Genius AB-3G
User Manual
GeniusAB-3G User Manual Rev 2 PN 3300041 18/11/21
Contents
Change History
How to use this Manual
Introduction
Warranties and Liabilities
Warranty & Liability Coverage
Safety Notices
Symbols
Safety Notice to Users
EU Declaration of Conformity
UK Declaration of Conformity
WEEE Compliance Statement
Technical Specification
Genius AB-3G
Unpacking
Fittings Kit Contents
Installation
Generator Environment
Generator Overview
General Dimensions
Rear Connections
Electrical Connection
Start-Up Sequence
Operating Modes
Normal Operation
Standby Mode
High Altitude Mode
HMI User menus
On Demand Gas
Generator Cycling
Unusual Operation
Fault Warning
Connecting to the application
Tubing Lengths
Service Requirements
Service Schedule
Peak Protected
Cleaning
Troubleshooting
Page 2
3
3
4
5
5
7
7
7
8
9
10
11
11
12
13
14
14
15
15
16
17
18
20
20
20
20
21
23
23
23
24
25
25
26
26
27
28
29
Change History
Rev
1
Comment
Declarations Update
Name
L. Couttie
Date
20/04/2017
2
Declarations Update
L. Couttie
18/11/2021
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The Genius AB-3G nitrogen generator is designed specifically for use with Sciex LC/MS
instruments. This premium model offers safeguards that are unique to Peak Scientific
generators and effectively deliver reassurance and confidence for critical functions.
These model provide a source of nitrogen gas with other features including:
•
Reassurance – Designed with back-up compressor capacity
•
Fail-Safe – Continuous operation to allow for scheduled maintenance
•
Confidence - Designed and built around extensive tried and tested processes.
•
Intelligent – High duty service indicator
•
Independant - No external compressor required
•
Quiet - Insulated compressor compartment for minimum disruption
•
Mobile – Flexibility to position the generator where required
•
Simple installation – Generator designed as a plug and play system
•
Economical – More cost effective than any other gas supply method
•
Convenient – Gas on demand, no hazards, no need to worry about running out of
gas
The Genius AB-3G is based on proven technology, and selectively removes oxygen,
moisture and other gases to leave clean, dry, phthalate free nitrogen and dry air.
Internal air compressors make this unit independent from in-house air supplies and
fitted castors allow the user to easily position the unit in the lab.
To ensure this Generator model meets our high expectations with regards to reliability
and performance, we have tested it extensively at our manufacturing plant and with
end users around the world to ensure reliability and longevity of the system.
Page 4
Warranties and Liabilities
Warranty & Liability Coverage
1.
Peak warrants that, subject to the provisions in this statement, purchased Peak generators, whether
purchased directly from Peak or indirectly via an approved, certified and trained distributor or partner
(referred to hereafter as a “Peak Partner”) will comply in all material respects with any specifications
referred to in your customer order confirmation and, subject to installation and operational guidelines
being followed as described in applicable product manuals, shall be free from any defects in quality
of materials or workmanship for a period of one year from the date of installation, provided this takes
place within 3 months of factory dispatch.
2. Where the purchased generator is from the Precision Hydrogen series, Peak further warrants that,
subject to installation and operational guidelines being followed as described in applicable product
manuals, the hydrogen cell shall be free from any defects in quality of materials or workmanship for a
total period of three years (inclusive of warranty period specified in clause 1) from date of installation,
provided this takes place within 3 months of factory dispatch.
3. Where the purchased generator is from the i-Flow 6000 series, Peak further warrants that, subject
to installation and operational guidelines being followed as described in applicable product manuals,
the generator shall be free from any defects in quality of materials or workmanship for a total period
of two years (inclusive of warranty period specified in clause 1) from the date of installation, provided
this takes place within 3 months of factory dispatch and the following provisions have also been met:
a. you must purchase a service plan, ensuring the generator is serviced by Peak or a Peak Partner on
or before the end of the first 12 months of your ownership, and serviced at least once during each
subsequent 12 month period thereafter;
b. the generator (and any associated equipment)
must have been commissioned by Peak or a Peak Partner;
c. the feed air or inlet air supply to the generator must comply with ISO 8573-1:2010 Class 1.2.1 at all
times;
d. your air compressor, dryer, filtration and oil removal systems must be deemed suitable for use
by Peak or a Peak Partner, and must be changed and serviced regularly, in line with the equipment
manufacturer’s recommended guidelines; and
e. any generator failure or fault that is deemed to have been caused by the failure of any upstream
equipment, component, part or system (such as air compressor, air treatment or filtration) will be
excluded from the warranty described herein.
