<p>Precision 500 230v Hydrogen Generator User Manual
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Contents
Change History How to use this Manual Introduction Warranties and Liabilities Safety Notices Symbols Safety Notice to Users Declaration of Conformity Environmental Declaration Technical Specification Environment Generator Outlets Electrical Requirements General Principle of Operation General Safety Generator Safety Features Unpacking & Contents Unpacking Instructions Installation Unpacking General Dimensions Unit Controls Rear Connections Tubing H2 Ventilation Fittings Kit Automatic Filling Electrical Connection Start-up Sequence Liquid Leak Detectors H2 Detector Option Gas Traps Expanding Capacity Rear Connections Supply Line Tubing Switching the unit off Cleaning Replacement of Deioniser Resin Column Refilling the Deioniser Resin Column Health and Safety Data- Deioniser Resin IMPORT ANT DOCUMENTS Service Requirements Trouble Shooting 3 3 4 5 6 6 6 8 9 10 10 10 10 10 11 12 13 14 14 15 16 17 18 18 19 19 19 20 21 22 22 23 24 25 25 26 27 27 28 29 30 32 33 34
Page 2
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Change History
Rev. 1 Comment Altered D.O.C. Name Liam Couttie Date 17/09/12
How to use this Manual
This manual is intended for end users and has been written so that it can either be read as a step by step guide to installation and usage or as a reference document where you can skip to the relevant information. Users of a hard copy version can refer to the contents page to find the relevant information. Users of the soft copy version can use the hyperlinks from the contents page as well as the hyperlinks between sections. Please review each of the following sections carefully. Thank you for selecting Peak Scientific to meet your Gas Generation needs, and should you require any further assistance or support please do not hesitate to contact Peak Scientific or Peak Partner from which you purchased your Generator.
Page 3
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Introduction
The Precision 500 has been developed to cater for the requirements of GCMS instruments. This model provides a source of High Purity Hydrogen gas. Other features include: • • • • Touch screen interface >99.9999% purity Robust dryer technology Expandable; Inter-connectable generators for increased capacity.
With the Precision 500 based on proven technology, it produces Hydrogen by electrolysis using a Proton exchange membrane (PEM) cell. Moisture removal is achieved by a pressure swing adsorption process, using molecular sieve. To ensure this Generator model meets our high expectations with regards to reliability and performance, we have tested this model extensively at our manufacturing plant to ensure reliability and longevity of the system.
Page 4
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Warranties and Liabilities
1. The Company warrants that it has title to the Goods. 2. Subject to the provisions of this clause the Company warrants that the Goods shall comply in all material respects with any specification referred to in the Order Confirmation (as the same may be amended) and shall, subject thereto, be free from defects in material and workmanship for the lesser of a period of twelve months from the date of delivery or thirteen months from the date of dispatch from the factory. 3. Save as provided in this clause and except where the Goods are sold to a person dealing as a consumer (within the meaning of the Unfair Contract Terms Act 1977) all warranties, conditions or other terms implied by statute or common law are hereby expressly excluded save to the extent they may not be lawfully excluded. When the Goods are sold to a consumer within the meaning of the Unfair Contract Terms Act 1977 their statutory rights are not affected by the provisions of this clause. 4. In the event of the Customer making a claim in respect of any defect in terms of clause 2 hereof the Customer must. 1. Reasonably satisfy the Company that the Goods have been properly installed, commissioned, stored, serviced and used and without prejudice to the generality of the foregoing that any defect is not the direct or indirect result of lack of repair and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or incorrect spare parts 2. Allow the company to inspect the Goods and/or any installation and any relevant packaging as and when reasonably required by the Company. 5. Subject to the Company being notified of any defect as is referred to in sub-clause 2 hereof within a reasonable time of it becoming apparent and subject always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace or repair the defective Goods or refund a proportionate part of the Price. The Company shall have no further liability to the Customer (save as mentioned in sub-clause 6 hereof). 6. The Company shall be liable to indemnify the Customer in respect of any claim for death or personal injury to any person in so far as such is attributable to the negligence or breach of duty of the Company or any failure by the Company to comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable
in respect of any claim by the Customer for costs, damages, loss or expenses (whether direct, indirect, consequential or otherwise) or indemnity in any respect howsoever arising including, but not by way of limitation, liability arising in negligence (other than pursuant to clause 6 above) that may be suffered by the Customer or any third party.
