Nitrogen N*10DR Nitrogen Generator User Manual UM – N*10DR – Rev 4
1
Nitrogen N*10DR Nitrogen Generator Contents
UM – N*10DR – Rev 4
Contents
Page
2 3 4 4 5 5 6 7 8 9 10 10 11 11 12 12 13 14 15 16 17
Contents Document Change History Warranty Statement Safety Notice to Users RoHS Compliance Statement WEEE Compliance Statement Technical Specification Introduction Unpacking and Installation Removal of Transit Bracket Useful Information Electrical Connection Generator Environment General Dimensions Generator Connections Generator Controls Operation Control Panel / Fault Diagnosis Additional Points to Note Routine Maintenance Maintenance Log
2
Nitrogen N*10DR Nitrogen Generator
History Document Change History
UM – N*10DR – Rev 4
Rev No.
1 2 3 4
Change
First Release Notes added for - Indoor use only & Altitude Update USA address Updated solenoid valve part #
Initials
G.L. T.G. S.K S.K
Date
16.10.06 10.09.07 25/08/2008 26/01/2009
3
Nitrogen N*10DR Nitrogen Generator
Warranties & Liabilities
1. The Company warrants that it has title to the Goods.
UM – N*10DR – Rev 4
2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory. 3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act 1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause. 4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must:4.1. reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts; and 4.2.allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company. 5. Subject to the Company being notified of any defect as is referred to in sub-clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof). 6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof. 7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
These instructions must be read thoroughly and understood before installation and operation of your Peak N*10DR generator. Use of the generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations.
SAFETY NOTICE TO USERS
4
Nitrogen N*10DR Nitrogen Generator
Declaration of Conformity
UM – N*10DR – Rev 4
5
Nitrogen N*10DR Nitrogen Generator
Technical Specification
UM – N*10DR – Rev 4
Generator Environment
Minimum operating ambient temperature Maximum operating ambient temperature Maximum relative humidity Maximum Altitude 5° (41° C F) 25° (77° C F) 70% 2000 Meters
Nitrogen Outlet
Maximum flow Maximum Pressure Particles Outlet 1/4" BSP Pressure Gauges Pthalates Suspended Liquids 30 l/min (1.1 CFM) 6.89 bar (100 psi) < 0.01 µm 1 1 None None
Electrical Requirements
230V (50/60 Hz) Compressor fuse on Control PCB PLC / fan fuse on Control PCB Electrical connection Noise Level 8.4 amps 8.0 amps 2.0 amps
Single phase power cord 57dBA @ 1m
Dimensions in cm (inches) W x D x H Weight Shipping weight
General
94.5 x 59 x 70.7 (37 x 23 x 28) 145 kg (319 lb) 190 kg (419lb)
6
Nitrogen N*10DR Nitrogen Generator
Introduction
UM – N*10DR – Rev 4
Welcome to the User Manual for the Peak Scientific N*10DR Nitrogen Generator. Enclosed in this manual you will find the information required to ensure that your generator is operated & serviced according to our recommended guidelines which will prepare you for long and trouble free Nitrogen generation. Please review each of the following sections carefully and ensure that the maintenance log at the rear of this manual is updated for future reference. Thank you for selecting Peak Scientific to meet your gas generation needs, and should you require any further assistance or support please do not hesitate to contact us at the addresses displayed on the front cover of this manual.
7
Nitrogen N*10DR Nitrogen Generator
Unpacking and Installation
UM – N*10DR – Rev 4
Unpacking and Installation Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ label for signs of rough handling prior to unpacking -
ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND PEAK SCIENTIFIC OR THE DISTRIBUTOR FROM WHERE THE UNIT WAS PURCHASED. Follow the un-packing instructions posted on the side of the crate. It will require two people to lift the crate clear and to manoeuvre the generator onto the floor.
8
Nitrogen N*10DR Nitrogen Generator
Installation Removal of Transit Bracket Remove Rear Panel as follows:At the rear of the unit, turn the key in the lock anti-clockwise to unlock the rear panel. Hold the rear panel by the two handles and slowly swivel the bottom of it out and lower it until the top comes free from the chassis. Disconnect the plug at the LHS so that the rear panel can be completely removed.
UM – N*10DR – Rev 4
Disassemble Transit Bracket as follows – • Remove Screw 1 holding two pieces of transit bracket together. • Remove Screws 2,3,4,5,holding the Transit Bracket to the Compressors • Remove Screws 6,7 holding the Bracket to the internal side panels. • Remove transit bracket from unit.
6
2
3
1
4
5(Hidden)
7(Hidden)
Re-fit the rear panel in reverse order of removal. Important Note: The Transit Bracket must be removed prior to switching the unit on. Failure to do so will result in damage to the equipment. The Generator can now be moved to its final location on the castors provided. Note: - Included with the Generator is a pack containing hex key, manuals, fittings. Be careful not to discard these with the packing. Please save the product packaging and Transit Bracket for storage or future shipment of the compressor.