4. Peak also warrants that any replacement parts whether purchased (directly from Peak, or via a Peak
Partner) or supplied as part of any remedial action undertaken in line with the provisions of clauses 12
and 13, shall be free from any defects in quality of materials or workmanship for a period of 180 days
from the date of factory dispatch, provided its installation is performed by Peak or a Peak Partner.
5. This warranty does not exclude Peak’s liability in respect of any claim for death or personal injury
to any person, in so far as such can be attributed to negligence or breach of duty of care directly
resulting from failure of Peak to comply with the provisions in clauses 1, 2, 3 & 4.
Exclusions & Limitations
6. This warranty does not cover:
a. damage, deterioration or malfunction resulting from an alteration or modification to a generator
which has not been carried out by Peak or a Peak Partner;
b. damage, deterioration or malfunction resulting from what Peak reasonably believes to be abuse, or
misuse of a generator by you or any third party;
c. liability for accident or neglect (other than pursuant to clause 5);
d. maintenance or repairs which have not been carried out by Peak or a Peak Partner;
e. operation of a generator or exposure of a generator to environmental conditions that fall out-with
operational guidelines as specified in the applicable product user manual; and
f. lightning, power surges or any other acts of God or nature.
7.
This warranty is non-transferrable. Only the original owner of the generator may benefit from the
terms within this statement.
8. Peak shall not be liable in respect of any claim made for costs, damages, losses or expenses (whether
Page 5
consequential, direct, indirect or otherwise) or in any respect howsoever arising including, but not
limited to, liability from accident or negligence (other than pursuant to clause 5) that may be suffered
by you or any third party.
9. No person or entity is authorised to change the terms and conditions outlined in this warranty
statement in any respect, or to create any additional obligations or liabilities for any party involved.
10. This warranty statement supersedes any and all prior warranty agreements between the parties and
constitutes the complete, final and exclusive understanding of the parties with respect to the subject
matter. All prior negotiations, representations, or promises, whether oral or written, of either party
shall be deemed to have been merged herein.
11. If any part of this warranty statement is invalidated, for whatever reason, such part will be deleted
and the rest shall remain unaffected, continuing to be in full force and effect.
Delivery of Warranty Service
12. Subject to clause 13, and:
a. Peak being notified by you, within the duration of the applicable warranty period, of any defect
that you think is subject to any warranty valid under clauses 1, 2, 3 or 4; and
b. Peak being permitted to inspect the generators, parts and their installation (along with any relevant
packaging)
Peak shall at its option repair or replace defective generators or parts (including, if necessary, any
moving parts and irrespective of runtime). No additional charges will apply, for parts or delivery and,
where applicable, labour or travel. Peak will endeavour to deliver this service within 3 working days of
your notification.
13. Where, in Peak’s reasonable opinion, a defect is subject to an exclusion described in clause 6, Peak
reserves the right to charge for parts or delivery and, where applicable, you may also be charged by
Peak for call out, labour or travel in respect of any repair or replacement which you authorize Peak to
carry out.
Page 6
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which may
represent a potential hazard. The warnings detailed within this manual refer to the most
likely potential hazards, but by definition cannot be all inclusive. If the user employs
an operating procedure, item of equipment or a method of working which is not
specifically recommended by Peak Scientific, the user must ensure that the equipment
will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the safe
and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.
Safety Notice to Users
WARNING
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the generator
in a manner not specified by Peak Scientific MAY impair the SAFETY
provided by the equipment.
WARNING
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING
Page 7
EU Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment Type:
Nitrogen & Dry Air Generator
Model Designator:
Genius AB-3G & AB-3G Hi-Flow
To which this declaration relates, is in conformity with the following applicable EU
Directives, harmonized standards, and other normative requirements.
•
Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use.
