Installation Guide – SCTOCA Generator
FITTINGS KIT
x1
x2
Mains Cable – Euro
6mm Push-fit Fitting
x1
Mains Cable – UK
x1
x1
Mains Cable - US
CD containing User Manual
x 3m
6mm Teflon Tube
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
x 2m
6mm Nylon Tube
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 1 of 8
INSTALLATION
1.
2.
Unpack the generator from the shipping crate
and place on a flat surface. To remove the
transit brackets, firstly remove the lid.
3.
Using a Pozidrive # 2 screwdriver, remove
the 13 screws around the edge of the lid,
shown above.
4.
Remove the lid by lifting upwards.
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
Ensure to remove the earth lead on the left
side of the panel and the fan connector
leads on the right.
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 2 of 8
5.
6.
With the panel removed you now have access
to the transit brackets. These are located
under the compressor and are painted red. To
remove the transit bracket remove the 4
screws as indicated.
Once these screws have been removed
the transit bracket can be pulled out as
shown.
**NOTE**
Do not discard the transit brackets or screws as these will be needed again if the
unit is to be transported from its current location.
7.
8.
Re-attach the lid,
ensuring you connect
the earth lead, and fan
connector lead.
9.
From the fittings kit, select
one of the 6mm push-fit
elbow fitting and connect it
to the port marked ‘DRAIN’
on the rear of the unit
using a 16mm or 5/8”
spanner.
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
Connect the 2m Nylon tube to the
drain line, ensure the tube is
pushed fully in and gripped securely
by the fitting.
Fit the other end of the drain line to
a suitable drain connection or
container. The container must NOT
have an airtight seal as water and
air will be expelled periodically
under a slight pressure.
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 3 of 8
10.
11.
Select the appropriate mains cable from the fittings kit
and plug the IEC 60320/C19 socket into the mains
input at the rear of the generator.
12.
Plug the mains cable into an
appropriate 230VAC 50/60Hz
single phase power supply
13.
Switch the generator power ON at the
switch on the front panel.
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
During the initial 30 minutes the generator
should reach a pressure of 80psi, and a
temperature of 400°C.
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 4 of 8
14.
15.
If at any point during the 30 mins there is an
apparent problem reaching pressure or
temperature, contact Peak Scientific.
With pressure and temperature successfully
achieved, allow the generator to vent to
atmosphere for 16 hours, before connecting to
any equipment.
16.
17.
After 16 hours the system is now
operational and should be switched OFF.
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
To connect to the instrument Select the other
6mm push fit elbow fitting from the fittings kit
and connect it to the port marked ‘OUTPUT’
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 5 of 8
18.
19.
Select the appropriate size Teflon Tube**
and connect to the elbow port.
20.
Connect the other end of the Teflon tube to
the Instruments input.
21.
Switch the machine back ON
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
The temperature, and pressure will now
quickly reach ‘400°C’ and ‘80psi’ respectively
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 6 of 8
22.
CONGRATULATIONS
Your PEAK SCIENTIFIC gas generator is now fully installed, operational and ready to
supply gas on demand to your instrument.
General Notes
*
The generator will operate on voltages of 230VAC +/- 10%. This means it will
operate between 207V and 253V. If the generator is operated on a supply voltage of
less than 220V the internal compressors will work harder and the service life of the
compressors will be reduced. It is ok to operate the generator on a mains voltage
between 207V and 220V for a short period of time, however we would recommend
fitting the listed transformer at your earliest convenience. Mains voltages between
220V and 253V no transformer is required.
**
The 3m length of Teflon tube supplied in the fittings kit is to allow the
generator to be installed adjacent to the instrument it is supplying. If the generator is
to be installed further away from the instrument then attention needs to be paid to
pressure drops that could occur. The maximum distance away from the instrument
the generator can be placed still using 6mm (4mm internal diameter) or ¼” (3/16”
internal diameter) tubing is 10m. For further information relating to tubing lengths
please refer to the user manual.
For all other technical specifications, operating instructions, service
requirements, contact details and trouble shooting, please refer to
the user manual contained on the CD supplied in the fittings kit.
Please keep this for future reference.
23.
It is very important to register your generator with PEAK SCIENTIFIC. This will initiate your
warranty entitlement. Please use the form on the next page to register your generator. You
will need the generators serial number which can be found on the serial label on the rear of
the generator.
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 7 of 8
Product Warranty Registration
IMPORTANT
To register your Generator and guarantee Warranty entitlement, send the completed form to
Peak Scientific by:
Email to
Online
Phone to
Fax to
marketing@peakscientific.com
www.peakscientific.com/service-and-support/warranty_registration
+44 (0)141 530 4185
+44 (0)141 812 8200
Product Warranty Registration
Company:
Contact Name:
Address:
Email Address:
Generator Serial Number:
City/Town:
Postcode:
Country:
Model Type: SCTOCA
Telephone:
Installation Date (DD/MM/YYYY):
Contact Information
Technical Help Desk
USA
India
South Africa
tel: +44 (0)141 530 4180
tel: + 1 866 732 5427
tel: +91 (0)40 2780 0663
tel: +27 (0)83 296 4516
email: support@peakscientific.com
email: usasupport@peakscientific.com
email: indiasupport@peakscientific.com
email: peakafrica@peakscientific.com
Installation and Maintenance Visits:
tel: +44 (0)141 530 4190
Service Contract Quotation and Renewals:
tel: +44 (0)141 530 4185
email: maintenance@peakscientific.com
Spare Part Enquiries:
tel:+44 (0)141 530 4181
email: spares@peakscientific.com
Fountain Crescent
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Part No. IG-SCTOCA
Rev.
01
Page
Page 8 of 8