MS Table (All Models) - User Manual 

Full screen preview

Resource tools

Original PDF File

1.4 MB Download
  •  Share
Resource details

Resource ID

1927

Access

Open

Contributed by

Debora LU

Review date

03 May 19

Title

MS Table (All Models) - User Manual

Revision Number

2

Revision Details

Update flow rates - david lai

Keywords

MS, Mass, Spec, Table, User, Manual, All, Models, UM-0012

Product Range

Other

Language

UK English

Company Brand

Peak Scientific

Public / Private

public

Reviewed on 3rd may 2017 - LC

Extracted text

MS Bench (All Models)
User Manual

MS Bench (All Models) User Manual Rev 1 RSID 3023 EN 31/05/19

Contents
Change History
3
How to use this Manual
3
Introduction 4
Warranties and Liabilities
5
Safety Notices
6
Symbols 6
Safety Notice to Users
6
Declaration of Conformity
7
Environmental Declaration
8
Technical Specification
9
Unpacking 10
Fittings Kit Contents
11
Installation 12
Generator Environment
12
Generator Overview
13
General Dimensions
13
Rear Connections
14
Unit Controls
14
Drain Connection
15
Electrical Connection
16
Start-Up Sequence
17
Connecting to the application
18
Tubing Lengths
18
Normal Operation
19
On Demand Gas
19
Generator Cycling
19
Unusual Operation
19
Service Requirements
20
Service Schedule
20
Service Indication
21
Stage 1
21
Stage 2
21
Service Indication Reset
21
Peak Protected
22
Cleaning 23
Run Time Alarm
24
Troubleshooting 25

Page 2

Change History
Rev

Comment

Name

Date

How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.

Page 3

Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods
shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free
from defects in material and workmanship for the lesser of a period of twelve
months from the date of delivery or thirteen months from the date of dispatch
from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person
dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common
law are hereby expressly excluded save to the extent they may not be lawfully
excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of
this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of
clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality
of the foregoing that any defect is not the direct or indirect result of lack of repair
and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or
incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any relevant
packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in sub-
clause 2 hereof within a reasonable time of it becoming apparent and subject always
to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or
repair the defective Goods or refund a proportionate part of the Price. The Company
shall have no further liability to the Customer (save as mentioned in sub-clause 6
hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim
for death or personal injury to any person in so far as such is attributable to the
negligence or breach of duty of the Company or any failure by the Company to
comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect
of any claim by the Customer for costs, damages, loss or expenses (whether direct,
indirect, consequential or otherwise) or indemnity in any respect howsoever arising
including, but not by way of limitation, liability arising in negligence (other than
pursuant to clause 6 above) that may be suffered by the Customer or any third
party.

Page 4

Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.

Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the generator.

WARNING

A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.

CAUTION

A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.

Safety Notice to Users

WARNING

WARNING

These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.

WARNING

Page 5

Declaration of Conformity
Unit currently under accreditation.

Page 6

Environmental Declaration
Unit currently under accreditation.

Page 7

Technical Specification
Environment
MS Bench SCI 1

MS Bench G SCI 1

Minimum Operating Ambient Temperature

5°C (41°F)

Maximum Operating Ambient Temperature

30°C (86°F)

Maximum Altitude

2000 m

Maximum Relative Humidity

80% Non-Condensing

Minimum Storage Temperature*

-20°C (-4°F)

Maximum Storage Temperature*

60°C (140°F)

*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.

Generator Outlets
Curtain Maximum Flow^

19 L/min @ 4.48 bar (0.670
cfm @ 65 psi)

Source Maximum Flow^

26 L/min @ 6.89 bar (0.918
cfm @ 100 psi)

Exhaust Maximum Flow^

N/A

Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlet

25 L/min @ 4.13 bar (0.882
cfm @ 60 psi)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP

^NOTE – These performance characteristics are valid only when paired with an approved Sciex application. Please refer
to current application matrix for a list of supported products.

Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord (Supplied)
Pollution Degree
Insulation Category

100 - 240v ±10%

220 - 240v ±10%
50/60 Hz

2 Amps

7 Amps

C20 Plug
C19 socket to local connection (13A minimum)
2
II

General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)

Page 8

91.4 x 78.7 x 76.2 (36 x 31 x 30)
85 (187)
120 (264)
100 (220)
135 (297)

Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking.

Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator to
the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful
not to discard these with the packaging. 

Page 9

Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. ¼” Teflon Tubing* x 9m

2. 6mm PE Tubing* x 3m
3. ¼” Compression Fitting* x 3

4. 6mm Push-Fit Fitting*



x 1

5. Flow Control Silencer* x 3
6. UK Mains Power Cable 230v

x1

7. EU Mains Power Cable 230v

x1

8. US Mains Power Cable 230v

x1

* MS Bench G SCI 1 only
All of the generators output ports are located on the output panel at the rear of the
unit.