•
Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for Measurement, Control and Laboratory Use
– EMC Requirements. (Class A)
•
Restriction on the use of certain hazardous substances in electronic equipment
(RoHS) Directive 2011/65/EU as amended by EU 2015/863.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
16th November 2021
Page 8
UK Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment Type:
Nitrogen & Dry Air Generator
Model Designator:
Genius AB-3G & AB-3G Hi-Flow
To which this declaration relates, is in conformity with the following applicable UK
Statutory Instruments, Standards and other normative requirements.
•
The Electrical Equipment (Safety) Regulations 2016 (SI 2016 / 1101) as amended.
BS61010-1:2010 Safety Requirements for Electrical Equipment for Measurement
Control and Laboratory Use.
•
The Electromagnetic Compatibility Regulations 2016 (SI 2016 / 1091) as amended.
BS61326-1:2013 Electrical Equipment for Measurement , Control and Laboratory Use
– EMC Requirements.
•
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012 (SI 2012 / 3032) as amended.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
16th November 2021
Page 9
WEEE Compliance Statement
The Waste Electrical and Electronic Equipment (WEEE) Regulations SI 2013 No 3113
and or the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU
apply to all electrical and electronic equipment placed on the market in the UK and EU
covered by the scope of regulations which can be found in the Government Guidance
Notes (PDF) produced by the Department for Business Innovation and skills for the UK
and here for Europe.
All PEAK products that are subject to the WEEE directive are compliant with the
WEEE marking requirement. Such products are marked with the “crossed-out wheelie
bin” symbol (shown below) in accordance with European standard EN50419. All
old electrical equipment can be recycled. Please do not dispose of any electrical
equipment (including those marked with this symbol) in general rubbish bins. Please
contact your dealer or distributor for clarity.
Page 10
Technical Specification
Genius AB-3G
Environment
AB-3G
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
AB-3G Hi-Flow
5°C (41°F)
30°C (86°F)
2000 m
70%
-20°C (-4°F)
60°C (140°F)
* NOTE. When taken out of storage the generator should be allowed to acclimatize at room temperature for a
minimum of 3 hours before operation
Generator Outlets
Curtain Max Flow @ 80 psi
Source Max Flow @ 110 psi
Exhaust Max Flow @ 60 psi
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlets
Pressure Gauges
Start-Up Time
12 L/min (0.42 cfm)
12 L/min (0.42 cfm)
24 L/min (0.85 cfm)
24 L/min (0.85 cfm)
8 L/min (0.28 cfm)
14 L/min (0.49 cfm)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP
3
30 minutes
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord
Pollution Degree
Insulation Category
230 VAC ± 10%
50/60 Hz
9.5 Amps
C20 Plug
C19 socket to local connection (13A minimum)
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 11
90 x 73 x 75.3 (35.4 x 28.7 x 29.6)
145 kg (316lb)
200 kg (441lb)
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require two
people to remove the unit from the shipping crate and to manoeuvre the generator to
the desired location.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads for
UK, EU & US and also all the required fittings and warranty registration card. Be careful
not to discard these with the packaging.
Page 12
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. ¼” Teflon Tubing 4 x 3m
2. 6mm PE Tubing x 3m
3. ¼” Compression Fitting x 3
4. 6mm Push Fit Fitting
x 1
5. UK Mains Power Cable x 1
6. EU Mains Power Cable x 1
7. US Mains Power Cable 230v
x1
8. 8mm Hex Key x 1
All of the generators output ports are located on the output panel at the rear of the
unit.
Page 13
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator (like all sophisticated equipment) is affected by ambient
conditions. Note should also be taken to the proximity of Air Conditioning outlets.
These can sometimes give rise to “pockets” of air with high relative humidity. Operation
of the unit within such a pocket could adversely affect its performance. Consideration
should also be given to the air flow around the unit. It is recommended that an air gap
of 75mm (3”) should be maintained down both sides and at the rear of the unit. Please
refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
30 °C (86 °F)
Page 14
Generator Overview
General Dimensions
900 mm / 35.4”
730 mm / 28.7”
Peak Scientific
AB-3G
NORMAL OPERATION
Menu
753 mm / 29.6"
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING
Page 15
Rear Connections
Mains Cable
Page 16
6mm Push-Fit Fitting
¼" Compression Fitting
Electrical Connection
Connect the generator to an appropriate 110 or 230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 12 amps, can
be fitted by a qualified electrician.