Page 5
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Safety Notices
Symbols
This manual uses the following symbols to highlight specific areas important to the safe and proper use of the Generator
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause personal injury or in the worst case death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood or met. A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process or similar, which if not correctly performed or adhered to, could cause damage to the Generator or the Application. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the Generator has been removed before proceeding.
Safety Notice to Users
These instructions must be read thoroughly and understood before installation and operation of your Peak Precision 500 Hydrogen Generator. Use of the Generator in a manner not specified by Peak Scientific MAY impair the SAFETY provided by the equipment. When handling, operating or carrying out any maintenance, personnel must employ safe engineering practices and observe all relevant local health and safety requirements and regulations. The attention of UK users is drawn to the Health and Safety at Work Act 1974, and the Institute of Electrical Engineers regulations. If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment maybe impaired.
Page 6
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
ATTENTION - IMPORTANT NOTICE TO ALL USERS These instructions must be read thoroughly, section by section and understood before installation and operation of your Peak Scientific Hydrogen Generator. Please follow the appropriate safety standards for handling hydrogen gas and other gases as determined by the laws and regulations of your country. Please save the product packaging for storage or future shipment of the generator. 1. For your own personal safety and to prevent accidental damage to your generator please ensure that you read and understand our Directions for Use Manual.
2. Do not store away the generator when not in operation. The generator should be powered up at least twice a month and a small flow taken for 1 hour. It is recommended that you replace the water monthly to prevent biological growth developing in the tank and tubing. Failure to do this will greatly reduce the life of the generator and render the warranty void. 3. Do not place the external water bottle above the height of the generator. The bottle may be placed at the same level or below the generator. 4. Use only high grade de-ionised water (1µS/cm or better). Do not attempt to connect pressurised water to the generator. 5. It is recommended that you operate the generator at the lowest pressure permissible for your particular application. Do not automatically set the pressure to maximum if not required. 6. Failure to maintain the Deioniser Column as recommended will reduce the operational life of your generator and render its warranty void. 7. Do not attach any additional tanks or gas cylinders to your generator without first consulting Peak Scientific. 8. Ensure your entire installation, both pipe work and analysers, are completely gas tight. Even minor leaks will affect the operation of your generator and generally reduce its operational life. 9. Has your generator been correctly sized for your application? If in doubt, consult your dealer or Peak Scientific. 10. Ensure the ambient temperature inside your laboratory does not rise above 25°C/77°F. Extremes of operating temperature will reduce cell life.
Page 7
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Declaration of Conformity
We of declare that: Peak Scientific Instruments Ltd. Fountain Crescent, Inchinnan. Renfrewshire PA4 9RE Hydrogen Gas Generator Precision 500
Equipment Model
To which this declaration relates, is in conformity with the applicable EC Directives, harmonized standards, and other normative requirements.
•
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use. Part 1: General requirements.
Low Voltage Directive 2006/95/EC
•
EN 61326-1: 2006 Electrical Equipment for Measurement, Control and Laboratory Use – EMC Requirements. Part 1: General requirements.
Electromagnetic Compatibility Directive 2004/108/EC
All evaluation, testing and certification issued by:
York EMC Services (2007) Ltd Donibristle Industrial Park Dunfermline Fife KY11 9HZ Signed By: Name: Position: Done at: Date: Chris Pugh Engineering Director Peak Scientific Instruments Ltd, Inchinnan, Scotland. 2nd August 2012
Page 8
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Environmental Declaration
We Peak Scientific Instruments Ltd.
