Infinity XE 50X0 Series
User Manual
Copyright © Peak Scientific Instruments Ltd 2024 — Infinity XE 50X0 Series User Manual Rev 4 RSID 2750 EN —20/06/2024
Contents
Change History
3
How to use this Manual
3
Introduction
4
Warranties and Liabilities
5
Warranty & Liability Coverage
5
Safety Notices
7
Symbols
7
Safety Notice to Users
7
EU Declaration of Conformity
8
UK Declaration of Conformity
9
Environmental Declaration
10
Technical Specification
11
Unpacking
13
Fittings Kit Contents
14
Installation
15
Generator Environment
15
Generator Overview
16
General Dimensions
16
Rear Connections
17
Inlet Air Quality
17
Electrical Connection
18
Electrical Panel Access
18
Normal Operation
19
On Demand Gas
19
Eco Mode
19
Commissioning
20
Tubing Lengths
20
Service Requirements
21
Service Schedule
21
Peak Protected
22
Cleaning
23
Safely Isolated Condition
23
Safe State After Repair
23
Troubleshooting
24
Register Warranty
25
Supported Maintenance Guide
26
Service Parts List
29
Page 2
Change History
Rev
Comment
Name
Date
2
Spec Update
D.Lai
19/07/2021
3
Delerations Update
D.Lai
12/10/2021
4
SMG Added
C. Denholm
18/04/2021
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The Infinity XE 50 Series delivers performance and value beyond expectations,
offering a pure and reliable source of gas. These perpetual flow nitrogen generators
are suitable for a variety of applications in your laboratory.
These models provide a source of nitrogen gas with other features including:
•
Single Source Solution - One generator can cater for the requirement of multiple
applications & multiple laboratories on site
•
Durable - Few moving parts with minimum maintenance requirements
•
Quiet in operation – Silent whilst operating
•
Mobile – Flexibility to position the generator where required
•
Simple installation – Generator designed as a plug and play system
•
Economical – More cost effective than any other gas supply method
•
Convenient – Gas on demand, no hazards, no need to worry about running out of
gas
•
Eco-mode – Energy efficient eco-mode ensures low running costs
The Infinity XE 50 Series generators are based on proven technology, and selectively
remove oxygen, moisture and other gases to leave clean, dry, phthalate free nitrogen.
To ensure these Generator models meet our high expectations with regards to
reliability and performance, we have tested them extensively at our manufacturing
plant and with end users around the world to ensure reliability and longevity of the
system.
Page 4
Warranties and Liabilities
Warranty & Liability Coverage
1.
Peak warrants that, subject to the provisions in this statement, purchased Peak generators, whether
purchased directly from Peak or indirectly via an approved, certified and trained distributor or
partner (referred to hereafter as a “Peak Partner”) will comply in all material respects with any
specifications referred to in your customer order confirmation and, subject to installation and
operational guidelines being followed as described in applicable product manuals, shall be free
from any defects in quality of materials or workmanship for a period of one year from the date of
installation, provided this takes place within 3 months of factory dispatch.
2. Where the purchased generator is from the Precision Hydrogen series, Peak further warrants that,
subject to installation and operational guidelines being followed as described in applicable product
manuals, the hydrogen cell shall be free from any defects in quality of materials or workmanship
for a total period of three years (inclusive of warranty period specified in clause 1) from date of
installation, provided this takes place within 3 months of factory dispatch.
3. Where the purchased generator is from the i-Flow 6000 series, Peak further warrants that, subject
to installation and operational guidelines being followed as described in applicable product
manuals, the generator shall be free from any defects in quality of materials or workmanship
for a total period of two years (inclusive of warranty period specified in clause 1) from the date
of installation, provided this takes place within 3 months of factory dispatch and the following
provisions have also been met:
a. you must purchase a service plan, ensuring the generator is serviced by Peak or a Peak Partner on
or before the end of the first 12 months of your ownership, and serviced at least once during each
subsequent 12 month period thereafter;
b. the generator (and any associated equipment)
must have been commissioned by Peak or a Peak Partner;
c. the feed air or inlet air supply to the generator must comply with ISO 8573-1:2010 Class 1.2.1 at all
times;
d. your air compressor, dryer, filtration and oil removal systems must be deemed suitable for use
by Peak or a Peak Partner, and must be changed and serviced regularly, in line with the equipment
manufacturer’s recommended guidelines; and
e. any generator failure or fault that is deemed to have been caused by the failure of any upstream
equipment, component, part or system (such as air compressor, air treatment or filtration) will be
excluded from the warranty described herein.
