NM18LA - User Manual
Original PDF File1.2 MB |
Download |
View directly in browser1.2 MB |
View in browser |
Resource ID
202
Access
Open
Contributed by
Derek Rennie
Review date
01 May 24
Title
NM18LA - User Manual
Revision Number
Revision Details
Reviewed 31/12/2020 - BP
Keywords
User Manual,
Product Range
Language
Company Brand
Reviewed on 3rd may 2017 - LC
Extracted text
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Directions For Use
High Purity Nitrogen Generator
NM18LA (230V 50/60Hz AC)
Peak Scientific Instruments Ltd Fountain Crescent Inchinnan Business Park Inchinnan Renfrewshire PA4 9RE T. (Europe) +44 (0) 141 812 8100 T. (USA) +1 866 647 1649 (sales) T. (USA) +1 866 732 5427 (tech help) F. (Europe) +44 (0) 141 812 8200 info@peakscientific.com sales@peakscientific.com support@peakscientific.com
www.peakscientific.com
NM18LA 230v issue 7
1
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Document Change History
Issue No
1 2 3 4 5 6 7
Changed
Document Created Updated Jan 2003 Update (Inlet Filter Position) System Diagram Update Technical support phone number updated Weight updated New Style Front Added
Initials
SGM SGM SGM MK FAD FAD FAD
Date
01/03/02 04/09/02 13/01/03 16/07/04 10/11/04 16/11/04 07/04/05
NM18LA 230v issue 7
2
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Contents
Page No Document History Warranty Statement Introduction Unpacking & Installation Electrical Connection Air Connection Principle of Operation Commissioning Routine Maintenance Technical Specifications System Diagrams Maintenance Log Notes 2 4 5 5 5 6 6 7 8 11 12 14 15
NM18LA 230v issue 7
3
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Warranties and Liabilities
1) The Company warrants that it has title to the Goods. 2) Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material
respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory.
3) Save as provided in this clause and except where the Goods are sold to a person dealing as a
consumer (within the meaning of the Unfair Contract Terms Act 1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause.
4) In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the
Customer must:-
4.1) Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts; and 4.2) Allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company.
5) Subject to the Company being notified of any defect as is referred to in sub-clause 2 hereof within a
reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof).
6) The Company shall be liable to indemnify the Customer in respect of any claim for death or personal
injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof.
7) Save as provided in sub-clause 2 hereof the Company shall not be liable in respect of any claim by the
Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party,
SAFETY NOTICE TO USERS
These instructions must be read thoroughly and understood before installation and operation of your Peak Nitrogen Generator. Use of the Generator in a manner not specified by Peak Scientific Inst. MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations. WARNING: Nitrogen is not a poisonous gas, but if the concentration in the inhaled air becomes too high there will be a risk of asphyxiation.
NM18LA 230v issue 7
4
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
1
Introduction
The Peak Scientific Instruments Ltd NM18LA Gas Generator is designed specifically for use with Laboratory Analytical Instruments as a source of carrier gas.
2
Unpacking and Installation.
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage.
ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS PURCHASED.
After unpacking and a visual inspection, the unit should be placed in a ventilated area away from direct sunlight. Care should be taken not to obstruct the ventilation holes on the sides of the unit nor the fan outlet at the rear. The generator should be placed on a steady and level base. It has been designed to fit under most benches. The generator may be supplied with castors if this option has been requested. The front two castors are fitted with brakes. These should be applied when the unit is in use. Performance of the generator (like all sophisticated equipment) is affected by ambient temperatures. Continuous operation in ambient temperatures exceeding 25°C will lead to a reduction in capacity and prolonged operation in temperatures exceeding 28°C will shorten the life of the unit. Note should also be taken of the proximity of Air Conditioning outlets. These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the generator within such a pocket could adversely affect its performance.
3
Electrical Connection
Important Electrical Notice
This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be
EARTH/GROUND (E): LIVE (L): Neutral (N): Fuse
Green & Yellow Brown Blue
The generator protection fuse is in the pull out drawer of the mains inlet IEC euro connector located on the bottom right hand side of the cabinet adjacent to the off/on switch. The fuse is rated at 10.0 A. The drawer also holds a spare. Connect the generator to a single-phase supply using the power cord provided.
NM18LA 230v issue 7
5
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
4
Air Connection
The NM18LA Nitrogen Generator has its own in-built air compressor and therefore does not need an external source of compressed air. There is a single ¼” Female bulkhead connection on the left hand side of the unit. In addition this is a drain port to the rear of the unit. The generator may liberate a considerable amount of water through this drain, which should be led to a suitable drain connection or container.
5
Principle of Operation
Peak Scientific Instruments Membrane Generators utilize Hollow Fibre Membrane Technology to efficiently separate Nitrogen from other gases present in ambient air. The membrane operates on the principle of selective permeation in that so-called “Fast” gases such as H2O, CO2 & Oxygen will permeate through the membrane wall whilst so-called “Slow” gases will not and continue along the membrane tube and are thus available for collection and use.
Unloading Valve
S C I E N T I F I C
Heat Exchanger Pressure Regulator
Zero Air
Front Panel
RAC Filter
Membrane
Pressure Switch Filter Separator
Outlet
Receiver Compressor
NM18LA Pneumatic Diagram
Air is drawn into the system by the Compressor and passed via the Heat Exchanger and the Filter/Separator to the Membrane. After the Membrane the separated Nitrogen is passed into the Air Receiver. The Nitrogen is regulated and through an Activated Carbon Filter to remove any remaining impurities.
