INSTALLATION QUALIFICATION
Instrument Origin and Identification
Instrument Series:
Genius 30 Series
Model:
Genius 3010
Instrument Description:
Nitrogen Gas Generator
Manufacturer:
Peak Scientific Instruments Ltd
Fountain Crescent
Inchinnan Business Park
Inchinnan
Renfrew PA4 9RE
Scotland, UK
Service Contact:
Peak Scientific Instruments Ltd
Tel.
+44 (0)141-812-8100
Email support@peakscientific.com
Serial Number:
Condition when Received:
New
Used
Reconditioned
Date Received:
Installed Location:
Purchase Order Number:
Identification / Asset Number:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 1 of 8
INSTALLATION QUALIFICATION
Instrument Specification
Instrument Parameter
Specified Range
Acceptable
Output
Nitrogen Pressure
100 psig (6.90bar) Max
Nitrogen Flow Rate
64 l/min (ATP)
System Purge Time
30 minutes
Environment and Siting
Has the instrument been adequately acclimatized since transport or storage?
(Requires 3 hours before operation)
Yes
No
Operating Parameter
Specified Range
Ambient Temperature
Acceptable
5-35 ºC (41-95 ºF)
Maximum Relative Humidity
80% Non-Condensing
Maximum Altitude
2000 m (6500 ft)
Supply Voltage & Frequency
230V ±10%, 50/60Hz
Refer to User Manual for acceptable range
Siting – In-House / Large Space
Minimum Clearance around Machine – 75mm
Siting – In-House / Confined Space
Air Conditioner/Extraction Fan
Mains Power Outlet
Within 2m Radius to Generator
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 2 of 8
INSTALLATION QUALIFICATION
Instrument Delivery and Documentation
Unpack the instrument carefully. Are any items missing against the Packing List?
Yes
No
If yes, state missing items:
Is there any damage to the instrument?
Yes
No
Yes
No
If yes, describe damage:
Manufacturer informed:
Is all standard documentation included?
Present
Missing
User Manual (Available via download from Peak Extranet)
Warranty Registration Card
Installation Guide
Site Preparation Guide
(Available via Download from Peak Extranet)
Installation
Installed By:
User
Authorized Engineer
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 3 of 8
INSTALLATION QUALIFICATION
Summary Report
Instrument:
Genius 3010
Serial Number:
Assessment of Complete Installation Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument has been
satisfactorily delivered and installed. The instrument has passed the Installation Qualification
procedure and may now be submitted for Operational Qualification.
IQ Completed By:
Date:
Deviations Approved By:
Date:
IQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 4 of 8
OPERATIONAL QUALIFICATION
Pre-Run Checks
Pre-Run Check
OK
Discrepancy
N/A
LH Compressor Transit Brackets Removed
(Upper and Lower Compartments)
RH Compressor Transit Brackets Removed
(Upper and Lower Compartments)
Pressure gauges showing 0 psig
Drain line connected to suitable outlet
Assemble Silencer to “Nitrogen Out”
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 5 of 8
Functional Tests
Functional Test
Acceptance Criteria
Pass
Apply Mains Power / Switch On
Internal Volt meter reads
between 220V – 253V
Fail
Compressor(s) Cycle
Allow to run for 30 minutes.
No Alarms
(Ensure Side Panels are fitted)
Pressure builds to 100 PSI
Output pressure
Stable pressure shown on panel
pressure gauge (100 PSI)
After 30 minutes 1. Power-off / Isolate Generator from Mains Power / Allow pressure to drop to Zero
2. Remove Silencer from “Nitrogen Out”
3. Apply Mains Power / Switch On
4. Perform Calibration on Generator
Calibration
This instrument is supplied fully calibrated by the manufacturer using equipment traceable to
national standards.
Calibration Step
Pass
Fail
Calibrate output flow(s)
Serial No. of Calibrated Flow Meter -:
Calibration Due Date of Flow Meter -:
Calibrate output pressure(s)
Serial No. of Calibrated Pressure Gauge -:
Calibration Due Date of Pressure Gauge -:
Comments:
5. Connect Instrument to “Nitrogen Out”
6. Apply Mains Power / Switch On and allow to run for 3 minutes
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 6 of 8
Functional Test contd.
Acceptance Criteria
Pass
Output pressure
Stable pressure shown on panel
pressure gauge (100 PSI)
Output Connection Leak Tight
No Leaks
Fail
Comments:
Training
Date Instrument Released for Use:
Trained Operators
Read User
Manual
Practical
Training
Authorized Signature
Comments:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 7 of 8
OPERATIONAL QUALIFICATION
Summary Report
Instrument:
Genius 3010
Serial Number:
Assessment of Complete Operational Qualification:
No Deviations
Deviation
Deviations
Impact on Operation
Justification for Acceptance
Successful completion of the preceding activities and checks indicates that this instrument is
operating satisfactorily following delivery and installation. The instrument has passed the Operational
Qualification procedure and may now be released for use.
OQ Completed By:
Date:
Deviations Approved By:
Date:
OQ Approved By:
Date:
Fountain Crescent,
Inchinnan Business Park
Inchinnan, PA4 9RE
Scotland, UK
T: +44 (0)141 812 8100
F: +44 (0)141 812 8200
W: www.peakscientific.com
E: info@peakscientific.com
Doc. Ref.: BF-818
Revision: 2
Page 8 of 8