WARNING
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow
or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements for 230V generators is 195V - 253V 50/60Hz. For 195V - 219V a
06-3210 Dual Tap Transformer is required.
If the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment maybe impaired.
WARNING
Page 17
Start-Up Sequence
Once connected to the mains supply the generator can now be switched on. Flick the
power switch on the front panel of the generator to the on position
The first time the unit is switched on it will ask you to confirm that the transit brackets
have been removed. The unit will not start until the user has pressed the confirm
button.
The unit will then start an installation diagnostics sequence. The diagnostics will
take approximately 20 minutes to complete. A progress bar is shown to indicate the
progress of the tests. This is not an indication of time.
During the diagnostics the user will be asked to confirm that the tubing connected to
the drain outlet is secure and that if it is connected to a drain container that it is empty.
The diagnostics will be paused until the user presses confirm.
WARNING
During diagnostics air is expelled through the drain outlet at pressure.
Please ensure that the tubing is secure. If a drain bottle is used please
ensure that it is empty before carrying out a diagnostics check as
water may be sprayed from the container.
The drain bottle should NOT have an air tight seal.
WARNING
CAUTION
Page 18
It should be noted that the gas pressures and flows are factory set.
The pressures shown on the Generator front panel are in excess of
the maximum inlet pressure of the Mass Spectrometer. This is to allow
for known pressure drops. These settings have been approved by AB
Sciex.
CAUTION
Once the tubes are connected to the appliance, ensure that they are
thoroughly checked for being leak-tight. Even the slightest leak in the
gas supply between the Generator and the application can lead to a
reduction in efficiency, or possible alarm of the Generator.
The diagnostics sequence will only run on subsequent start-ups if there was a problem
detected before it was shut down. If no fault had previously been detected and the
installation diagnostics has been completed, a start-up screen will appear for 30
seconds to allow the user to initiate a diagnostics check. This is recommended if the
unit has been turned off for a long period or the generator has been moved.
Once diagnostics have completed the generator will take a couple of minutes to
become ready for use.
Once pressure is within the normal operating limits the buzzer will sound two short
beeps to signify the generator is ready for use. A reminder to register the Generator
with Peak Scientific will be displayed.
Once you have registered the generator pressing [Acknowledge] will prevent this
message appearing again. Pressing [Remind me later] will display the message again in
one and a half hours. You can register online at http://www.peakscientific.com/serviceand-support/warranty-registration
CAUTION
Page 19
Before the Generator is connected to the Mass Spectrometer the
Generator should be operated in isolation (i.e. not connected to the
Mass Spectrometer) for thirty minutes after the start up diagnostics
have completed. This is to ensure any impurities present are purged
from the system. Failure to do this may harm the Mass Spectrometer.
Operating Modes
The generator can operate in a number of modes depending on the gas demand
from the Mass Spectrometer and the health of the generator. The operating mode is
displayed on the main screen of the HMI. The operating mode will change automatically,
no user intervention is required.
Normal Operation
The Genius AB-3G Gas Generators are designed specifically to minimize operator
involvement. Given that the systems are installed as described in earlier sections and is
serviced in accordance with the specified maintenance recommendations (see Service
Requirements), then it should simply be a matter of turning the Generator on when it is
required.
The Generator will automatically produce the factory set flow and pressure as detailed
in the Technical Specifications.
Standby Mode
The generator will detect when the Mass Spectrometer enters standby mode. Due to
the reduced gas demand the Generator controls the compressors differently to help
extend compressor life.
It should be noted that even in standby the Mass Spectrometer still consumes a small
amount of Nitrogen and Dry Air. This in turn reduces the receiver pressure within the
Generator causing the compressors to start up. This is normal in Standby Mode.
The operating mode on the main screen will change to Standby Mode.
High Altitude Mode
High altitude mode allows the Generator to operate easier at high altitude. In this mode
both air compressors run together during Normal Operation. Consult with your local
Peak Representative before enabling this function.
CAUTION
It should be noted that this will increase the duty on the compressors
and reduce time between services. This mode should only be used
where required.
