Title
MS Bench - Installation Guide (All Models)
Revision Number
10
Revision Details
Gauge note Update - LC
Keywords
MS Bench, Table, SCI 1, G SCI 1, Sciex, Installation Guide, IG, Install, G SCI 2, SCI 2
Product Range
Other
Language
UK English
Company Brand
Peak Scientific
Part Number
IG-0028
Extracted text
MS Bench (All Models)
Installation Guide
MS Bench (All Models) Installation Guide Rev 10 RSID 3024 IG-0028 EN 21/03/24
Change History
Rev
Comment
Name
Date
1
Initial Release
L. Couttie
13/06/2019
2
Content Update
L. Couttie
20/11/2019
3
4
Fittings Kit Update
L. Couttie
03/02/2020
SCIEX Feedback Updates
L. Couttie
31/03/2020
5
Fittings Kit Update
L. Couttie
24/08/2020
6
Fittings Kit Update
L. Couttie
06/08/2021
7
Intended Use of Equipment Update
L. Couttie
25/05/2022
8
Drain connection warning added
L. Couttie
02/11/2022
9
Branding Update
L. Couttie
05/12/2023
10
Gauges Note Update
L. Couttie
21/03/2024
How to use this Installation Guide
This document is intended to be used only as a Quick Start Guide, to facilitate safe and
correct installation and initial configuration of your Peak Scientific gas generator.
For more extensive user operation directions, we recommend you download the User
Manual from the website.
Please visit www.peakscientific.com/downloads to
download the full User Manual for your gas generator.
Thank you for selecting Peak Scientific to meet your laboratory gas generation needs,
and should you require any further assistance or support please do not hesitate to
contact Peak Scientific or Peak Partner from which you purchased your generator.
Intended Use of Equipment
MS Bench SCI 1, is a support cabinet with in-built cooling provided by fans, its top surface
and slide out tray are available for use by the customer as required for their planned
application.
MS Bench G SCI 1 is a support cabinet with an in-built Nitrogen generator. Its top surface
and internal storage space are available for use by the customer as required. However, it
CANNOT be used to situate the MS due to vibration emitted by the internal generator.
The inbuilt generator provides an output of both compressed air and nitrogen for supply
to customer application on demand via its compressors.
MS Bench SCI 2 is a support cabinet with two internal compartments, one containing
inbuilt cooling provided by fans and a slide out tray, the other open for a larger roughing
pump. The top surface is available for the customer to use for their planned application.
MS Bench G SCI 2 is a support cabinet with an in-built nitrogen generator, with higher
gas flows than the G SCI 1. The top surface is available for the customer to use for their
planned application. As per G SCI 1 the MS Bench G SCI 2 is NOT suitable for situating
the MS.
Page
2
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this document detail the
most known potential hazards, but by definition cannot be all inclusive. If the user
employs an operating procedure, item of equipment or a method of working which
is not specifically recommended by Peak Scientific, the user must ensure that the
equipment will not be damaged or become hazardous to persons or property.
Symbols
This document uses the following symbols to highlight specific areas important to the
safe and proper use of the Generator
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the Generator or the Application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the Generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
WARNING
WARNING
WARNING
Page 3
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
‘The MS Bench SCI 1 & SCI 2 are designed to support most RUO SCIEX
MS Systems (except TripleTOF 5600+, 6600, and 6600+) per the
weights and dimensions provided in Sciex site prep manuals.
The MS Bench G SCI 1 & G SCI 2 are designed to support Sciex liquid
chromatography (LC) systems. The MS Bench G SCI 1 & G SCI 2 are
NOT suitable for placing the MS.
Fittings Kit
x1
x1
x3*
Flow Control SIlencer 02-1096
8mm Hex Key 00-0007
US Mains Cable C19 6-15 04-1024
x1*
x1
5/16” Compression Fitting
3303538
1x3m*
5/16” Teflon Tubing 3303537
UK Mains Cable C19 04-1025
x2*
x1
1/4” Compression Fitting
02-4421
2x3m*
1/4” Teflon Tubing 00-1266
EU Mains Cable C19 04-1026
x1**
US Mains Cable 120V C19 5-15 04-1035
x1*
1/4” - 6mm Push Fit Fitting
02-4600
1x3m*
6mm OD x 4mm ID PE 00-1282
* MS Bench G SCI 1 & G SCI 2 Only
Page 4
** MS Bench SCI 1 & SCI 2 Only
1.
0°
0°
Unpack the generator from the shipping crate and position on a flat surface, in
desired area.
2.
To extend the levelling feet, first pull the ratchet handle out.
