SCTOCA User Manual
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Contents
Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental Declaration Technical Specification Environment Generator Outlets Electrical Requirements General Unpacking Installation Generator Environment Removal of Transit Brackets Unit Controls & Rear Connections Fittings Kit Drain Connection Electrical Connection Start Up Procedure Pressure/Flow Adjustment Unusual Operation IMPORT ANT DOCUMENTS Normal Operation Generator Cycling Unusual Operation Principal of Operation 4 4 5 6 7 7 7 8 9 10 10 10 10 10 11 12 12 13 14 15 16 16 17 17 17 18 19 19 19 20
Page 2
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Service Requirements Service Schedule Compressor Service Service Plans Cleaning
21 21 21 22 23
Page 3
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Change History
Rev.
1 2 3 4
Comment
Filter Element Part No. Change
Name
Liam Couttie
Date
13/06/13
How to use this Manual
This manual is intended for end users and has been written so that it can either be read as a step by step guide to installation and usage or as a reference document where you can skip to the relevant information. Users of a hard copy version can refer to the contents page to find the relevant information. Users of the soft copy version can use the hyperlinks from the contents page as well as the hyperlinks between sections. Please review each of the following sections carefully. Thank you for selecting Peak Scientific to meet your Gas Generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or the Peak Partner from which you purchased your Generator.
Page 4
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Introduction
The SCTOCA 500cc generator has been developed to cater for systems which require a pure gas supply free of hydrocarbons with no carbon dioxide. This model provides a source of gas with other features including:
• • • • •
Quiet in operation – noise level of 57 db(A) Anti-vibration – maximum reduction of vibration Minimum user intervention Improved drainage – reduction of moisture carry over and thus increased reliability Robust control system – improves safety and reliability of units.
With the SCTOCA based on proven technology, it removes moisture, hydrocarbons and CO2. Its own internal air compressor makes this unit independent from in–house air. To ensure this Generator model meets our high expectations with regards to reliability and performance, we have tested this new model extensively at our manufacturing plant and with end users around the world to ensure reliability and longevity of the system.
Page 5
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Warranties and Liabilities
1. The Company warrants that it has title to the Goods. 2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory. 3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act 1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause. 4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must. 1. Reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts 2. Allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company. 5. Subject to the Company being notified of any defect as is referred to in sub-clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof). 6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable
in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
Page 6
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Safety Notices
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the Generator:
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met. A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the Generator or the Application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met. Caution, risk of electric shock. Ensure power to the Generator has been removed before proceeding.
Table 1 - Safety Symbols
Safety Notice to Users
These instructions must be read thoroughly and understood before installation and operation of your Peak SCTOCA Generator. Use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations.
Page 7
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Declaration of Conformity
We of declare that: Peak Scientific Instruments Ltd. Fountain Crescent, Inchinnan. Renfrewshire PA4 9RE TOC Gas Generator SCTOCA
Equipment Model
To which this declaration relates, is in conformity with the applicable EC Directives, harmonized standards, and other normative requirements.
•
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use.
Low Voltage Directive 2006/95/EC
•
Electromagnetic Compatibility Directive 2004/108/EC
EN 61326-1: 2006 Electrical Equipment for Measurement, Control and Laboratory Use – EMC Requirements.
All evaluation, testing and certification issued by: York EMC Services Ltd. Donibristle Industrial Park Dunfermline Fife KY11 9HZ
Signed By: Name: Position: Done at: Date: Chris Pugh Engineering Director Peak Scientific Instruments Ltd, Inchinnan, Scotland. 20 of November 2011
th
Page 8
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Environmental Declaration
We of declare that: Peak Scientific Instruments Ltd. Fountain Crescent, Inchinnan. Renfrewshire PA4 9RE
Equipment Model
Generator SCTOCA
Is fully compliant with the following Directives: 2002/96/EC WEEE (Waste of Electrical and Electronic Equipment) 2002/95/EC RoHS (Restriction of Hazardous Substances) Peak Scientific Instruments Ltd fully complies with its obligations towards the European WEEE (Waste of Electrical and Electronic Equipment) Directive 2002/96/EC. These obligations are being met within the B2B compliance group. Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to ensure that our products comply with the principles and requirements of the European RoHS (Restriction of Hazardous Substances) Directive 2002/95/EC. Similar directives in the United States and China, for example, have also been captured within this program. Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website on www.peakscientific.com Signed By: Name: Position: Done at: Date: Chris Pugh Engineering Director Peak Scientific Instruments Ltd, Inchinnan, Scotland. 6 December 2010
th
Page 9
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Technical Specification
Environment
Minimum operating ambient temperature Maximum operating ambient temperature Maximum relative humidity Maximum altitude Minimum storage temperature* Maximum storage temperature* 5°C (41°F) 30°C (86°F) 70% 2000 Meters -20° 60°
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.