Page 10

Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to the
application it is supplying. If this is not convenient then the unit can be sited elsewhere,
however, consideration should be made of the lengths of pipe runs as pressure drops
can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the
air flow around the unit. It is recommended that an air gap of 75mm (3”) should be
maintained down both sides and at the rear of the unit. Please refer to the drawing on
the following page for the general dimensions of the unit.
Minimum Operating Ambient Temperature:

5 °C (41 °F)

Maximum Operating Ambient Temperature:

30 °C (86 °F)

Page 11

Generator Overview
General Dimensions
914 mm / 36”

787 mm / 31”

762 mm / 30"
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING

Page 12

MS Bench G SCI 1 Rear Connections

IEC Power Inlet
Power Switch

Drain Outlet
Curtain Outlet
Source Outlet
Exhaust Outlet

MS Bench G SCI 1 Unit Controls
Power LED (Green)
Service/High Duty LED (Yellow)

Page 13

MS Bench SCI 1 Rear Connections

IEC Power Inlet
Power Switch

MS Bench SCI 1 Unit Controls
Fan Power LED

Page 14

Drain Connection (MS Bench G SCI 1 Only)
Fit the 6mm push fit fitting to the drain port located on the output panel. Tighten
using a 16mm or 5/8” spanner. Use the 6mm tubing to connect this to a suitable drain
connection or container. It should be noted that the Generator can expel a considerable
amount of water from this (dependant on ambient humidity).

CAUTION

Page 15

If a container is used it should be emptied at regular intervals. The
container must NOT have an air tight seal as water and air are expelled
at pressure.

Electrical Connection
Connect the generator to an appropriate single-phase supply, refer to the generator
serial plate for input specification and ensure your supply matches the requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.

WARNING

This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.

EARTH/GROUND (E):-

Green & Yellow or

Green

LIVE (L):-

Brown

or

Black

Neutral (N):-

Blue

or

White



Electrical requirements are 220 - 240v ±10% (MS Bench G SCI 1) & 100 - 240v ±10% (MS
Bench SCI 1). Running at voltages outwith this is not recommended. Extended periods
at extremes can have a detrimental effect on the operation and life of the generator.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING

Page 16

Start-Up Sequence (MS Bench G SCI 1 Only)

CAUTION

Before the Generator is connected to the application, the Generator
should be operated in isolation (i.e. not connected to the application)
for thirty minutes. This is to ensure any impurities present are purged
from the system. Failure to do this may harm the application.

Once this is done, the Generator can be re-connected to the mains and switched on.
Pressure will start to build in the internal storage tanks. Once this pressure is reached,
the compressors will continue to run until the internal tank upper pressure limit is
reached and the compressor has run for a period of at least two minutes.
The compressors will then rest until the internal tank lower pressure limits is reached.
Once this limit is reached the compressors will switch back on again. This compressor
cycling is normal and will continue throughout the operation of the Generator.
When the system has been operated for a period of 30 minutes, all the internal pipework and storage tanks will have been purged with Nitrogen.
The Generator is now purged and the tubes can be connected at the rear of the unit.

Page 17

Connecting to the application (MS Bench G SCI 1 Only)
Once the initial purge run of 30 minutes has completed, and the generator has been
running for 1 hour, (required to reach purity for the application) it is ready to be
connected to the application(s)
The pressure in the internal storage tanks must be allowed to dissipate
before connecting the generator to the application(s)
CAUTION

Attach the ¼” compression fittings to the outlets of the generator. Using the ¼” tubing
supplied, connect the outlets of the generator to the inlets on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.

CAUTION

Once the tubing is connected to the application, please ensure that it
is thoroughly checked for being leak-tight. Even the slightest leak in
the gas supply between the generator and the application can lead to
a reduction in efficiency.

Tubing Lengths

CAUTION



The diameter of the tubing which will be connected to the gas outlets
is important and is determined by the length of tubing required. Failure
to follow these recommendations could lead to pressure between
generator and application.

< 10 meters:

Use ¼”/3⁄16” (¼” O/D, 3⁄16” I/D) P.T.F.E. tubing.


> 10 - 40 meters: Use 3⁄8”/ 5⁄16” (3⁄8” O/D, 5⁄16” I/D). Tubing and fittings not supplied
in the fittings kit.

> 40 metres:


Please contact Peak Scientific with the relevant distance and
we will calculate the flow resistance and the tubing size required.

A combination of ¼”/3⁄16” and 3⁄8”/5⁄16” tubing may be used to ensure that there is no large
diameter tubing within the lab (i.e. for the first 20 meters from the generator use 3⁄8”/5⁄16”
and the final 10 meters to the application use ¼”/3⁄16” tubing). Keep the connections and
bends to a minimum.

