Infinity 1045
User Manual
Copyright © Peak Scientific Instruments Ltd 2021 — Infinity 1045 User Manual Rev 2 RSID 179 EN —12/10/2021
Contents
Change History
3
How to use this Manual
3
Introduction
4
Warranties and Liabilities
5
Warranty & Liability Coverage
5
Safety Notices
7
Certificate of Compliance
8
Technical Specification
9
Unpacking
10
Fittings Kit Contents
11
Installation
12
Generator Overview
13
Air Connection
14
Commissioning
15
Service Requirements
16
Peak Protected
17
Cleaning
18
Safely Isolated Condition
18
Troubleshooting
19
Notes
21
Page 2
Change History
Rev
6
Comment
Declarations Update
Name
D.Lai
Date
18/12/2020
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The Infinity 1045 compressorless nitrogen generator has been specifically designed to
supply the Bruker EVOQ Triple Quadrupole Mass Spectrometer with carrier nitrogen
gas and a separate dry air output.
The Infinity Series are compressor-free gas generators which require a clean source of
in-house air to operate effectively. With very few moving parts, these generators are
effectively silent in operation and need only minimal care to ensure continued high
performance and extended product life
Page 4
Warranties and Liabilities
Warranty & Liability Coverage
1.
Peak warrants that, subject to the provisions in this statement, purchased Peak generators, whether
purchased directly from Peak or indirectly via an approved, certified and trained distributor or partner
(referred to hereafter as a “Peak Partner”) will comply in all material respects with any specifications
referred to in your customer order confirmation and, subject to installation and operational guidelines
being followed as described in applicable product manuals, shall be free from any defects in quality
of materials or workmanship for a period of one year from the date of installation, provided this takes
place within 3 months of factory dispatch.
2. Where the purchased generator is from the Precision Hydrogen series, Peak further warrants that,
subject to installation and operational guidelines being followed as described in applicable product
manuals, the hydrogen cell shall be free from any defects in quality of materials or workmanship for a
total period of three years (inclusive of warranty period specified in clause 1) from date of installation,
provided this takes place within 3 months of factory dispatch.
3. Where the purchased generator is from the i-Flow 6000 series, Peak further warrants that, subject
to installation and operational guidelines being followed as described in applicable product manuals,
the generator shall be free from any defects in quality of materials or workmanship for a total period
of two years (inclusive of warranty period specified in clause 1) from the date of installation, provided
this takes place within 3 months of factory dispatch and the following provisions have also been met:
a. you must purchase a service plan, ensuring the generator is serviced by Peak or a Peak Partner on
or before the end of the first 12 months of your ownership, and serviced at least once during each
subsequent 12 month period thereafter;
b. the generator (and any associated equipment)
must have been commissioned by Peak or a Peak Partner;
c. the feed air or inlet air supply to the generator must comply with ISO 8573-1:2010 Class 1.4.1 at all
times;
d. your air compressor, dryer, filtration and oil removal systems must be deemed suitable for use
by Peak or a Peak Partner, and must be changed and serviced regularly, in line with the equipment
manufacturer’s recommended guidelines; and
e. any generator failure or fault that is deemed to have been caused by the failure of any upstream
equipment, component, part or system (such as air compressor, air treatment or filtration) will be
excluded from the warranty described herein.
4. Peak also warrants that any replacement parts whether purchased (directly from Peak, or via a Peak
Partner) or supplied as part of any remedial action undertaken in line with the provisions of clauses 12
and 13, shall be free from any defects in quality of materials or workmanship for a period of 180 days
from the date of factory dispatch, provided its installation is performed by Peak or a Peak Partner.
5. This warranty does not exclude Peak’s liability in respect of any claim for death or personal injury
to any person, in so far as such can be attributed to negligence or breach of duty of care directly
resulting from failure of Peak to comply with the provisions in clauses 1, 2, 3 & 4.
Exclusions & Limitations
6. This warranty does not cover:
a. damage, deterioration or malfunction resulting from an alteration or modification to a generator
which has not been carried out by Peak or a Peak Partner;
b. damage, deterioration or malfunction resulting from what Peak reasonably believes to be abuse, or
misuse of a generator by you or any third party;
c. liability for accident or neglect (other than pursuant to clause 5);
d. maintenance or repairs which have not been carried out by Peak or a Peak Partner;
e. operation of a generator or exposure of a generator to environmental conditions that fall out-with
operational guidelines as specified in the applicable product user manual; and
f. lightning, power surges or any other acts of God or nature.
7.
This warranty is non-transferrable. Only the original owner of the generator may benefit from the
terms within this statement.