Generator Outlets
Maximum Output Flow Purge Flow Rate Maximum Output Pressure Pressure Dew Point Particles Gas Outlets Drain Outlets Pressure Gauges Phthalates Suspended Liquids 500CC/Min 2600CC/Min 80psig -70°C (94°F) <0.01 µm 1 x ¼” BSP 1 x ¼” BSP 1 NONE NONE
Electrical Requirements
@230V AC+-10% (50/60Hz) Fuse Electrical Connection Noise Level 4 Amps 6.3 Amps Single Phase Power Cord 57 dBA @ 1m
General
Dimensions in cm (inches) W x D x H Weight Shipping weight 361.5 x 570 x 714.5 (14.2 x 22.4 x 28.1) 55kg (121lbs) 80kg (176lbs)
Page 10
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ label for signs of rough handling prior to un-packing –
Any damage should be reported immediately to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased. Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the Generator onto the floor. Please save the product packaging for storage or future shipment of the Generator. Note: Included with the Generator is a “Fittings Kit” containing mains power leads for UK, EU and US also all the required fittings. Be careful not to discard these with the packaging.
Page 11
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to the instrument it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe. Performance of the generator (like all sophisticated equipment) is affected by ambient conditions. Note should also be taken to the proximity of Air Conditioning outlets. These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the unit within such a pocket could adversely affect its performance. Consideration should also be given to the air flow maintained down both sides, at the rear and across the top of the unit. Please refer to the drawing below for the general dimensions of the unit.
General Dimensions
361.5mm
570mm
Page 12
714.5mm 2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Removal of Transit Brackets
The transit bracket must be removed prior to switching the unit on. Failure to do so will result in damage to the equipment. This will void the warranty on the generator and will result in a chargeable repair.
1.
Use a Phillips screw driver to remove the side cover from the cabinet ensuring that the earth and fan cables are disconnected.
2. Remove the four screws with a #3 Pozidrive Screwdriver. 3. Slide the transit bracket out from under the compressor by pulling it up and towards you. 4. Retain the transit brackets as these must be refitted if the generator is to be transported again.
During and after operation of the generator, the internal components of the generator will be extremely hot for several hours. Access should be limited to suitably trained personnel. Allow 6 hours for the components to cool down.
Page 13
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Unit Controls & Rear Connections
Temperature Gauge
Pressure Gauge
Power Switch
Fan- Do not block
Fan- Do not block
Power Inlet
Gas & Drain Outlets
Page 14
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Fittings Kit
Supplied in the Fittings Kit are all the fittings required to connect the SCTOCA to the application. The contents of the Fittings Kit are as follows. Description User manual CD Cable UK Cable Euro Cable US ¼” 6mm Push fit elbow 6mm Teflon tubing 6mm Nylon tubing Quantity 1 1 1 1 2 3m 3m
All of the Generator output ports are located at the rear of the Generator.
Page 15
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Drain Connection
Fit the 6mm push fit fitting to the drain port located at the rear of the generator and connect this to a suitable drain connection or container. It should be noted that the generator can expel a considerable amount of water from this (dependant on ambient humidity).
If a container is used it must be emptied at regular intervals. The container must NOT have an airtight seal as water and air are expelled at pressure.
Electrical Connection
Connect the generator to a 230 volt single-phase supply using the power cord provided. If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can be fitted by a qualified electrician. This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED. Before connecting to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):LIVE (L):NEUTRAL (N):-
Green & Yellow Brown Blue
or or or
Green Black White
Our electrical requirements are 230VAC nominal ±10%. This means that the generator can accommodate transients between 207VAC and 253VAC. However, running continuously at voltages less than 220VAC is not recommended and extended periods at these extremes can have a detrimental effect on the operation and life of the Generator. If when the Generator is switched on, the input voltage is 219V or less, then we would recommend fitting a transformer. This can be ordered directly from Peak Scientific with the order number below. Product Description Dual Tap Transformer 200V – 230 Part Number 06-3200
Page 16
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Start Up Procedure
For a more detailed procedure, please refer to the SCTOCA installation guide supplied with the generator. Make sure the mains power supply is switched on. Switch on power switch located at front of generator. Confirm LED/Temperature display is illuminated. Almost immediately, the compressor should be heard running. During the initial 90-120 minutes the required temperature of 400°C should be reached. Allow a total 24 hours running, the generator can be switched off before connecting to the application.
