INSTALLATION QUALIFICATION
Instrument Origin and Identification
Instrument Series:
Genius
Model:
1024
Instrument Description:
Nitrogen Gas Generator
Manufacturer:
Peak Scientific Instruments Ltd
Fountain Crescent
Inchinnan Business Park
Inchinnan
Renfrew PA4 9RE
Scotland, UK
Service Contact:
Peak Scientific Instruments Ltd
Tel.
+44 (0)141-812-8100
Email support@peakscientific.com
Serial Number:
Condition when Received:
New
Used
Reconditioned
Date Received:
Installed Location:
Purchase Order Number:
Identification / Asset Number:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 1 of 9
INSTALLATION QUALIFICATION
Instrument Specification
Instrument Parameter
Specified Range
Acceptable
Output
Curtain Pressure
65 psi (4.48 bar)
Curtain Flow Rate
19 l/min (ATP)
Source Pressure
100 psi (6.89 bar)
Source Flow Rate
26 l/min (ATP)
Exhaust Pressure
60 psi (4.13 bar)
Exhaust Flow Rate
25 l/min (ATP)
System Purge Time
30 minutes
Environment and Siting
Has the instrument been adequately acclimatized since transport or storage?
(Requires 3 hours before operation)
Yes
No
Operating Parameter
Specified Range
Ambient Temperature
Acceptable
5-30 ºC (41-86 ºF)
Maximum Relative Humidity
80% Non-Condensing
Maximum Altitude
2000 m (6500 ft)
Supply Voltage & Frequency
230V ±10%, 50/60Hz
Refer to User Manual for acceptable range
Siting – In-House / Large Space
Minimum Clearance around Machine – 75mm
Siting – In-House / Confined Space
Air Conditioner/Extraction Fan
Mains Power Outlet
Within 2m Radius to Generator
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 2 of 9
Comments:
INSTALLATION QUALIFICATION
Instrument Delivery and Documentation
Unpack the instrument carefully. Are any items missing against the Packing List?
Yes
No
If yes, state missing items:
Is there any damage to the instrument?
Yes
No
Yes
No
If yes, describe damage:
Manufacturer informed:
Is all standard documentation included?
Present
Missing
User Manual (available via download form Peak Extranet)
Installation Guide
Warranty Registration Card
Installation
Installed By:
User
Authorized Engineer
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 3 of 9
INSTALLATION QUALIFICATION
Summary Report
Instrument:
Genius 1024
Serial Number:
Assessment of Complete Installation Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument has been
satisfactorily delivered and installed. The instrument has passed the Installation Qualification
procedure and may now be submitted for Operational Qualification.
IQ Completed By:
Date:
Deviations Approved By:
Date:
IQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 4 of 9
OPERATIONAL QUALIFICATION
Pre-Run Checks
Pre-Run Check
OK
Discrepancy
N/A
Compressor Transit Brackets Removed
Pressure gauges showing 0 psig
Gas Outlets Disconnected
Drain line connected to suitable outlet
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 5 of 9
Functional Tests
Functional Test
Acceptance Criteria
Apply Mains Power / Switch On
Internal Volt meter reads
between 220V – 253V
Allow to run for 10 minutes.
Pressure builds to 65, 100 and
60 psi on Curtain, Source and
Exhaust gauges respectively
(Ensure Side Panels are fitted)
Output pressures
Pass
Fail
Stable pressure shown on panel
pressure gauges
After 30 minutes 1. Perform Calibration on the Generator
Calibration
This instrument is supplied fully calibrated by the manufacturer using equipment traceable to
national standards. Calibration directly following installation is not usually required.
Calibration Step
Pass
Fail
Calibrate output flows
Serial No. of Calibrated Flow Meter -:
Calibration Due Date of Flow Meter -:
Calibrate output pressures
Serial No. of Calibrated Pressure Gauge -:
Calibration Due Date of Pressure Gauge -:
Comments:
2. Power-off / Isolate Generator from Mains Power / Allow pressure to drop to Zero
3. Connect Gas lines to instrument
4. Apply Mains Power / Switch On
5. Allow to run for 10 minutes
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 6 of 9
Functional Test contd.
Acceptance Criteria
Pass
Output pressures
Stable pressure shown on panel
pressure gauges
Output Connections Leak Tight
No Leaks
Fail
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 7 of 9
OPERATIONAL QUALIFICATION
Training
Date Instrument Released for Use:
Trained Operators
Read User
Manual
Practical
Training
Authorized Signature
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 8 of 9
OPERATIONAL QUALIFICATION
Summary Report
Instrument: Genius 1024
Serial Number:
Assessment of Complete Operational Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument is
operating satisfactorily following delivery and installation. The instrument has passed the Operational
Qualification procedure and may now be released for use.
OQ Completed By:
Date:
Deviations Approved By:
Date:
OQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-934
Revision: 1
Page 9 of 9