of Fountain Crescent, Inchinnan. Renfrewshire PA4 9RE declare that: Equipment Model Hydrogen Gas Generator Precision 500
Is fully compliant with the following Directives: 2002/96/EC WEEE (Waste of Electrical and Electronic Equipment) 2002/95/EC RoHS (Restriction of Hazardous Substances) Peak Scientific Instruments Ltd fully complies with its obligations towards the European WEEE (Waste of Electrical and Electronic Equipment) Directive 2002/96/EC. These obligations are being met within the B2B compliance group. Peak Scientific Instruments Ltd has developed all reasonable ‘due diligence’ controls to ensure that our products comply with the principles and requirements of the European RoHS (Restriction of Hazardous Substances) Directive 2002/95/EC. Similar directives in the United States and China, for example, have also been captured within this program. Where a specific certificate of compliance is required, this can be requested, on a product serial number basis, directly from Peak Scientific Instruments Ltd, by contacting us through our website on www.peakscientific.com Signed By: Name: Position: Done at: Date: Chris Pugh Engineering Director Peak Scientific Instruments Ltd, Inchinnan, Scotland. 2
nd
August 2012
Page 9
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Technical Specification
Environment
Minimum operating ambient temperature Maximum operating ambient temperature Maximum relative humidity Minimum storage temperature* Maximum storage temperature* 10°C (50°F) 25°C (77°F) 90% Non-Condensing 10 C (50 F) 25 C (77 F)
o o o o
*NOTE – When taken out of storage the Generator should be allowed to acclimatize at room temperature for a minimum of 3 hours before operation.
Generator Outlets
Max Flow Rate Purity Gas outlet Delivery pressure 500 cc/min 99.9999% 1 x 1/8” Swagelok compression fitting 0-100 psi (0-6.9 bar)
Electrical Requirements
Voltage Frequency Fuse Current Input connection Power cord (Supplied) Pollution degree Insulation category 230V 50/60 Hz T10AH250V 10 Amps C14 Connection C13 socket to local connection (10A minimum) 2 II
General
Generator dimensions in cm W x D x H Generator weight Shipping crate dimensions in cm W x D x H Shipping weight Water purity requirement 38 x 55.3 x 48.4 (14.9” x 21.7” x 19.1”) 55 kg (121 lb) 55 x 80 x 68 (21.6” x 31.4” x 26.7”) 66 kg (145 lb) 1 Mohm-cm
Page 10
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Principle of Operation
The Peak Scientific Hydrogen Generator uses a proton exchange membrane (PEM) that is extensively used in industrial and laboratory applications. The reliability of this type of membrane is assured as it was originally developed for NASA. To produce the highest purity of hydrogen that an analytical laboratory requires, Peak Scientific employs the services of a platinum catalyst in the electrode area. Electrolysis of water is the best method of producing high purity hydrogen on demand. The most important element of the generator is the electrolyser cell where the electrolysis reaction takes place. This consists of two electrodes (an anode and a cathode), which are separated by the ion exchange membrane. When a continuous voltage is applied to the electrodes on the electrolyser cell, the following reactions take place: At the anode (the positively charged electrode), the water molecules loose two electrons forming an oxygen molecule and four hydrogen ions.
Anode 2 H2O - 4e = O2 + 4 H+
The oxygen that is produced in this half of the reaction is safely vented to atmosphere at the back of the generator. The four hydrogen ions that have been produced then pass through the ion exchange membrane (attracted by the negatively charged cathode) and collect four electrons reducing them to two hydrogen molecules.
Cathode 4 H+ + 4e = 2H2
The hydrogen produced is separated from the oxygen by the ion exchange membrane, which is impervious to molecular oxygen. Deionised water is the raw material of the system and must have a purity of 1µs/cm or better to prevent damage to the membrane. Failure to use this grade of deionised water will invalidate the warranty of your generator. The water is automatically loaded (either at power-up or periodically during operation) and the rate, at which the water is consumed, depends on the amount of hydrogen used - 1cc of water produces 1244cc of hydrogen. Some electrical energy is lost as heat, which is removed from the generator by the ventilation system. Hydrogen is generated on demand. By maintaining the user’s preset pressure and using this as a reference value, the generator will produce the exact amount of hydrogen required. As soon as the demand from the consuming application stops, then no gas is generated and the generator will maintain the supply lines at the preset pressure.