Page 18

Normal Operation (MS Bench G SCI 1 Only)
The generator is designed specifically to minimize operator involvement. Given that the
system is installed as described in earlier sections and is serviced in accordance with
the specified maintenance recommendations (see Service Requirements), then it should
simply be a matter of turning the generator on when it is required.
The generator will automatically produce the factory set flow and pressure as detailed
in the Technical Specifications.

On Demand Gas (MS Bench G SCI 1 Only)
When the application requires a gas flow, the generator will produce gas on demand.
The generator will continue to supply gas whilst the application is in a ‘Standby’ mode.
As soon as a demand is detected from the application, the generator will match the
applications requirements.

Generator Cycling (MS Bench G SCI 1 Only)
The generator is designed to operate in either a cyclic mode or run continuously, this
will be dependent on the output flow.
The cyclic rate of the compressors will depend on the gas demand of the application;
the generator will cycle when the upper tank pressure set limits are achieved within a
3-minute time period.
When the application demands increase, the compressor duty will be higher and the
rest period of the compressor cycle will be shorter. If operated at maximum flow, the
generator will run continuously to maintain the upper tank pressures. Another cause
for continuous operation would be if the generator is installed at high altitude, in an
extreme environment or subjected to low voltage supply. The compressors will undergo
periods where they do not cycle.

Unusual Operation (MS Bench G SCI 1 Only)
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.

Page 19

Fan Power LED (MS Bench SCI 1 Only)
The MS Bench Bench is fitted with a fan power LED on the front panel.
The LED should be illuminated when the table is supplied with power. This is indication
that the internal cooling fans are being supplied with power, and are running correctly.

Page 20

Safe Operation of Pump Tray (MS Bench SCI 1 Only)
For safe installation or removal of roughing pumps, we recommend use of the sliding
function of the pump tray, as shown below.
This should only be attempted once the Mass Spec has been
positioned on top of the table.
WARNING

Remove the front panel, ensuring that the fan power is disconnected.
Depress both yellow tabs simultaneously and pull tray outwards until it clicks into
position.
Once pumps have been installed, ensure correct reconnection of the fan power lead.

Page 21

Service Requirements (MS Bench G SCI 1 Only)
Service Schedule
Purchase Interval

Component

Visit

12 Months

Genius 1024 Annual Maintenance Kit*

www.peakscientific.com/ordering

* Please quote your generator serial number when ordering your Annual Maintenance
Kit to ensure selection of the appropriate kit for your generator.

Page 22

Service Indication (MS Bench G SCI 1 Only)
The generator has the following Service Indication Stages:-

Stage 1
Once either compressor requires a service the LED indicator (yellow) on the front of the
generator will illuminate.
This is to make the user aware that a service of the generator is due and should be
planned at the earliest convenience. The generator will continue to operate as normal
with the LED on.  

Stage 2
If the service is not completed the generator will continue to run. After 2 weeks, the
service LED indicator (yellow) will start to flash.
This is to make the user aware that the service of the generator is now overdue and
must be completed immediately to ensure the continuous trouble free operation of the
generator.

Service Indication Reset
Once the service has been completed the Service Indication LED can be reset in the
main control PCB. This will be performed by the Peak Service Engineer or trained
service representative that completes the service operation.

Page 23

Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.

[Peak Protected] can provide…

Installation

A dedicated Peak engineer will visit
your lab to install and setup your
generator

IQ/OQ

Certified assurance for applications
requiring documented qualification

Complete plan

Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance

Spares

Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime

Premium Protected

Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance

Technical Support Hotline

Around the clock support by phone
or online with our global technical
helpdesk

To find out more about protecting your investment visit: www.peakscientific.com/protected

Page 24

Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the generator.

CAUTION

Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.

CAUTION

Page 25

Troubleshooting (MS Bench G SCI 1 Only)
Problem

Possible Solution
•

Ensure power cord is plugged into the Generator
and that the power socket is turned on.

•

Check the fuse in the power cord plug.

•

Contact your service provider.

Compressors are running but pressure is not building.

•

Contact your service provider.

The mass spec is reporting low pressure.

•

Contact your service provider.

•

A compressor(s) due for service. Contact your
service provider.

•

Refer to Service Indication section of this manual for
further information.

•

A compressor(s) has not cycled for 8 hours. Refer to
the Run Time Alarm section of this manual.

•

Ensure ambient temperatures are within the
specification.

•

Turn the Generator off and on to reset the alarm and
contact your service provider.

The Generator will not switch on and the power switch
does not illuminate.

Yellow “SERVICE” LED on front panel is on constantly.

Yellow “SERVICE” LED on front panel is flashing.

Page 26

Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.

Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.

Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?

Yes

What gas requirements do
you have in your lab?

Hydrogen

No

Nitrogen

Zero Air

Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only






Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 27

Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com

Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
MS Bench (All Models) User Manual Rev 1 RSID 3023 EN 31/05/19