The main screen will now display “High Altitude Mode” instead of “Normal Operation”
Page 20
HMI User menus
The AB-3G’s advanced control system employs an easy to navigate menu system to
give the user additional information about the Generator. The menu is accessed by
pressing the [Menu] button on the main screen.
User Menu
The user menu gives access to the following functions
Run hours
Displays:
•
Unit on time – The total time the unit has been turned on.
•
The date the next annual service is due.
•
Compressor run times – The time that each compressor has actually been running.
Temperatures
The AB-3G monitors ambient (room) temperature and the temperature within the
compressor cabinet inside the generator as part of the temperature alarm system. The
information on this screen is a live reading of these two temperature inputs. The alarm
set points can be accessed via this screen also.
Page 21
System status
The status of each component within the Generator is displayed. This includes the three
compressors and all of the solenoid valves used within the system. This can be used
should a fault be detected to show which components have failed.
Event logs
The event logs record the date and time of all system events. These are split into two
logs
•
System Faults
•
Temperature alarms
System faults log the error code along with the date and time. Recent error codes
should be noted and passed to your service provider if you are reporting a fault. The
last twenty faults are logged.
Temperature alarms will record every time a temperature warning or over temperature
shutdown occurs. The last ten faults are logged.
Set time/date
The date and time, as displayed on the main screen, is set here. Setting the date and
time ensures that any events a correctly logged.
Comp hours reset
When a compressor is serviced or replaced the run hours of the compressor need to
be reset here. This screen is password protected. A password will be included in the
instruction manual provided with your replacement compressor or compressor service
kit.
High Altitude Mode
This allows High Altitude Mode to be enabled or disabled
Event message
This displays details of the current error if either of the fault icons are displayed on the
main screen.
Service Engineer
This is a password protected area of the controls. A caution screen is displayed to
allow the user to go back to the main menu. Access is only granted to qualified service
personnel. Please contact Peak Scientific if you require training.
Page 22
On Demand Gas
The generator will produce nitrogen on demand. If the application is operating
and requires a gas flow, the system it is connected to will supply this to suit the
requirements of the application. If the application requirement for gas stops, the system
will also stop, once it has reached its upper set limit in the internal storage tanks. If the
demand from the application starts again, the system will detect the demand for gas
and will automatically start again to suit the demand.
Generator Cycling
The Generator is designed for the internal compressors to cycle. This cycling reduces
the duty (run time) on the compressors. The rate at which they cycle will be dependent
on the gas required to satisfy the demand of the application. If the application demands
the maximum gas flow of the Generator, the compressor duty will be higher, (the rest
period in the compressor cycle will be shorter). If the application demand is lower than
the maximum gas flow, then the duty on the compressors will decrease, (the rest period
in the compressor cycle will be longer). If the Generator is installed in an extreme
environment or is subjected to low supply voltage or high altitude the compressors may
undergo periods where they do not cycle.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.HeaderHeader
Page 23
Fault Warning
Should the unit detect a fault, a message will be displayed on the HMI to explain the
fault and inform the user of the current status of the machine. There are three levels of
fault.
•
Minor fault
•
Redundancy mode
•
System failure
•
Temperature alarm
•
Run time alarm
A minor fault is shown when one of the valves used during diagnostics has failed.
These are set to fail in the normal operating mode therefore the system can function
as normal. You should contact your service provider to rectify the problem at your
convenience
Redundancy mode is likely to indicate a failed compressor. The unit will continue to run
with the remaining two compressors. An intermittent audible alarm will also sound to
alert the user. This alarm can be silenced.
A system failure means that the unit can no longer supply gas and will be shut down.
An audible alarm will sound constantly to alert the user. This can be silenced.
A temperature alarm is to ensure the generator is not being operated over the specified
maximum temperature (30°C). The temperature within the internal compressor
cabinet is also monitored to ensure that they are properly cooled. If the Generator is
operated above the maximum temperatures then damage can be caused to the internal
compressors and the membrane separators.
Run time alarm monitors the length of time that the compressors have been running. If
everything is working properly the compressors should cycle regularly. If a compressor
has not cycled but is still maintaining pressure the Generator will continue to operate. A
run time alarm will generally mean there is a problem with the system which may lead
to system failure. Your service provider should be contacted immediately.