4. Peak also warrants that any replacement parts whether purchased (directly from Peak, or via a Peak
Partner) or supplied as part of any remedial action undertaken in line with the provisions of clauses
12 and 13, shall be free from any defects in quality of materials or workmanship for a period of 180
days from the date of factory dispatch, provided its installation is performed by Peak or a Peak
Partner.
5. This warranty does not exclude Peak’s liability in respect of any claim for death or personal
injury to any person, in so far as such can be attributed to negligence or breach of duty of care
directly resulting from failure of Peak to comply with the provisions in clauses 1, 2, 3 & 4.
Exclusions & Limitations
6. This warranty does not cover:
a. damage, deterioration or malfunction resulting from an alteration or modification to a generator
which has not been carried out by Peak or a Peak Partner;
b. damage, deterioration or malfunction resulting from what Peak reasonably believes to be abuse,
or misuse of a generator by you or any third party;
c. liability for accident or neglect (other than pursuant to clause 5);
d. maintenance or repairs which have not been carried out by Peak or a Peak Partner;
e. operation of a generator or exposure of a generator to environmental conditions that fall out-with
operational guidelines as specified in the applicable product user manual; and
f. lightning, power surges or any other acts of God or nature.
7.
This warranty is non-transferrable. Only the original owner of the generator may benefit from the
terms within this statement.
Page 5
8. Peak shall not be liable in respect of any claim made for costs, damages, losses or expenses
(whether consequential, direct, indirect or otherwise) or in any respect howsoever arising including,
but not limited to, liability from accident or negligence (other than pursuant to clause 5) that may
be suffered by you or any third party.
9. No person or entity is authorised to change the terms and conditions outlined in this warranty
statement in any respect, or to create any additional obligations or liabilities for any party involved.
10. This warranty statement supersedes any and all prior warranty agreements between the parties
and constitutes the complete, final and exclusive understanding of the parties with respect to the
subject matter. All prior negotiations, representations, or promises, whether oral or written, of either
party shall be deemed to have been merged herein.
11. If any part of this warranty statement is invalidated, for whatever reason, such part will be deleted
and the rest shall remain unaffected, continuing to be in full force and effect.
Delivery of Warranty Service
12. Subject to clause 13, and:
a. Peak being notified by you, within the duration of the applicable warranty period, of any defect
that you think is subject to any warranty valid under clauses 1, 2, 3 or 4; and
b. Peak being permitted to inspect the generators, parts and their installation (along with any
relevant packaging)
Peak shall at its option repair or replace defective generators or parts (including, if necessary, any
moving parts and irrespective of runtime). No additional charges will apply, for parts or delivery
and, where applicable, labour or travel. Peak will endeavour to deliver this service within 3 working
days of your notification.
13. Where, in Peak’s reasonable opinion, a defect is subject to an exclusion described in clause 6, Peak
reserves the right to charge for parts or delivery and, where applicable, you may also be charged by
Peak for call out, labour or travel in respect of any repair or replacement which you authorize Peak
to carry out.
Page 6
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which
may represent a potential hazard. The warnings detailed within this manual refer to
the most likely potential hazards, but by definition cannot be all inclusive. If the user
employs an operating procedure, item of equipment or a method of working which
is not specifically recommended by Peak Scientific, the user must ensure that the
equipment will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the
safe and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
WARNING
Page 7
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the
generator in a manner not specified by Peak Scientific MAY impair the
SAFETY provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment maybe
impaired.
EU Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen Gas Generator
Models:
Infinity XE 50 Series
To which this declaration relates, is in conformity with the following applicable EU
Directives, harmonized standards, and other normative requirements.
•
Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use.
•
Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for Measurement, Control and Laboratory Use
– EMC Requirements. (Class A)
•
Restriction on the use of certain hazardous substances in electronic equipment
(RoHS) Directive 2011/65/EU as amended by EU 2015/863.
•
FCC 47 CFR Part 15 class A
Unintentional radiators; Conducted and Radiated emissions limits.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
Page 8
UK Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen Gas Generator
Models:
Infinity XE 50 Series
To which this declaration relates, is in conformity with the following applicable UK
Statutory Instruments, Standards and other normative requirements.
•
The Electrical Equipment (Safety) Regulations 2016 (SI 2016 / 1101) as amended.
BS61010-1:2010 Safety Requirements for Electrical Equipment for Measurement
Control and Laboratory Use.
•
The Electromagnetic Compatibility Regulations 2016 (SI 2016 / 1091) as amended.
BS61326-1:2013 Electrical Equipment for Measurement , Control and Laboratory Use
– EMC Requirements.