Pressure/Flow Adjustment
The system is configured in the factory to give standard outlet pressure and flow rate (see technical specification). These settings should never require adjustment.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact your authorised service provider or Peak Scientific as soon as possible.
Page 17
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
IMPORTANT DOCUMENTS Warranty Entitlement
To register your generator for your warranty entitlement, send the completed form to Peak Scientific by: • • • • Email Online Phone Fax
warranty@peakscientific.com http:/ /www.peakscientific.com/service-and-support/warranty_registration
+44 (0)141 530 4185 +44 (0)141 812 8200
PRODUCT WARRANTY REGISTRATION
COMPANY: ADDRESS: EMAIL ADDRESS: CONTACT NAME:
CITY/TOWN: POSTCODE: COUNTRY: TELEPHONE:
GENERATOR SERIAL NUMBER:
MODEL TYPE: INSTALLATION DATE (DD/MM/YYYY):
Important Please Note:
You have 1 month to register your Peak Scientific product from the date of shipment. If you wish to defer installation of your generator you must notify Peak Scientific within 1 month of the shipment date. This can be done by emailing warranty@peakscientific.com Once registered the warranty will be honoured for a period of 12 months after the installation date. For any generators that remain unregistered the warranty will begin from date of shipment. Thank you on behalf of Peak Scientific.
Page 18
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Normal Operation
The SCTOCA Gas Generator is designed specifically to minimize operator involvement. Given that the system is suitably matched to the application, installed as described in earlier sections, and is serviced in accordance with the specified maintenance recommendations; it should act as a long trouble free source of air to meet your needs. In normal operation, the ON/OFF switch will be illuminated. The pressure gauge will show 80psi and the temperature display will show approximately 400°C. The temperature may fluctuate up or down by up to 4°C. The generator will automatically produce the factory set flow and pressure as detailed in the Technical Specifications. It should not be altered at any time.
Generator Cycling
The generator is designed for the internal compressor to cycle. This cycling reduces the duty (run time) on the compressor. The rate at which it cycles will be dependent on the gas required to satisfy demand. If the application demands the maximum gas flow of the generator, the compressor duty will be higher, (the rest period in the compressor cycle will be shorter). Please note that even when no gas is consumed from the generator, it will continue to cycle such to ensure immediate readiness when demand resumes.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact your local representative or peak scientific as soon as possible. If the temperature display indicates that the system is not heating or the gauge shows that the pressure is not building, please contact your local representative or Peak Scientific as soon as possible.
Page 19
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Principal of Operation
The inlet air first passes through a pre-filtration process where it undergoes particulate removal (≤0.01 micron) and bulk moisture removal (by a coalescing element). A membrane dryer further reduces the moisture content before the filtered air is stored in a receiver to be consumed as required. The SCTOCA generator works on the basis of two fundamental processes to produce the output gas. The first process utilizes a ‘Zero Air’ catalytic combustion chamber. This works on the principle of catalytic oxidation where hydrocarbons from the incoming compressed air supply, from the internal receiver, are cracked to carbon dioxide and water. The hydrocarbon level in the form of methane is reduced to <0.1ppm. For this process to work the catalyst requires to be heated to approximately 400°C. Because water and also carbon dioxide are created in the catalytic process there is a necessity to effectively remove them both. Moisture/ CO2 Removal:The second process utilizes a ‘Pressure Swing Adsorption’ (PSA) method to further treat the air. This is where contaminant gases and moisture can be selectively adsorbed from compressed air into a porous crystalline sieve material. The adsorption process is aided by the electrostatic interaction between the adsorbent sieve material and the gaseous adsorbate. The Peak Scientific Instruments Ltd. PSA dryer system utilizes the ‘Skarstrom’ process where there are two columns of adsorbent used alternatively and described as follows:Air is fed at pressure into one of the columns and the outlet air from this column is split into two paths. Most of the air is directed to the outlet of the generator for use by the consuming instrument. The remainder, 2.6lpm, is fed back down the ‘off line’ column to purge out the adsorbed impurities and subsequently regenerate the column. After a time period the columns swap over to ensure on going regeneration of the previously used column thus maintaining output gas quality. During this process CO2 is reduced to a level of <1ppm and water moisture is removed to a level of -70°C dew point (2.5ppm @ATP).