There is a ring pull on the handle to aid removal.
Page 5
3.
Then, ratchet the mechanism to the right in order to lower the
levelling foot until it reaches the floor.
Note: There is a switch on the end of the tab. Pressing this will reverse the direction
of the ratchet mechanism, ratcheting to the left will then raise the foot from the floor.
4.
With the foot in position, push the tab back into the ratchet mechanism.
For MS Bench SCI 1 & SCI 2 proceed to step 8.
For MS Bench G SCI 1 & G SCI 2, continue following all remaining steps.
Page 6
5.
There are Transit Bolts located on the underside of the generator. The G SCI 1 has 2
bolts and the G SCI 2 has 3 bolts.
These MUST be unscrewed and removed before operating the generator.
Please retain the bolts for future transportation.
6.
02-4600
02-1096
Select the ¼”- 6mm Push Fit Fitting 02-4600 and attach to the Drain port at the
rear of the generator.
Attach the three Flow Control Silencer 02-1096 fittings to the Curtain, Source &
Exhaust Ports.
Page 7
7.
00-1282
Connect the 6mm PE Tubing 00-1282 to the ¼”- 6mm Push Fit Fitting 02-4600.
Fit the other end of the tubing to a suitable drain connection or container.
WARNING: The container used MUST be a plastic material and MUST NOT be glass
for safety reasons. The container MUST NOT have an airtight seal as water and air
will be expelled periodically under slight pressure.
8.
04-1024/04-1025/04-1026
Select the appropriate IEC 60320/C19 mains cable from the fittings kit and plug the
cable into the mains input at the rear of the generator.
Page 8
9.
UK
EU
CAUTION
US
230V
230VAC 50/60Hz
Plug the mains cable into the corresponding 220 - 240v ±10% (MS Bench G SCI 1 &
G SCI 2) & 100 - 240v ±10% (MS Bench SCI 1 & SCI 2) single phase power supply.
CAUTION ensure the correct cable has been selected for your electrical supply
10.
Switch the generator power ON using the switch on the rear panel.
For MS Bench SCI 1 & SCI 2 proceed to step 18.
Page 9
11.
45
minutes
Leave the generator switched on for a period of 45 minutes to purge the system.
Note: Gauges are for indication purposes only.
12.
G SCI 1
G SCI 2
Remove the front panel and check gauges have risen to the factory set pressures
stated in the User Manual. The gauges are located in the bottom left hand corner
of the G SCI 1. For G SCI 2 models, the gauges are located behind the fascia panel
which can be removed by pulling towards you.
Page 10
Replace any front panels before continuing.
13.
Switch the generator power OFF using the switch on the front panel.
14.
The purge is now completed and the silencer fittings can be removed from the
Curtain, Source & Exhaust ports.
Page 11
15.
02-4421
3303538
Select the 5/16” Compression Fitting 3303538 & 1/4” Compression Fittings 02-4421
and connect to the CURTAIN and SOURCE & EXHAUST ports respectively.
16.
00-1266
3303537
Attach the 5/16” Teflon Tubing 3303537 to the fitting attached to the CURTAIN port,
then attach the 1/4” Teflon tubing 00-1266 to the fittings attached to the SOURCE &
EXHAUST. Connect the other end of the tubing to the corresponding
gas inlets of your instrument.
Page 12
17.
Switch the generator power ON using the switch on the rear panel.
18.
/
CONGRATULATIONS
Your Peak Scientific MS Bench is now fully installed and operational.
Page 13
General Notes
Electrical requirements are 220 - 240v ±10% (MS Bench G SCI 1 & G SCI 2) &
100 - 240v ±10% (MS Bench SCI 1 & SCI 2). Running at voltages outwith this is not
recommended. Extended periods at extremes can have a detrimental effect on the
operation and life of the Generator.
1.
The Teflon tube supplied in the fittings kit is to allow the generator to be installed
adjacent to the instrument it is supplying. If the generator is to be installed further
away from the instrument then attention needs to be paid to pressure drops that could
occur. The maximum distance away from the instrument the generator can be placed
still using 6mm (4mm internal diameter) or ¼” (3/16” internal diameter) tubing is 10m.
For further information relating to tubing lengths please refer to the user manual.
2.
Please familiarise yourself with the full User Manual available at www.peakscientific.
com/downloads. This includes all other technical specifications, operating
instructions, service requirements, contact details and trouble shooting.
Page 14
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/register.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 15
Please visit www.peakscientific.com/downloads to
download the full User Manual for your gas generator.
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com.
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
MS Bench (All Models) Installation Guide Rev 10 RSID 3024 IG-0028 EN 21/03/24