Extracted text
PG14L & PG28L
Purge gas generator for FT-IR
Part Number : 99-9999
Service Kit : 99-9999
Your local gas generation partner
Description
The PG14L and PG28L gas generators from Peak Scientific are
suitable for laboratory detectors requiring CO2 free gas such as
FT-IR analysis.
Using Pressure Swing Adsorption technology to remove organic
contaminants and moisture from air, these generators provide
a consistent flow of air to ensure little background noise in
laboratory analysis.
Requiring minimal maintenance, the PG14L and PG28L do require
an external source of air to operate, Peak Scientific will be able to
help with recommending a suitable air source if your laboratory
does not have one.
Applications
)
LD-IR & FT-IR Instruments
)
FT-IR microscopes
)
Laboratory applications requiring CO2 free air
Key Features
Purge Gas (PG) generators from Peak Scientific offer gas
which is free of organic contaminants such as carbon
dioxide
Ideal for commercial labs with a high throughput rate of
samples
High performance filters and PSA ensure maximum
reduction of moisture (< -70ºC dew point).
Few moving parts means minimum maintenance
requirements and 24/7 performance
Position generator where required, wall mountable for
small labs
Most cost effective than any other gas supply as the FT-IR
detector needs a constant flow of gas
Completely silent in operation
12 month full comprehensive on-site warranty
Copyright © Peak Scientific Instruments Ltd 2024 — PG 14-28 Data sheet Rev 5 RSID 2297 EN —20/12/2024
Technical Specifications
PG14L
PG28L
Max Flow Rate
up to 14 L/min
up to 28 L/min
Max Pressure
up to 100 psi /6.9 bar
up to 100 psi /6.9 bar
Gas Outlets
1/4” BSP Female
1/4” BSP Female
Dewpoint
-70 °C / -94 °F
-70 °C / -94 °F
Particles
< 0.01 µm
< 0.01 µm
100 psi/6.9 bar - 120 psi/8.3 bar
100 psi/6.9 bar - 120 psi/8.3 bar
Operating Temperature
5-25 °C \ 41-77 °F
5-25 °C \ 41-77 °F
Electrical Requirements
Universal 120 / 240VAC / 50 / 60 Hz / 0.8 A
Universal 120 / 240 VAC / 50 / 60Hz / 0.8 A
660 x 459 x 162 mm
26.0 x 18.1 x 6.4’’
910 x 459 x 162 mm
35.8 x 18.1 x 6.4’’
14.4kg / 31.8lbs
18.6kg / 41.1lbs
45-1010
45-1060
65-0555 (230v) / 65-1555 (120v)
N/A
Min/Max Air Inlet Pressure
Generator Dimensions (HxWxD)
Generator Weight
Ordering Information
Part Number
Precision air compressor (suitable
for PG14L only)
External Regulator & Gauge Kit for
FT-IR/LD-IR Applications
3306030
Annual Service
FPPM Maintenance Plan
Complete Maintenance Plan
www.peakscientific.com/ordering
Supported Maintenance Plan
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in laboratories around the
world, and come backed by a 12 month warranty. Beyond this period however you can ensure that your investment continues
to be [Protected] by our comprehensive generator care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance whilst giving you
the reassurance of instant access to worldwide technical support and priority on-site response in the untimely event of a
breakdown.
Peak Scientific UK
Tel: +44 (0)141 812 8100
Fax: +44 (0)141 812 8200
Peak Scientific North America
Tel: +1 866 647 1649
Fax: +1 978 608 9503
Product Certifications
Peak Scientific’s Quality
Management System
conforms to: ISO:9001:2008
For a full list of our worldwide
office locations, please visit:
Web: www.peakscientific.com
Email: marketing@peakscientific.com
Copyright 2024 Peak Scientific - Rev - PG 14-28 Data sheet Rev 5 RSID 2297 EN - 20/12/24