Page 11
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
General Safety
The construction of the Precision 500 generator conforms to CEI / OSHA regulations as they do not store any appreciable amount of hydrogen internally. The generator only produces gas when the application places a demand on the unit. The unit also meets the European Electromagnetic Compatibility and Low Voltage Directives. The Generator is CE approved and has the CE mark attached. The maximum quantity of hydrogen stored in the generator at any one time is 400cc at a pressure of 6.9bar (100psi). This means that the Hydrogen Generator can be safely installed in your laboratory and does not classify the surrounding area as dangerous. All the Precision 500 hydrogen generators have the necessary control circuitry, both electrical and mechanical, to ensure the shutdown of the instrument in the event of a fault or faults occurring.
Page 12
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Generator Safety Features
The Precision 500 hydrogen generator has the following safety features to ensure safe and reliable operation: 1. To avoid any build up of hydrogen from leaks inside the unit; the generator features forced air ventilation. This ensures that a hydrogen / oxygen mixture cannot form inside the generator.
2. On power-up the generator self checks for internal leaks and if required it will shutdown at the end of the System Check informing the operator with a visual alarm. 3. The generator will detect any hydrogen leaks that occur within the generator, the user’s application and gas lines. In the event of a major leak the generator will automatically shutdown within 20 minutes. 4. The hydrogen pressure inside the generator should only reach a maximum of 7.5 bar (110psi) and this is controlled in two ways: a. Gas pressure is monitored and controlled electronically by a pressure sensor b. A mechanical pressure switch (Fail Safe) will electrically isolate the hydrogen cell should the internal pressure reach 7.5 bar (110psi) 5. When the generator is in alarm status the following occurs: a. The current to the cell is stopped so no hydrogen is generated therefore preventing a hazardous situation. b. The over pressure relief valve opens allowing the unit to depressurise. c. An audible and visual alarm relevant to the fault condition is displayed. The generator can sense for internal and external leaks, excessive gas demand and lack of water.
Page 13
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Unpacking & Contents
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any signs of transit damage. Any damage should be reported immediately to the carrier and Peak Scientific or the distributor from whom the unit was purchased. If possible, unpack the generator and allow to stand at room temperature for a few hours. This will allow any condensation that may have formed during transportation to evaporate from inside the generator.
Contents
Each shipping box should contain the following: Hydrogen Generator Hex Key L-Shape Long Arm 2.5mm Hose Barb ¼” Tygon Tubing (use as Water Feed Hose) Deioniser Resin Refill Kit Power Cables (EU and UK) Installation Guide– Precision 500 User Manual CD– Precision 500
Unpacking Instructions
The Precision 500 generator weights approximately 55Kg, and as such should only be lifted by two people; do not attempt to lift on your own, as you will significantly increase the chance of injury & damage to yourself and other around you. 1. Remove the 8 screws encircled in red around the bottom of the crate lid and lift upwards.
2. Next, with someone on either side of the unit, position your hands underneath the unit as shown by the arrows, ready to lift. 3. Ensuring your knees are bend and your back is straight, lift the unit to the desired location.
Page 14
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Installation
Location
The generator should be located on a flat, level surface for operation - this will ensure that the water level sensors within the unit operate correctly. Some components in the generator produce heat during operation and therefore an adequate airflow must be maintained around the generator to allow the ventilation system to perform efficiently. The rear of the generator will become warm to the touch during operation - a minimum clearance of 15 cm (6") from other bodies is recommended. Do not locate the generator in a sealed or unventilated room, or in close proximity to any ignition sources. Caution - Remove Transport Plugs Before installation, remove the blanking plugs fitted to the vents at the rear of the generator. Do not obstruct or connect the vents to any application. Failing to observe this could lead to permanent damage of the generator. Do not expose the unit to freezing conditions. Please adhere to the operating temperatures below: Minimum Operating Ambient Temperature: Maximum Operating Ambient Temperature: 10 °C (50 °F) 25 °C (77 °F)
Note: It is recommended that the unit is left to run for 12 hours before utilisation.
Page 15
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is advisable to fully inspect the unit for any sign of transit damage. Check ‘SHOCKWATCH’ label for signs of rough handling prior to un-packing –
Any damage should be reported immediately to the carrier and Peak Scientific or the Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two people to remove the unit from the shipping crate and to manoeuvre the Generator onto the bench. Please save the product packaging for storage or future shipment of the Generator. Note: Included with the Generator is a “Fittings Kit” containing mains power leads for UK and EU, and also all the required fittings. Be careful not to discard these with the packaging.