•
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012 (SI 2012 / 3032) as amended.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
Page 9
WEEE Compliance Statement
The Waste Electrical and Electronic Equipment (WEEE) Regulations SI 2013 No 3113
and or the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU
apply to all electrical and electronic equipment placed on the market in the UK and EU
covered by the scope of regulations which can be found in the Government Guidance
Notes (PDF) produced by the Department for Business Innovation and skills for the UK
and here for Europe.
All PEAK products that are subject to the WEEE directive are compliant with the
WEEE marking requirement. Such products are marked with the “crossed-out wheelie
bin” symbol (shown below) in accordance with European standard EN50419. All
old electrical equipment can be recycled. Please do not dispose of any electrical
equipment (including those marked with this symbol) in general rubbish bins. Please
contact your dealer or distributor for clarity.
Page 10
Technical Specification
Infinity XE 5010 & 5020
Environment
5010
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Relative Humidity
5020
5°C (41°F)
30°C (86°F)
80%
*
Maximum safety ambient temperature 30°C
**
When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Inlet Pressure
Maximum Air Inlet Pressure
Minimum Air Inlet Flow
Minimum Air Quality
4.1 bar (60 psi)
9.9 bar (145 psi)
Dependant on required flow and purity
ISO 8573-1:2010 Class [1.4.1]
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet)
Maximum Outlet Flow (95% Nitrogen)
Outlet Nitrogen Purity
Start-Up Time
Particles
Pressure Gauges
0.7bar (10psi)
130 L/min
260 L/min
95 - 99.5%
30 minutes
<0.01µm
2
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
100-240 v +/- 10%
60 Hz / 50 Hz
0.5A
C20 Plug
C19 socket to local connection (13A min 16A Max)
2
I
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 11
50 x 50 x 99 (19.6 x 19.6 x 39.0)
50.4 kg (111 lb)
54.6 kg (120 lb)
67 kg (148 lb)
72 kg (159 lb)
Technical Specification
Infinity XE 5030 & 5040
Environment
5030
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Relative Humidity
5040
5°C (41°F)
30°C (86°F)
80%
*
Maximum safety ambient temperature 30°C
**
When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation.
Inlet Conditions
Minimum Air Inlet Pressure
Maximum Air Inlet Pressure
Minimum Air Inlet Flow
Minimum Air Quality
4.1 bar (60 psi)
9.9 bar (145 psi)
Dependant on required flow and purity
ISO 8573-1:2010 Class [1.4.1]
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet)
Maximum Outlet Flow (95% Nitrogen)
Outlet Nitrogen Purity
Start-Up Time
Particles
Pressure Gauges
1.4bar (20psi)
390 L/min
520 L/min
95 - 99.5%
30 minutes
<0.01µm
2
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category
100-240 v +/- 10%
60 Hz / 50 Hz
0.5A
C20 Plug
C19 socket to local connection (13A min 16A Max)
2
I
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 12
50 x 50 x 99 (19.6 x 19.6 x 39.0)
58.8 kg (130 lb)
63 kg (138 lb)
77 kg (170 lb)
82 kg (180 lb)
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator to
the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads
for UK, EU & US and also all the required fittings and warranty registration card. Be
careful not to discard these with the packaging.
Page 13
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. 6mm PE Tubing x 3m
2. 12mm PE Tubing x 6m
3. 6mm Push-Fit Fitting x 1
4. 12mm Push-Fit Fitting x 2*
5. 16mm Push-Fit Fitting x 1**
6. Tee fitting
x 1
7. Shut-off Valve x 1
8. UK Mains Power Cable x 1
9. EU Mains Power Cable x 1
10. JP Mains Power Cable x 1
11. US Mains Power Cable 230v
x1
12. US Mains Power Cable 110v
x1
13. 4mm Hex Key x 1
14. 5mm Hex Key x 1
15. 8mm Hex Key x 1
All of the generators output ports are located on the output panel at the rear of the
unit.
* x1 with 5030 & 5040
** 5030 and 5040 only
Page 14
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the
air flow around the unit. It is recommended that an air gap of 75mm (3”) should be
maintained down both sides and at the rear of the unit. Please refer to the drawing
below for the general dimensions of the unit.
The generator should be situated in a well ventilated environment, with at least a
75mm (3”) air gap all around it and should be positioned such that it can be easily
disconnected from the mains supply if required
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
30 °C (86 °F)
Page 15
Generator Overview
General Dimensions
500 mm / 19.6”
500 mm / 19.6”
990 mm / 39"
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING
Page 16
Rear Connections
Mains Cable
12mm Push-Fit Fitting (5010,5020)
16mm Push-Fit Fitting (5030,5040)
Tee Fitting/Shut-Off Valve/12mm Push-Fit Fitting
6mm Push-Fit Fitting
Inlet Air Quality
The Nitrogen Generator should be connected to a ISO 8573-1:2010 (1:4:1) source of
compressed air. A minimum pressure of 60 psig is required for efficient operation of
the Generator. Any doubts as to the suitability of your compressed air supply should
be referred to the factory for advice.