NM18LA 230v issue 7
6
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
6
Commissioning
With the Generator installed as described earlier connect power to the unit and turn it ON. Disconnect the Nitrogen Outlet connection to allow the generator to vent to atmosphere until the unit is stabilised. The Generator has been pre-set in the factory to give the specified output flow-rate and pressure in line with known requirements; however some minor adjustment might be necessary. Once the Membrane reaches the design pressure the Generator will stabilise and produce pure Nitrogen. Maximum purity will be achieved after around 1 hour. After this time the generator can be re-connected to the application. The design of the generator is that it can deliver up to 18 Litres / min of Nitrogen at up to 100 psig. Should the demand for Nitrogen be less than the rated output flow, or indeed should the demand stop the generator will continue to operate without any problems. The generator is protected from over-pressure and will un-load should demand be reduced. After a period of around 5 minutes with no load the generator will stop.
Pressure & Flow Settings
Output pressures and flow rates are controlled by pressure regulators located inside the front cover.
Flow Regulator
Pressure Regulator
The regulators are pre-set at the factory and should not require adjustment.
NM18LA 230v issue 7
7
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
7
Routine Inspection
by a TECHNICALLY COMPETENT PERSON with the Generator safely isolated.
WARNING: Servicing and/or repair of the Generator should only be undertaken
Due to the simplicity of the design and the small number of moving parts the Nitrogen Generator will have a long and trouble free life. However the following components should be replaced as follows:
Compressor Inlet Filter Filter Separator & Silencer Element Compressor Units (the lesser of)
Every 6- months Every 12- months Every 6000Hours or 18- months
Service kits are available for all routine maintenance; please contact the factory for further details.
FAILURE TO FOLLOW THE PRESCRIBED MAINTENANCE PLAN WILL INVALIDATE THE PRODUCT WARRANTY Inlet Filter / Separator Elements
These should be changed at intervals as indicated below. In addition filter bowls should be cleaned and the operation of the auto-drains should be checked.
Compressor Inlet Filter
This should be changed at 6-month intervals. Part Number is 02-4187 and the filter is located as shown.
Inlet Filter (02-4187)
Remove the cover by rotating it anti-clockwise ¼ turn. The element can then be removed. Re-fitting is the reverse procedure.
NM18LA 230v issue 7
8
PEAK Scientific Instruments Ltd
Instructions for use Manual
Filter Separator Elements
High Purity Nitrogen Generator 7th April 2005
The Coalescing Filter Element (02-4335) should be changed at 12-month intervals.
The filter housing is removed by depressing the black slide and twisting the bowl through a ¼ turn. The element then un-screws counter clockwise. Re-assembly is the reverse procedure. IMPORTANT!!! Ensure that the system is NOT under pressure before attempting to remove the bowl and that the air supply is shut-off.
NM18LA 230v issue 7
9
PEAK Scientific Instruments Ltd
Instructions for use Manual
Active Carbon Filter
High Purity Nitrogen Generator 7th April 2005
The Element is as shown, and should be changed at 12-month intervals. The part number is 00-4425
Active Carbon Element (00-4425)
The filter housing is removed by un-screwing in a counter clockwise direction. IMPORTANT!!! Ensure that the system is NOT under pressure before attempting to remove the bowl and that the air supply is shut-off.
NM18LA 230v issue 7
10
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
8
Technical Specifications
General Details
5 oC (41 oF) 35 oC (95 oF)
Minimum Operating Ambient Temperature Maximum Operating Ambient Temperature Inlet Conditions (Free of oil and bulk moisture) Minimum Air Inlet Pressure Maximum Air Inlet Pressure Minimum Air Inlet Flow Rate Outlet Gas Maximum Pressure Drop (Outlet-Inlet) δP Maximum Gas Outlet Pressure Maximum Gas Outlet Flow (Nitrogen) Start up time for Purity Particles
N/A N/A N/A
N/A 100 psig 18 Litres/min (ATP) 30 minutes 0.01um
Electrical Requirements @230V ac (50/60Hz) Fuse (230V 50Hz ac) Electrical Connection 3.6 Amps 10 Amps IEC-Euro connector
General Dimensions W x D x H cm (inches) Kg (lbs) Kg (lbs) 88 x 43 x 41 (34.6 x 17 x 16) 41 (90) 79 (173.8)
Weight Shipping Weight
NM18LA 230v issue 7
11
NM18LA 230v issue 7
Cooling Fan
PEAK Scientific Instruments Ltd Instructions for use Manual
Pressure Switch
Contactor
Unloading Valve
Cooling Fan
Cooling Fan
High Purity Nitrogen Generator 7th April 2005
Cooling Fan Compressor On/Off Switch Euro Connector
12
NM18LA Electrical Diagram
NM18LA 230v issue 7
Zero Air
PEAK Scientific Instruments Ltd Instructions for use Manual
Unloading Valve
S C I E N T I F I C
Heat Exchanger Pressure Regulator
Front Panel
RAC Filter
Membrane
Pressure Switch Filter Separator
Outlet
High Purity Nitrogen Generator 7th April 2005
Receiver Compressor
NM18LA Pneumatic Diagram
13
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Maintenance Record Log
Model. Serial number
Work done
Remarks
Date Name
NM18LA 230v issue 7
14
PEAK Scientific Instruments Ltd
Instructions for use Manual
High Purity Nitrogen Generator 7th April 2005
Notes
NM18LA 230v issue 7
15