Page 16
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
General Dimensions
The Generator must always be placed on a flat, level surface. Failure to do so will affect the performance of the Generator.
Page 17
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Unit Controls
Touch Screen
Figure 1: Unit controls
Rear Connections
H2 Exhaust O2 Vent H2 Vent Master Connection (Daisy Chain) Slave Connection (Daisy Chain) H2 Detector Connection
H2 Output Fuse
Water Supply Drain Fill
Power Power Inlet Switch
Figure 2: Rear Connections
Page 18
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Tubing
For the hydrogen supply line, it is recommended that only clean grade copper or stainless steel tubing should be used. Plastic tubing material such as Teflon®, polyvinyl chloride, or Tygon® should not be used when plumbing GCs since these materials will allow air and water to diffuse into the gas lines. In addition, plastic tubing can give off organic impurities which can cause ghost peaks and baseline instability. Furthermore Peak Scientific recommend that Swagelok® fittings are used, and that the fittings are free from Teflon® tape.
H2 Ventilation
Peak Scientific recommends that both the H2 Exhaust and H2 Vent ports at the rear of the unit should be connected to an exhaust hood, or other ventilation system, to ensure safe disposal of vented gasses. Tubing sizes for the vent ports are 1/8” OD for the H2 Exhaust and 3/8” OD for the H2Vent.
Fittings Kit
Supplied in the Fittings Kit are all the fittings required to connect the Precision 500 Hydrogen Generator to the application. The contents of the Fittings Kit are as follows: 1. 2. 3. 4. 5. 6. 7. 8. Hex Key L-Shape Long Arm 2.5mm Hose Barb ¼” Tygon Tubing (use as Water Feed Hose) Deioniser Resin Refill Kit C13 Mains Cable (UK) C13 Mains Cable (Euro) Installation Guide– Precision 500 User Manual CD– Precision 500 x1 ×1 × 1m x1 ×1 ×1 ×1 ×1
All of the Generator output ports are located on the output panel at the rear of the Generator.
Page 19
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Automatic Filling
The Precision Hydrogen generator uses an automatic water filling system, whereby sensors in the unit’s internal water tanks detect when the water level reaches a minimum. This then refills the tank using a rear connection to an external DI water supply. The unit should be connected to the Water supply as shown below. Using the length of Tygon® tubing and the hose barb fitting from the fittings kit, attach the barb to the tubing. Then connect one end to the water supply and the other to the Fill port at the rear of the Precision 500.
Page 20
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Electrical Connection
Connect the Generator to a 230 volt single-phase supply using the power cord provided. If the appropriate power cord is not supplied; a new plug, rated to at least 13 amps, can be fitted by a qualified electrician.
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE EARTHED. Before connecting the unit to the mains supply, please check the information on the serial plate. The mains supply must be of the stated AC voltage and frequency.
EARTH/GROUND (E):LIVE (L):Neutral (N):-
Green & Yellow Brown Blue
or or or
Green Black White
Our electrical requirements are 230VAC nominal +/- 10%. This means that the Generator can accommodate transients between 207VAC and 253VAC. However, running continuously at voltages less than 220V is not recommended and extended periods at these extremes can have a detrimental effect on the operation and life of the Generator.
Page 21
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Start-up Sequence
CHECK: 1. Transport plugs have been removed
2. H2 Outlet is connected to application 3. Mains power is connected Turn the power switch ON. The digital display will light up, the cooling fans will rotate and the internal pumps may automatically run to load water if the internal tanks are not full. On power up the generator performs a diagnostic check including a leak test, this should take approximately six minutes. Upon successful completion of the leak test, the operator is presented with the main screen and the generator is ready to deliver hydrogen. The next screen displayed to the user will be the Home screen, where two options will be available, Start and Menu. The first time the unit is switched on the user should select the Menu option which allows the desired Unit of measure and delivery pressure to be selected. Also displayed on this screen are the Shutdown and Back options. Shutdown will depressurize the generator. This should be selected before turning the unit off using the Power switch at the rear of the unit. Finally, the Back option returns the user to the home screen where the Start option will allow the flow of hydrogen to the application. Upon selection the unit will take a few moments to adjust to the desired delivery pressure. If the unit has been used previously the Start button will initiate the supply of hydrogen to the application at the last used pressure. Note: If leak test is unsuccessful, and alarm will sound and a red failure screen will be displayed complete with a mute button. The user is prompted at this point to contact their Peak service provider.