A MAXIMUM OIL VAPOUR CONTENT OF 0.01 mg per m^3 is PERMISSIBLE
Air Inlet and Nitrogen Outlet connections are at the rear of the generator in the middle right side of the panel. The Air inlet connection is the right port, 12mm Push-Fit
Fitting (5010,5020) 16mm Push-Fit Fitting (5030,5040), Outlet port is 12mm Push-Fit
Fitting
WARNING
Page 17
Ensure all inlets and outlets are connected to correct sources and
applications”, “All Connections should only be carried out by trained
personnel”, “Generator must be switched off and unplugged prior to
any cleaning or maintenance operations
Electrical Connection
Connect the generator to an appropriate 85-264 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps,
can be fitted by a qualified electrician.
If an appropriate mains power cords is not supplied or a substitute one is used then
ensure that all components of it the plug, cord and connector have adequate ratings
for the generator and appropriate approvals for the country of use. Failure to do so
could cause damage to the generator or risk overloading of the power cord
WARNING
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow
or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical Panel Access
Changing settings within this enclosure may adversely affect the
operation of the generator. This enclosure should not be accessed
unless by or under direct instruction from a trained individual.
WARNING
Page 18
Normal Operation
The Infinity XE 50 Series Gas Generators are designed specifically to minimize
operator involvement. Given that the systems are installed as described in
earlier sections and is serviced in accordance with the specified maintenance
recommendations (see Service Requirements), then it should simply be a matter of
turning the Generator on when it is required.
The Generator will automatically produce the factory set flow and pressure as detailed
in the Technical Specifications.
On Demand Gas
The generator will produce nitrogen on demand. If the application is operating
and requires a gas flow, the system it is connected to will supply this to suit the
requirements of the application. If the application requirement for gas stops, the
system will also stop. If the demand from the application starts again, the system will
detect the demand for gas and will automatically start again to suit the demand.
Eco Mode
When demand from the instrument is below 1.2L/min the Infinity XE generator will
stop demanding air from the external compressor. If demand from the application
starts again, the system will detect the demand for gas and will automatically start up
again.
Page 19
Commissioning
With the generator installed as previously described, connect the application to the
Nitrogen output port and open the air supply to the unit. On reaching operational
pressure the generator will produce Nitrogen at the customer-specified purity. The
design of the generator is such that it will deliver up to the rated output flow of
Nitrogen, determined by demand of the consuming equipment. Rated flow, purity
and pressure are customer-specific and dependent on inlet conditions being met as
detailed on the individual generator
The design of the generator is that it will deliver variable purity dependant on the
inlet pressure and outlet flow rate. Outlet pressure may be adjusted by the pressure
regulator on the rear of the unit, allowing for up to 20 psi pressure drop across the
system. Should the demand for the nitrogen stop, the generator will automatically go
into "eco-mode" (refer to page 18), shutting off the air inlet. Once the demand restarts
the generator automaticcaly resumes producing Nitrogen. The generator is protected
from over pressurising.
Tubing Lengths
CAUTION
The diameter of the tubing which will be connected to the gas outlet
is important and is determined by the length of tubing required.
Failure to follow these recommendations could lead to pressure
between generator and application.
> 10 - 40 meters:
Use 12/10 (12mm O/D, 10mm I/D) Tubing.
> 40 metres:
Please contact Peak Scientific with the relevant distance
andwe
will calculate the flow resistance and the
tubing size required.
The imperial equivalents are: 12/10 = 1/2” O/D, 3/8” I/D.
Air input tubing for 5030 & 5040 is not supplied. 16mm OD, 12mm ID PTFE Tubing is
reccomended.
Page 20
Service Requirements
Service Schedule
Purchase Interval
Component
Visit
12 Months
Infinity XE 50 Series Annual Maintenance
Kit
www.peakscientific.com/ordering
Page 21
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit
your lab to install and setup your
generator
IQ/OQ
Certified assurance for applications
requiring documented qualification
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance
Spares
Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime
Premium Protected
Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance
Technical Support Hotline
Around the clock support by phone
or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 22
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the generator.