Once the fault message has been acknowledged by the user an icon will be displayed
on the main screen as a reminder.
Page 24
Connecting to the application
Once the initial purge run of 30 minutes has completed, and the generator has been
running for 1 hour, it is ready to be connected to the application(s).
The pressure in the internal storage tanks must be allowed to dissipate
before connecting the generator to the application(s)
CAUTION
Attach the appropriate fittings to the outlets of the generator. Using the tubing
supplied, connect the outlets of the generator to the inlets on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.
CAUTION
Once the tubing is connected to the application, please ensure that it
is thoroughly checked for being leak-tight. Even the slightest leak in
the gas supply between the generator and the application can lead to
a reduction in efficiency.
Tubing Lengths
CAUTION
The diameter of the tubing which will be connected to the gas outlet is
important and is determined by the length of tubing required. Failure
to follow these recommendations could lead to pressure between
generator and application.
< 10 meters:
Use 6/4 (6mm O/D, 4mm I/D) P.T.F.E. tubing.
> 10 - 40 meters:
Use 10/8 (10mm O/D, 8mm I/D). Tubing and fittings not
supplied in the fittings kit.
> 40 metres:
Please contact Peak Scientific with the relevant distance andwe
will calculate the flow resistance and the tubing size required.
A combination of 6/4 and 10/8 tubing may be used to ensure that there is no large
diameter tubing within the lab (i.e. for the first 20 meters from the Generator use 10/8
and the final 10 meters to the application use 6/4 tubing). Keep the connections and
bends to a minimum.
The imperial equivalents are: 6/4 = 1/4” O/D, 3/16” I/D.
10/8 = 3/8” O/D, 5/16” I/D.
Page 25
Service Requirements
Service Schedule
Purchase Interval
Component
Visit
12 Months
Genius AB-3G Annual Maintenance Kit
www.peakscientific.com/ordering
Service Indication
The Generators will display the service icon when a service is due.
Page 26
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit
your lab to install and setup your
generator
IQ/OQ
Certified assurance for applications
requiring documented qualification
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance
Spares
Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime
Premium Protected
Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance
Technical Support Hotline
Around the clock support by phone
or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 27
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the generator.
CAUTION
Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Page 28
Troubleshooting
Problem
The Generator will not switch on and the
power switch does not illuminate.
Possible Solution
•
Ensure power cord is plugged into the
Generator and that the power socket is
turned on.
•
Check the fuse in the power cord plug.
•
Contact your service provider.
•
Disconnect power cord from the rear of
the Generator. Open the left hand panel
are check that both circuit breakers are
turned on (switch in the up position).
Reconnect power cord.
•
Contact your service provider.
•
Check the HMI screen to ensure that
diagnostics are not running. The pressure
gauges on the front panel will read zero
for the majority of the diagnostics run.
•
Run diagnostics.
•
Contact your service provider
•
This is normal. When the Mass
Spectrometer is in standby it still requires
a gas supply, the Generator changes its
operation to best cope with this reduced
demand.
•
Contact your service provider.
The Generator will not switch on but the
power switch is illuminated.
Compressors are running but pressure is not
building.
The HMI screen says “Standby Mode” but
the compressors are still running.
The mass spec is reporting low pressure.
Page 29
Go Online or Complete and Return
We know that registering any of your recently purchased products is not the
first thing on your mind- but it is very important to both of us. Not all warranties
are alike and Peak Scientific stand out against other gas suppliers as we offer a
comprehensive, quick response, on-site warranty. This means that in the very
unlikely case that your gas generator develops a fault we have rapid support
teams on-hand around the world who are able to come to your lab and get you
back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support packages.
Peak [Protected] aftercare support can guarantee an on-site response within
72 hours*, genuine parts from our ISO9001 approved factory and a 95% first-time
fix rate. See our enclosed Peak [Protected] leaflet for further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once registered
the warranty will be honoured for a period of 12 months. If you wish to defer the installation of your
generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.com.
For generators that remain unregistered after 1 month from the shipment date, the warranty will be
considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 30
Page 31
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid response
service to our customers. With [Peak Protected], your laboratory’s productivity becomes our top
priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
GeniusAB-3GUserManualRev2PN330004118/11/21