Liquid Leak Detectors
Liquid leak detectors should not be used on the Precision 500. These solutions are easily drawn into the high purity hydrogen gas system and can cause irreparable damage to a GC column. If a liquid solution has to be used, then it should be IPA based, or a portable hydrogen detector should be used.
Page 22
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
H 2 Detector Option
The Precision 500 can be purchased along with a H2 detector unit. The detector is intended for use with a GC, it is recommended that the detector unit is placed above the height of the GC outlet. The detector can be wall mounted. This unit is plugged, from the ‘In’ port at the rear on the detector, into the ‘H2 Detector’ port at the rear of the generator, as shown below. This connection will provide the detector unit with power and allows communication between the Precision 500 and the detector unit. The detector will take samples continuously and both the detector and the Precision 500 will alarm if dangerous levels of hydrogen are detected, the Precision 500 will then shut down. Multiple detectors can be connected to one Precision 500, up to 4 in number. If this is desired, connect the first detectors ‘Out’ port to the second detectors ‘In’ port, as illustrated below. Precision 500 H2 Detector In Port Out Port H2 Detector Port
Page 23
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Gas Traps
It is recommended that Moisture, Oxygen and Hydrocarbon traps are used on the gas supply line between the Peak generator and the instrument to provide back-up contaminant protection for your instrument. The hydrogen gas produced by the Precision 500 will be sufficiently pure for these contaminant traps to last a long time, but should a leak ever occur on the gas line, the traps will ensure that any harmful contaminants are caught, therewith preventing potential damage to the instrument.
Page 24
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Expanding Capacity
The Precision 500 has an expandable feature, where the flow capacity of the unit is increased, by connecting up to 16 units in a Master-Slave format.
Rear Connections
In order to use this feature the units must be connected together via the ‘Master’ port on the first Precision 500, and the ‘Slave’ port at the rear of the next Precision 500 unit. The first Precision 500 unit is automatically defined as the Master, and before the connections are made and H2 is supplied, the desired pressure must be set on all units. This must be the same pressure for all. The user can now press Start on the Master unit to begin the flow of H2. The Master will now manage all the other units. The Data Cables (08-8902) used to connect units together can be purchased from Peak Scientific.
Master Port
Slave Port
Master
Slave
EOL
Page 25
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Supply Line Tubing
Peak scientific recommends using only clean grade copper or stainless steel for the supply tubing. The ring main and series lines should be ¼” OD with the Tpieces 1/8" on the incoming connections from the generators and 1/4" on the main line connections. The supply line for an expanded capacity system should be connected using t pieces as shown below, in the recommended ring main layout. This allows for the pressure to equalise.
To Instruments
Alternatively the system can be connected in a series as shown in the image below. If a series layout is chosen, the Master must be the unit closest to the instrument.
MASTER
To Instruments
Should any of the generators fail, hydrogen will continue to be supplied and the remaining units will increase supply to compensate. If the failed unit is the Master, the next machine in line will take over this role.
Page 26
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Switching the unit off
The unit can be switched off at any time using the following procedure. Firstly press the Stop button on the Touch screen, then press Menu, and finally select Shut Down. The unit will take a few moments to depressurize after which the power switch at the rear of the generator can be switched off.
Cleaning
Clean the outside of the Generator only using warm soapy water and a clean damp cloth. Ensure the cloth is thoroughly rung out to remove excess fluid prior to use. Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the Generator.
Under no circumstances should any solvents or abrasive cleaning solutions be used as these can contain fumes that could be harmful to the Generator.
Page 27
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Replacement of Deioniser Resin Column
The resin, in the Deioniser column, in the unit should be replaced every 6 months. Listed below are the instructions for replacing the Deioniser column.
1.
Switch the generator off.
2. Using a 2.5mm Hex key remove the screws located in the four bottom corners of the generators cover, as shown in the first image.