Page 5
8. Peak shall not be liable in respect of any claim made for costs, damages, losses or expenses (whether
consequential, direct, indirect or otherwise) or in any respect howsoever arising including, but not
limited to, liability from accident or negligence (other than pursuant to clause 5) that may be suffered
by you or any third party.
9. No person or entity is authorised to change the terms and conditions outlined in this warranty
statement in any respect, or to create any additional obligations or liabilities for any party involved.
10. This warranty statement supersedes any and all prior warranty agreements between the parties and
constitutes the complete, final and exclusive understanding of the parties with respect to the subject
matter. All prior negotiations, representations, or promises, whether oral or written, of either party
shall be deemed to have been merged herein.
11. If any part of this warranty statement is invalidated, for whatever reason, such part will be deleted
and the rest shall remain unaffected, continuing to be in full force and effect.
Delivery of Warranty Service
12. Subject to clause 13, and:
a. Peak being notified by you, within the duration of the applicable warranty period, of any defect
that you think is subject to any warranty valid under clauses 1, 2, 3 or 4; and
b. Peak being permitted to inspect the generators, parts and their installation (along with any relevant
packaging)
Peak shall at its option repair or replace defective generators or parts (including, if necessary, any
moving parts and irrespective of runtime). No additional charges will apply, for parts or delivery and,
where applicable, labour or travel. Peak will endeavour to deliver this service within 3 working days of
your notification.
13. Where, in Peak’s reasonable opinion, a defect is subject to an exclusion described in clause 6, Peak
reserves the right to charge for parts or delivery and, where applicable, you may also be charged by
Peak for call out, labour or travel in respect of any repair or replacement which you authorize Peak to
carry out.
Page 6
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING
Page 7
EU Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen & Air Generator
Models: Infinity 1045
To which this declaration relates, is in conformity with the following applicable EU
Directives, harmonized standards, and other normative requirements.
•
Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use.
•
Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for Measurement, Control and Laboratory Use
– EMC Requirements. (Class A)
•
Restriction on the use of certain hazardous substances in electronic equipment
(RoHS) Directive 2011/65/EU as amended by EU 2015/863.
•
FCC 47 CFR Part 15 class A
Unintentional radiators; Conducted and Radiated emissions limits.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
UK Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen & Air Generator
Models: Infinity 1045
To which this declaration relates, is in conformity with the following applicable UK
Statutory Instruments, Standards and other normative requirements.
•
The Electrical Equipment (Safety) Regulations 2016 (SI 2016 / 1101) as amended.
BS61010-1:2010 Safety Requirements for Electrical Equipment for Measurement
Control and Laboratory Use.
•
The Electromagnetic Compatibility Regulations 2016 (SI 2016 / 1091) as amended.
BS61326-1:2013 Electrical Equipment for Measurement , Control and Laboratory Use
– EMC Requirements.
•
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012 (SI 2012 / 3032) as amended.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
WEEE Compliance Statement
The Waste Electrical and Electronic Equipment (WEEE) Regulations SI 2013 No 3113
and or the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU
apply to all electrical and electronic equipment placed on the market in the UK and EU
covered by the scope of regulations which can be found in the Government Guidance
Notes (PDF) produced by the Department for Business Innovation and skills for the UK
and here for Europe.
All PEAK products that are subject to the WEEE directive are compliant with the
WEEE marking requirement. Such products are marked with the “crossed-out wheelie
bin” symbol (shown below) in accordance with European standard EN50419. All
old electrical equipment can be recycled. Please do not dispose of any electrical
equipment (including those marked with this symbol) in general rubbish bins. Please
contact your dealer or distributor for clarity.
Technical Specification
Infinity 1045
Environment
Infinity 1045
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Relative Humidity
5 °C (41°F)
35°C (95°F)
70% Non-Condensing
Inlet Conditions
Minimum Air Inlet Pressure
Maximum Air Inlet Pressure
Minimum Air Inlet Flow
Minimum Air Quality
8 bar (116 psi)
10 bar (145 psi)
L/min
ISO8573 - 1:2010 Class 1.4.1
Generator Outlets
Maximum Pressure Drop (Inlet-Outlet)
Maximum Outlet Flow (Nitrogen)
Maximum Outlet Flow (Dry Air)
Max Gas Outlet Pressure (Nitrogen)
Max Gas Outlet Pressure (Dry Air)
Start-Up Time
Particles
Pressure Gauges
8 psi
32 L/min 1.13 cfm
50 L/min 1.77 cfm
90 psi
110 psi
60 mins
<0.01μm
2
General
Dimensions WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 11
730 x 424 x 250 mm (28.8 x 16.6 x 9.9'')
39 kg (86 lb)
kg ( lb)
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator to
the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful
not to discard these with the packaging.