9
Nitrogen N*10DR Nitrogen Generator
Installation Useful Information
UM – N*10DR – Rev 4
The diameter of the tubing which will be connected to the gas outlets is important and is determined by the length of tubing required. Failure to follow these recommendations could lead to accelerated compressor wear. < 10 meters. > 10 - 40 meters. > 40 metres. Use 6/4 (6mm O/D, 4mm I/D) P.T.F.E. tubing. Use 10/8 (10mm O/D, 8mm I/D). Please contact Peak with the relevant distance and we will calculate the flow resistance and the tubing size required.
A combination of 6/4 and 10/8 tubing may be used to ensure that there is no large diameter tubing within the lab (i.e. for the first 10 meters use 6/4 tubing and the final 20 meters use 10/8). Keep the connections and bends to a minimum. The imperial equivalents are:6/4 = 1/4” O/D, 3/16” I/D. 10/8= 3/8” O/D, 5/16” I/D.
Electrical Connection Important Electrical Notice This unit is classified as SAFETY CLASS 1 equipment. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency. The Electrical requirement is:EARTH/GROUND (E):LIVE (L):Neutral (N):230V AC (50/60 Hz), 8.4 Amps or or or Green Black White
Green & Yellow Brown Blue
Connect the compressor to a single-phase supply using the power cord provided.
10
Nitrogen N*10DR Nitrogen Generator
Generator Environment
UM – N*10DR – Rev 4
The Generator is designed for indoor use only. The unit should be installed adjacent to the Mass Spectrometer it is supplying. If this is not convenient then the unit can be sited elsewhere, however, consideration should be made of the lengths of pipe runs as pressure drops can result from extended runs of pipe. Please see the “Useful Information” section for further details. Performance of the generator (like all sophisticated equipment) is affected by ambient conditions. Note should also be taken to the proximity of Air Conditioning outlets. These can sometimes give rise to “pockets” of air with high relative humidity. Operation of the unit within such a pocket could adversely affect its performance. Consideration should also be given to the air flow around the unit. It is recommended that an air gap of 75mm (3”) should be maintained down both sides, at the rear and across the top of the unit. Please refer to the drawing below for the general dimensions of the unit. Maximum Ambient Conditions: 25° (dry bulb) 70%RH (Max) C
General Dimensions
Unit must always be placed on a level surface. Failure to do so will effect the performance of the Generator.
11
Nitrogen N*10DR Nitrogen Generator
Connections
UM – N*10DR – Rev 4
Connections to Rear of Unit
Serial Plate
Power Cord (230V ONLY) 04-1038 (EU) 04-1029 (US) 04-1030 (UK) 9-way D-type connector for PC Comms. Generator Controls Rear panel lock
Water outlet 1/4” BSPP female Nitrogen outlet 1/4” BSPP female
Front panel lock Pressure Gauge Service light
On/Off Switch Alarm light Keypad
12
Nitrogen N*10DR Nitrogen Generator
Operation
UM – N*10DR – Rev 4
The N*10DR generator is designed specifically to minimize operator involvement. Given that the system is installed as described in earlier sections and is serviced in accordance with the following maintenance recommendations then it should simply be a matter of turning the generator on. The generator will automatically produce the factory default flow and pressure. Start Up Sequence When the unit is switched on the fans should power up and the display on the front panel should read –
Compressor Test
The system will now run through an automatic compressor check procedure. Each of the four compressors is tested in turn to ensure that it is working. The compressor check takes a minute or so and then the system will turn on fully and “PLEASE WAIT” will be displayed while pressure builds. The pressure can be monitored on the front panel gauge. Gas Generation When the system is running, the maximum number of compressors operating at any time is two. To ensure the longevity of the system, the compressors cycle in pairs, with a pair running for a three minutes while the other pair are turned off to cool, then the running pair are switched off to cool and the idle pair are switched on. The display will read Peak Scientific Mode: APCI
*Note: The outlet should be vented to atmosphere for at least 30 seconds to purge the system of impurities. If only a small volume of gas is being drawn from the system, it will automatically change into ESI mode. The display will read Peak Scientific Mode: ESI
Whilst running in this mode, if too much gas is then being drawn from the system, it will automatically change into APCI mode to meet the increased demand. The user can manually select ESI or APCI Compressor mode by pressing the relevant button on the control panel. If ESI mode is selected while the system requires two compressors, it will automatically return to APCI mode.
ESI APCI
13
Nitrogen N*10DR Nitrogen Generator
Operation
UM – N*10DR – Rev 4
Control Panel / Fault Diagnosis
If there is a problem with the system, the user will be informed by one of the two available indicators SERVICE : Audible buzzer and light on the front panel ALARM : Audible buzzer and light on the front panel. To stop the audible buzzer press as long as the unit has power.