3. Once the screws have been removed lift the cover off the unit by pulling upwards. 4. Locate the used Deioniser Column, drain tank or clamp prior to removing the tubing from the top and bottom of the column as showing in image 3..
5. Remove the used column by pulling it towards you. 6. Fit a re-filled Deioniser Column into the generator and reconnect the tubing,
7. Replace the cover and screws. 8. Switch the unit on.
Page 28
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Refilling the Deioniser Resin Column
The resin, in the Deioniser column, in the unit should be replaced every 6 months. Below are instructions for replacing the resin in the column.
1.
Unscrew the end cap of the Deioniser Column, this should be hand tight, and then remove the rubber O ring and sponge.
2. Pour the saturated resin out of the column and dispose of it. (See MSDS on P.30&31)
3. Remove the sponge from the bottom of the column.
4. Put a new sponge (05-3290) at the bottom of the column, then open the replacement gel pack (05-4110) and pour its contents into the column.
5. Fit a new sponge at the top of the column, refit the O ring and screw the lid back on the column, until the O ring is in contact with the cartridge. Ensure the lid is not over tightened, as this could damage the thread.
Page 29
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Health and Safety Data- Deioniser Resin
This sheet to be retained by the 'Health & Safety Officer' The recipient of this Health & Safety Data sheet is required to ensure that all personnel handling, using or likely to come into contact with the material are made aware of the Health & Safety information contained herein. General Description Bag containing ion exchange resin. 60% Cation and 40% Anion by mass. Chemical Components Polystyrene divinylbenzene cation and ion exchange resin. Form Flash Point Colour Ignition Temperature Odour Explosion Limits Decomposition Temperature Vapour Pressure @ 25 °C Viscosity @ 25°C Solubility in Water Specific Gravity @ 25°C pH @ 25 °C Eye Protection Storage Solid beads N/A Tan 427 °C (Estimated) None N/A Over 220 °C 17mm CFH @ 20 °C (Water) N/A Negligible 1.1-1.3 7-9.5 in aqueous slurry Safety Glasses (BS2092C) Store in a cool place above 0 °C, out of direct sunlight Toxic combustion products may include sulphur dioxide and sulphur trioxide This product can be disposed as ordinary waste if local regulations permit
Fire & Explosion Hazards
Disposal
Page 30
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Emergency & First Aid Procedures Spillages Floor may be slippery. Use care to avoid falls. Wear protective glasses. Sweep up and transfer to containers for recovery and disposal. Extinguishing Media Carbon Dioxide (CO2), Water Fog and Dry Powder. Fire Fighting and Protection Wear HSE approved, pressure demand, self-contained breathing apparatus or equivalent. First Aid Skin Contact Wash effected areas with soap and water. Consult doctor if irritation persists. Eye Contact Product can cause severe eye irritation. Flush eyes with large amounts of water for at least 15 minutes. Consult a doctor if irritation persists. Inhalation This product should present no hazard if operated under normal conditions. If irritation persists, move patient to fresh air. Ingestion If resin has been confined to the mouth, administer 250 ml. of water as a mouthwash. Avoid swallowing. This product is considered non-hazardous if ingested in small amounts; however, it is advisable to seek medical advice. Toxicological Data No toxicity data available on this product Occupation exposure limits STEL/TLV = None established. Incompatible Materials (to avoid hazardous reactions) Avoid contact with concentrated nitric acid and any other strong oxidising agents at all times.
Page 31
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
IMPORTANT DOCUMENTS Warranty Entitlement
To register your generator for your warranty entitlement, send the completed form to Peak Scientific by: • • • • Email Online Phone Fax
warranty@peakscientific.com http:/ /www.peakscientific.com/service-and-support/warranty_registration
+44 (0)141 530 4185 +44 (0)141 812 8200
PRODUCT WARRANTY REGISTRATION
COMPANY: ADDRESS: EMAIL ADDRESS: CONTACT NAME:
CITY/TOWN: POSTCODE: COUNTRY: TELEPHONE:
GENERATOR SERIAL NUMBER:
MODEL TYPE: INSTALLATION DATE (DD/MM/YYYY):
Important Please Note:
You have 1 month to register your Peak Scientific product from the date of shipment. If you wish to defer installation of your generator you must notify Peak Scientific within 1 month of the shipment date. This can be done by emailing warranty@peakscientific.com Once registered the warranty will be honoured for a period of 12 months after the installation date. For any generators that remain unregistered the warranty will begin from date of shipment. Thank you on behalf of Peak Scientific.