Page 12
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. ¼” Compression Fittings
x2
2. ¼” Teflon Tubing x 8m
3. Screw N8 x 1 ½ Supa Counter Sunk
x3
4. Raw-Plug for N8 Screw (521-3087)
x3
5. Installation Guide–Infinity 1045
x1
6. User Manual CD– Infinity 1045
x1
All of the generators output ports are located on the output panel at the rear of the
unit.
Page 13
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator is affected by ambient conditions. Note should also be
taken to the proximity of Air Conditioning outlets. These can sometimes give rise to
“pockets” of air with high relative humidity. Operation of the unit within such a pocket
could adversely affect its performance. Consideration should also be given to the
air flow around the unit. It is recommended that an air gap of 75mm (3”) should be
maintained down both sides and at the rear of the unit. Please refer to the drawing
below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (35 °F)
Maximum Operating Ambient Temperature:
41 °C (95 °F)
Page 14
Generator Overview
General Dimensions
250.00 mm
330.00 mm
730.00 mm
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING
Page 15
Air Connection
Infinity 1045 Generator should be connected to a clean, dry, OIL - FREE source
of compressed air. A minimum inlet pressure of 116 psig (8 barg) is required. Any
doubts as to the suitability of your compressed air supply should be referred to Peak
Scientific or any of their authorised partners.
The generator has a Breathing Air Filter with ¼” BSPT connection to the left side
of the unit. The Compressed Air supply should be connected here. This filter will
drain moisture and is equipped with an automatic drain. The drain should be led to a
convenient place. There are 2-off ¼” BSPT Female bulkhead connections to the right
of the unit. Fittings are provided to connect these ports to the gas connections of the
Mass Spectrometer.
LC/MS with 2-off Connections (e.g. LC-TQ)
Connect the 2-off gas ports at the right side of the generator to their respective ports
on the Mass Spectrometer using the 2-off ¼” BSPT x ¼” Swagelok fittings. Use ¼”
tubing throughout.
Page 16
Commissioning
With the Generator installed (as described earlier), connect compressed air to the unit
and open the air supply to the Unit. Disconnect the Outlet connections to allow the
generator to vent to atmosphere until the unit is stabilised. The Generator has been preset in the factory to give the specified output flow-rate and pressure in line with known
requirements. Once the Membranes reach the design pressures the Generator will
stabilise and produce Nitrogen. Maximum purity will be achieved after approximately 1
hour. The generator can then be re-connected to the application.
The design of the generator is such that it can deliver up to the supplies specified in
the Pressure & Flow Settings table below. Should the demand for gasses be less than
the rated output flow at any time, the demand will be determined by the consuming
equipment. In circumstances of no demand, the Generator remains operational and will
provide Nitrogen immediately as demand resumes.
Page 17
Service Requirements
Service Schedule
Purchase Interval
Component
Visit
12 Months
Infinity 1045/1046 Annual Maintenance Kit
www.peakscientific.com/ordering
Page 18
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit
your lab to install and setup your
generator
IQ/OQ
Certified assurance for applications
requiring documented qualification
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance
Spares
Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime
Premium Protected
Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance
Technical Support Hotline
Around the clock support by phone
or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 19
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
CAUTION
Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Safely Isolated Condition
The unit is in a safely isolated condition when it is disconnected from its application and
fully de-pressurised. Directions for isolating the Generator are shown below.
WARNING
Failure to place the Generator in a safely isolated condition when
instructed to do so may lead to personal injury or injury to others and
even death.
1. Disconnect from air supply.
2. Ensure the output pressure gauge reads zero. (If gauge does not fall to zero, open
the manual ball valve at the outlet port, to allow trapped gas to escape.)
3. Disconnect from the application.
Page 20
Troubleshooting
Problem
The output pressure has dropped below its
original set point.
Page 21
Possible Solution
•
Check the pressure regulator at the back
of the unit has not been adjusted.
•
Check to see if the inlet pressure gauge
has dropped below the set air supply
pressure.
•
If so, check your air supply does not have
a pressure drop over the length of tubing
and has no leaks.
•
Contact your service provider.
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 22
Notes
Page 23
Peak Scientific gas generators define the benchmark in reliability, convenience and performance in
laboratories around the world, and come backed by a 12 month warranty. Beyond this period however
you can ensure that your investment continues to be [Protected] by our comprehensive generator
care cover.
Our world-class aftercare support packages deliver a program of scheduled preventative maintenance
whilst giving you the reassurance of instant access to worldwide technical support and priority on-site
response in the untimely event of a breakdown.
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2021 — Infinity 1045 User Manual Rev 2 RSID 179 EN —12/10/2021