CAUTION
Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Safely Isolated Condition
The unit is in a safely isolated condition when it is disconnected from its application
and fully de-pressurised. Directions for isolating the Generator are shown below.
WARNING
Failure to place the Generator in a safely isolated condition when
instructed to do so may lead to personal injury or injury to others and
even death.
1. Switch generator off.
2. Disconnect from air supply.
3. Ensure the output pressure gauge reads zero. (If gauge does not fall to zero, open
the manual ball valve at the outlet port, to allow trapped gas to escape.)
4. Disconnect from the application.
Safe State After Repair
To guarantee the generator is in a safe state after a repair, please ensure the following
conditions are met;
•
All earth leads have been reconnected.
Earth leads should be connected to the earth tabs located on the door panels, as
shown below.
•
The electricity cable has been correctly refitted.
The mains cable should be plugged into the IEC power inlet located on the rear
panel of the generator. This unit is classified as SAFETY CLASS 1. THIS UNIT MUST
BE EARTHED.
Page 23
Troubleshooting
Problem
The Generator will not switch on and the
power switch does not illuminate.
Possible Solution
•
Ensure power cord is plugged into the
Generator and that the power socket is
turned on.
•
Check the fuse in the power cord plug.
•
Contact your service provider.
•
Disconnect power cord from the rear of
the Generator. Open the left hand panel
are check that the circuit breaker is turned
on (switch in the up position). Reconnect
power cord.
•
Contact your service provider.
•
Check the pressure regulator at the back
of the unit has not been adjusted.
•
Check to see if the inlet pressure gauge
has dropped below the set air supply
pressure.
•
If so, check your air supply does not have
a pressure drop over the length of tubing
and has no leaks.
•
Contact your service provider.
The Generator will not switch on but the
power switch is illuminated.
The output pressure has dropped below its
original set point.
Page 24
Go Online or Complete and Return
We know that registering any of your recently purchased products is not
the first thing on your mind- but it is very important to both of us. Not all
warranties are alike and Peak Scientific stand out against other gas suppliers as
we offer a comprehensive, quick response, on-site warranty. This means that
in the very unlikely case that your gas generator develops a fault we have rapid
support teams on-hand around the world who are able to come to your lab and
get you back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support
packages. Peak [Protected] aftercare support can guarantee an on-site
response within 72 hours*, genuine parts from our ISO9001 approved factory
and a 95% first-time fix rate. See our enclosed Peak [Protected] leaflet for
further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once
registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation
of your generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.
com. For generators that remain unregistered after 1 month from the shipment date, the warranty will
be considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 25
Supported Maintenance Plan
Introduction
The following is a guide for maintenance of the Infinity XE 50 series as provided for by
the Peak Service Plan.
This procedure is to be performed on an annual basis
1. Check that the customer’s application is not in use and that it is OK to disconnect
the Gas Supply. Check that the Gas Demand is turned OFF.
2. If possible, shut OFF the customer’s air supply to the unit. Turn the Generator OFF.
3. Turn the red ball valve at the N2 output to release all pressure between the
generator and the customer’s application and disconnect once fully depressurised.
Check depressurisation with the gauges on the front of the generator.
4. Remove the side panels and disconnect the earth tabs.
Page 26
Stage 1 filtration – 0-01μm Particulate
5. Remove the four M5 bolts on the top of the filter using the 5mm hex key.
6. Remove the bowl from the filter assembly and replace the filter element 3300301.
Reverse steps 5 to reassemble.
Page 27
Stage 2 filtration – Active Carbon
7. Remove four M5 bolts on the top of the filter using the 5mm hex key.
8. Remove the bowl from the filter assembly and replace the filter element 3300302.
Reverse steps 7 to reassemble.
9. Attach the side panels to the generator. ENSURE EARTH TABS ARE RE-ATTACHED.
10. Carefully open the customer’s air supply to the generator.
11. Reconnect to the customer’s unit and turn the Generator ON.
Page 28
Service Parts List
Service parts Infinity XE 50 Series
Service
Component
Part No.
Qty.
12 months
Element, filter 0.01 μm
3300301
1
12 months
Element, filter active carbon
3300302
1
As an alternative to purchasing the 12 month service items individually an Annual
Service Kit is available as one part number. This contains all the filters required for this
Generator and offers a costs saving over buying the components separately.
Purchase interval
12 months
Page 29
Component
Infinity XE 50 Annual Service Kit
Part No.
3300317
Qty.
1
Page 30
Page 31
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid
response service to our customers. With [Peak Protected], your laboratory’s productivity becomes
our top priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2024 — Infinity XE 50X0 Series User Manual Rev 4 RSID 2750 EN —20/06/2024