Page 32
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Service Requirements
Service Schedule
Purchase Interval 12 months 24 months Component Precision 500 Annual Service Kit Recirculation Pump Table 1: Service Schedule Part No. 08-8899 05-5057 Qty. 1 3
Annual Service Kit Contents 2 x Deioniser Resin Refill Pack 2 x Column Sponge
Service Plans
Peak Scientific offer two service plans. The Complete Service Plan, specifically designed for Generators operated in critical environments, also includes full breakdown cover, guaranteed response times and Generator upgrades if available. Our Standard Service Plan, covering the basic needs of our Generators, features special deals on spare parts and breakdowns. If you want to know more about our Service Plan options and how we ensure that your instrument can run with the maximum uptime and performance, please contact us at maintenance@peakscientific.com
Page 33
2012 © Peak Scientific – Rev 1 – 17/09/12
Precision 500 User Manual
Trouble Shooting
Problem The Generator will not switch on and the power switch does not illuminate. Diagnostics Failed Alarm Cell Over Voltage Alarm Low Water Supply Alarm Over Pressure Alarm Capacity Excess Alarm Water Level Fault Hydrogen Detected Possible Solution • • • • • • • • • • • • Ensure power cord is plugged into the Generator and that the power socket is turned on. Check the fuse in the power cord plug. Contact your service provider. Contact your service provider. Contact your service provider. Check Water Supply level, and fill if necessary, Contact your service provider. Contact your service provider. Check external pipework for leaks. Contact your service provider. Contact your service provider. Check for external leaks.
Page 34
2012 © Peak Scientific – Rev 1 – 17/09/12
Peak Scientific UK Fountain Crescent Inchinnan Business Park Inchinnan PA4 9RE Scotland, UK Tel: +44 (0)141 812 8100 Fax: +44 (0)141 812 8200 Peak Scientific Germany Herriotstrasse 1 60528 Frankfurt
Peak Scientific North America 19 Sterling Road Suite #1 Billerica, MA 01862 USA Tel: +1 866 647 1649 Fax: +1 978 608 9503 Peak Scientific China Room 606 Building 1 Lane 2277 Zuchongzhi Road Pudong New Area Shanghai 201203, China Tel: +86 21 5079 1190 Fax: +86 21 5079 1191 Peak Scientific Brazil Peak Scientific Brasil Av. Quieroz Filho 1700 – torre Sunny-Conj. 602 São Paulo SP Brasil Tel: +1 866 647 1649
Peak Scientific India 202, Amsri Shamira Old Lancer Line Opp. St. Mary’s Degree College S.D. Road Secunderabad 500 025, India Tel: +91 40 2780 0663 Fax: +91 40 2780 0663 Peak Scientific Mexico Solon 352 Col. Los Morales Polanco 11530 Mexico, D.F. Mexico Tel: Fax: +1 866 647 1649 +1 978 608 9503
Peak Scientific Africa PO Box 478 Somerset Mall 7137 South Africa Tel: +27 (0)2185 16542 Fax: +27 (0)8654 64473 Peak Scientific Australia PO Box 65 Belmont Victoria 3216 Australia Tel: +61 1300 965 352
Germany Tel: +49 (0)69 677 33 490 Fax: +49 (0)69 677 33 200 Peak Scientific Taiwan 4F.-6, No.736 Zhongzheng Rd. Zhonghe Dist. New Taipei City Tel: +886 2 8226 2383 Fax : +886 2 8226 9499
Peak Scientific Japan K.K. 2-7-56, 2F Fuji Building 28 Kita Aoyama, Minato-Ku Tokyo, Japan 107-0061 Tel: +81 3-6864-0468
Peak Scientific Singapore 3 Science Park Drive #03-14 The Franklin Science Park Drive 1 Singapore, 118223 Tel: +65-6777 1966 Fax: +65-6777 1978
2012 © Peak Scientific – Rev 1 – 17/09/12
</p>