(Post March '14 Manufactured Generators)
*
Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 times, this is a
more cost effective solution, however a degree of technical expertise is required and can be time consuming
Please contact your service provider for more information.
Service Indication
The Generators have the following Service Indication Stages:-
Stage 1
Once either compressor requires a service the LED indicator (yellow) on the front of the
Generator will illuminate.
This is to make the user aware that a service of the Generator is due and should be planned
at the earliest convenience. The Generator will continue to operate as normal with the LED
on.
Stage 2
If the service is not completed the Generator will continue to run. After 2 weeks, the service
LED indicator (yellow) will start to flash.
This is to make the user aware that the service of the Generator is now overdue and must be
completed immediately to ensure the continuous trouble free operation of the Generator.
Service Indication Reset
Once the service has been completed the Service Indication LED can be reset in the
main control PLC. This will be performed by the Peak Service Engineer or trained service
representative that completes the service operation.
High Duty Indication
The Genius 3010/3013/3020 & 3022 have a ‘HIGH DUTY’ indicator on the front
panel. This indicator monitors the running condition of the internal compressors and
Page 23
illuminates when they have been operating continually for a period of 8 hours.
There are a number of extreme conditions that can affect the duty cycling of the
compressors inside the generator. These extremes include high ambient temperatures,
low supply voltages, a high flow requirement and locations of high altitude. Any one
of these or a combination of a number of these extremes can be evident in a customer
site.
While the generator is designed to operate fully in these conditions it should be noted
that the effect of these extremes can force the compressors to run continually. The
compressors are fully capable of running continually with no detrimental effect.
High Duty Indication Reset
If the compressors have been operating continually for a period of 8 hours the indicator
light will be illuminated. Once the compressors return to a cycling mode the indicator
light will automatically switch off. There is no manual intervention required.
Indication of Fault
In most installations the ‘HIGH DUTY’ indicator will never be illuminated. If your Genius
3010/3013/3020 & 3022 Generator has been operating for a significant period of
time, then suddenly develops a ‘HIGH DUTY’ indication with no apparent change in
environmental conditions or flow requirements, it may be indicative of a problem with
the system such as an external leak or imminent compressor failure.
If this is the case you should contact Peak Scientific or your service provider.
It should be noted that the Generator will continue to operate so long as it can maintain
pressure during this indication.
Page 24
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit
your lab to install and setup your
generator
IQ/OQ
Certified assurance for applications
requiring documented qualification
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance
Spares
Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime
Premium Protected
Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance
Technical Support Hotline
Around the clock support by phone
or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 25
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the generator.
CAUTION
Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Page 26
Troubleshooting
Problem
Possible Solution
• Ensure power cord is plugged into the
Generator and that the power socket is
turned on.
The Generator will not switch on and the
power switch does not illuminate.
The Generator will not switch on but the
power switch is illuminated.
• Check the fuse in the power cord plug.
• Contact your service provider.
• Disconnect power cord from the rear of
the Generator. Open the left hand panel
are check that both circuit breakers are
turned on (switch in the up position).
Reconnect power cord.
• Contact your service provider.
Compressors are running but pressure is not
building.
The mass spec is reporting low pressure.
Yellow “SERVICE” LED on front panel is on
constantly.
*Yellow “SERVICE” LED on front panel is
flashing.
Yellow “HIGH DUTY” LED on front panel is on
constantly.
• Check pressure gauges are showing
normal pressure.
• Contact your service provider.
• Contact your service provider.
• Annual service is due. Contact your service
provider.
• Refer to Service Indication section of this
manual for further information.
• Annual service is overdue. Contact your
service provider urgently.
• Refer to Service Indication section of this
manual for further information.
• A compressor(s) has not cycled for a
period of 8 hours. Refer to the High Duty
Indication section of this manual.
• Ensure ambient temperatures are within
the specification.
• Ensure there are no leaks between the
Generator and the mass spec.
Page 27
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 28
Page 29
Page 30
Page 31
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2018 — Genius3010-3013-3020-3022 User Manual Rev 2 RSID 120 PN UM-0008 EN —01/02/2018