ALARM RESET
on the Control Panel. Note the light will stay illuminated
Interrogating the Control Panel to find fault Diagnostic: To display the current status of each compressor, the system and the inlet filter. The compressors are numbered 1 to 4 from the right-hand side looking from the rear and are identified by holes in the panel they are sitting on.
MENU
Mode Diagnostic
ENTER
Compressor 1: OK Compressor 2: OK
Compressor 3: OK Compressor 4: OK
System: OK Inlet filter: OK
To exit.
MENU ENTER
Operating hours: To display the current operating hours of each compressor / membrane in hours and minutes.
MENU
Diagnostic Operating Hours
ENTER
Comp 1: 29Hr 24Min Comp 2: 27Hr 30Min
Comp 3: 29Hr 24Min Comp 4: 27Hr 30Min
Memb 1: 29Hr 24Min Memb 2: 27Hr 30Min
To exit.
MENU ENTER
System hours: To display the time since the last service and the total system operating time in days.
MENU
Operating Hours System Hours
ENTER
Time since last service: 21 days
Total system run time: 30 days
To exit.
MENU ENTER
To display the software revision.
MENU
System Hours Software Rev. V1.55
To exit.
MENU ENTER
Mode: To display the current operating mode.
MENU
Mode Diagnostic
ENTER
Peak Scientific Mode: ESI
14
Nitrogen N*10DR Nitrogen Generator
Additional Points
UM – N*10DR – Rev 4
Additional Points to Note Pressure / Flow Adjustment The system is configured in the factory to give standard Outlet Pressure and Flow Rate (see ‘Technical Specification’ page 6). These settings should NEVER require adjustment during normal operation. During service/fault diagnosis the settings can be changed by adjusting the pressure regulator and flow controller situated on the Electrical Panel (see page 18). Both will be in the locked position on receipt of the unit. To unlock and reset the following steps should be completed – Pressure Regulator •Pull adjustment knob upwards away from the bracket to release the lock. •Rotate to increase/decrease pressure. •Push adjustment knob downwards towards the bracket to lock. Flow Controller •Rotate knurled locking nut at base of adjustment knob to release lock. •Rotate to increase/decrease flow. •Rotate knurled locking nut at base of adjustment knob to lock. Unusual Operation If at any time the generator begins to emit excessive noise or vibration, then it should be switched off and you should contact your local representative or Peak Scientific as soon as possible. System Drain Please ensure that the drain port at the rear of the generator is led to a suitable connection or container make sure the container is not sealed tight. It should be noted that the generator will expel considerable amounts of water from this port. If a container is used it should be emptied at regular intervals. Service Ensure that the generator is serviced in accordance with the maintenance recommendations.
15
Nitrogen N*10DR Nitrogen Generator
Maintenance
UM – N*10DR – Rev 4
Routine Maintenance
Servicing and/or repair of the Generator should only be undertaken by a TECHNICALLY COMPETENT PERSON, with the generator in its safely isolated condition.
WARNING
Definition: The unit is in a Safely Isolated Condition when it is disconnected from its application, fully de-pressurised and isolated from the Electrical Supply. Directions for isolating the generator are shown below. Isolating the Generator: a) Switch off the unit. b) Unplug the generator from the mains supply and remove the power cord from the rear of the unit. c) Ensure the pressure gauge on the front panel reads zero. (If gauge does not fall to zero, loosen outlet fitting slightly to allow trapped gas to escape.) d) Disconnect from the application. e) Important Note: Allow 10 minutes for unit to fully de-pressurise. After this time pull out the ‘Quick Release Ring’ on the Pressure Relief valve situated on the tank. Ensure tank is fully depressurised before releasing the ring. Maintenance Schedule
SERVICE INTERVAL COMPONENT
ELIMINISER FILTER (x1) 1 YEAR COALESCER FILTER (x1) FLOAT ASSEMBLY VALVE (x2) RAC Filter (x1) 1 YEAR OR ON SYSTEM ALARM ANNUAL SERVICE KIT INLET FILTER (x1) INCLUDING ABOVE SOLENOID VALVE 2/2 G1/8" N.C 2.4mm (x1) 2 YEARS SOLENOID VALVE 2/2 G1/8M N.O 1.6mm (x2) COMPRESSOR ASSEMBLY (x4) ALTERNATIVE TO COMPRESSOR ASSEMBLY COMPRESSOR RE-FIT (x4)*
SAFELY ISOLATED CONDITION
PART NO.
02-4366 02-4364 02-4464 00-4425 02-1054 (08-4714) 02-4692 02-6000 08-8069 06-5529 On Request
*Compressors can be re-fitted as an alternative to replacement up to a maximum of 3 times, this is a more cost effective solution, however a degree of technical expertise is required and can be time consuming.
16
Nitrogen N*10DR Nitrogen Generator
Maintenance Log
UM – N*10DR – Rev 4
Maintenance Log for Serial Number _____________
Work Done Remarks Name Date
17