Page 20
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Service Requirements
Service Schedule
Service Interval Component Coalescing filter element Inlet filter element 12 months RAC Filter element Silencer Fixed orifice fitting Table 2: Service schedule Part No. 00-0032 02-4640 00-4425 02-6005 02-1272 Qty. 1 1 1 1 2
As an alternative to purchasing the 12 month service items individually an Annual Service Kit is available as one part number. This contains all the filters required for this Generator and offers a costs saving over buying the components separately.
Purchase Interval 12 months Component SCTOCA 500cc annual service kit Table 3: Annual Service Kit Part No. 08-8326 Qty. 1
Compressor Service
Compressor Service 12 months Component Compressor assembly **Compressor re-fit Part No. 08-8314 06-5542 Qty. 1 1
** Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 times, this is a more cost effective solution, however a degree of technical expertise is required and can be time consuming. Please contact your service provider for more information.
Page 21
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Service Plans
Peak Scientific offer two service plans. The Complete Service Plan, specifically designed for Generators operated in critical environments, also includes full breakdown cover, guaranteed response times and Generator upgrades if available. Our Standard Service Plan, covering the basic needs of our Generators, features special deals on spare parts and breakdowns. If you want to know more about our Service Plan options and how we ensure that your instrument can run with the maximum uptime and performance, please contact us at maintenance@peakscientific.com
Page 22
2013 © Peak Scientific – Rev 1 – 13/06/13
SCTOCA User Manual
Cleaning
Clean the outside of the Generator only using warm soapy water and a clean damp cloth. Ensure the cloth is thoroughly rung out to remove excess fluid prior to use. Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the Generator.
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the Generator.
Page 23
2013 © Peak Scientific – Rev 1 – 13/06/13
Peak Scientific UK Fountain Crescent Inchinnan Business Park Inchinnan PA4 9RE Scotland, UK Tel: +44 (0)141 812 8100 Fax: +44 (0)141 812 8200 Peak Scientific Germany Herriotstrasse 1 60528 Frankfurt
Peak Scientific North America 19 Sterling Road Suite #1 Billerica, MA 01862 USA Tel: +1 866 647 1649 Fax: +1 978 608 9503 Peak Scientific China Room 606 Building 1 Lane 2277 Zuchongzhi Road Pudong New Area Shanghai 201203, China Tel: +86 21 5079 1190 Fax: +86 21 5079 1191 Peak Scientific Brazil Peak Scientific Brasil Av. Quieroz Filho 1700 – torre Sunny-Conj. 602 São Paulo SP Brasil Tel: +1 866 647 1649
Peak Scientific India 202, Amsri Shamira Old Lancer Line Opp. St. Mary’s Degree College S.D. Road Secunderabad 500 025, India Tel: +91 40 2780 0663 Fax: +91 40 2780 0663 Peak Scientific Mexico Solon 352 Col. Los Morales Polanco 11530 Mexico, D.F. Mexico
Peak Scientific Africa PO Box 478 Somerset Mall 7137 South Africa Tel: +27 (0)2185 16542 Fax: +27 (0)8654 64473 Peak Scientific Australia PO Box 65 Belmont Victoria 3216 Australia
Germany
Tel: +49 (0)69 677 33 490 Fax: +49 (0)69 677 33 200 Peak Scientific Taiwan 4F.-6, No.736 Zhongzheng Rd. Zhonghe Dist. New Taipei City
Tel: Fax:
+1 866 647 1649 +1 978 608 9503
Tel:
+61 1300 965 352
Peak Scientific Japan K.K. 2-7-56, 2F Fuji Building 28 Kita Aoyama, Minato-Ku Tokyo, Japan 107-0061
Peak Scientific Singapore 3 Science Park Drive #03-14 The Franklin Science Park Drive 1 Singapore, 118223
Tel: +886 2 8226 2383 Fax: +886 2 8226 9499
Tel: +81 3-6864-0468
Tel: +65-6777 1966 Fax: +65-6777 1978
2013 © Peak Scientific – Rev 1 – 13/06/13