MS Table (All Models) - User Manual
Resource ID
1927
Access
Open
Contributed by
Derek Rennie
Updated date
13 October 21
Title
MS Table (All Models) - User Manual
Keywords
MS, Mass, Spec, Table, User, Manual, All, Models, UM-0012
Product Range
Other
Language
UK English
Company Brand
Peak Scientific
Reviewed on 3rd may 2017 - LC
Extracted text
MS Table (All Models)
User Manual
Copyright © Peak Scientific Instruments Ltd 2021 — MS Table User Manual | Rev 3 | PN UM-0012| RSID 1927 | EN —06/10/2021
Contents
Change History
How to use this Manual
Introduction
Warranties and Liabilities
Safety Notices
Symbols
Safety Notice to Users
Declaration of Conformity
Environmental Declaration
Technical Specification
Technical Specification
Fittings Kit Contents
Unpacking Instructions
Installation
Generator Environment
Generator Overview
1.5L & 3.5L General Dimensions
Rear Connections
Drain Connection
Electrical Connection
Air Connection
Air Purity
Class 1 Particulate
Class 4 Water
Class 1 Oil
Connecting to the application
Tubing Lengths
Normal Operation
Unusual Operation
Relocation of MS Table
Fan Power LED
Safe Operation of Pump Tray
Infinity 1031 Installation
Service Requirements
Service Schedule
Peak Protected
Cleaning
Troubleshooting
Page 2
3
3
4
5
6
6
6
7
8
9
10
12
12
13
13
14
14
15
15
16
17
17
17
17
17
18
18
19
19
19
19
20
21
22
22
23
24
25
Change History
Rev
Comment
Name
Date
Rev 1
Update flow rates
David Lai
25 Aug 2017
Rev 2
Declaration update
David Lai
12/10/2021
How to use this Manual
This manual is intended for end users and has been written as a reference document
where you can skip to the relevant information.
Users can refer to the contents page to find the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your gas generation needs, and should
you require any further assistance or support please do not hesitate to contact Peak
Scientific or the Peak Partner from which you purchased your generator.
Page 3
Introduction
The MS table has been designed as a Sciex exclusive product to house the Triple Quad™,
QTRAP™ or IVD MS systems. The MS table comes in a standard and premium form, the
premium version including an inbuilt Infinity 1031 allowing the supply of curtain, source
and exhaust gases. Furthermore, the MS Table also functions as an accessible sound
proof enclosure for the required vacuum pumps for these applications.
Other features include:
•
Table designed to hold Sciex Mass Spectrometer
•
Can be supplied on its own or with an integrated Infinity 1031 compressor-less
nitrogen generator
•
Reduced lab footprint
•
Minimal set-up required
•
Sound proofed, reducing noise of generator
To ensure these generator models meet our high expectations with regards to
reliability and performance, we have tested them extensively at our manufacturing
plant and with end users around the world to ensure reliability and longevity of the
system.
Page 4
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the Goods
shall comply in all material respects with any specification referred to in the Order
Confirmation (as the same may be amended) and shall, subject thereto, be free
from defects in material and workmanship for the lesser of a period of twelve
months from the date of delivery or thirteen months from the date of dispatch
from the factory.
3. Save as provided in this clause and except where the Goods are sold to a person
dealing as a consumer (within the meaning of the Unfair Contract Terms Act
1977) all warranties, conditions or other terms implied by statute or common
law are hereby expressly excluded save to the extent they may not be lawfully
excluded. When the Goods are sold to a consumer within the meaning of the Unfair
Contract Terms Act 1977 their statutory rights are not affected by the provisions of
this clause.
4. In the event of the Customer making a claim in respect of any defect in terms of
clause 2 hereof the Customer must.
1. Reasonably satisfy the Company that the Goods have been properly installed,
commissioned, stored, serviced and used and without prejudice to the generality
of the foregoing that any defect is not the direct or indirect result of lack of repair
and/or servicing, incorrect repair and/or servicing, use of wrong materials and/or
incorrect spare parts
2. Allow the company to inspect the Goods and/or any installation and any
relevant packaging as and when reasonably required by the Company.
5. Subject to the Company being notified of any defect as is referred to in subclause 2 hereof within a reasonable time of it becoming apparent and subject
always to the terms of sub-clause 4 hereof, the Company shall, in its option, replace
or repair the defective Goods or refund a proportionate part of the Price. The
Company shall have no further liability to the Customer (save as mentioned in subclause 6 hereof).
6. The Company shall be liable to indemnify the Customer in respect of any claim
for death or personal injury to any person in so far as such is attributable to the
negligence or breach of duty of the Company or any failure by the Company to
comply with the provisions of sub-clause 2 hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in respect
of any claim by the Customer for costs, damages, loss or expenses (whether direct,
indirect, consequential or otherwise) or indemnity in any respect howsoever arising
including, but not by way of limitation, liability arising in negligence (other than
pursuant to clause 6 above) that may be suffered by the Customer or any third
party.
Page 5
Safety Notices
Peak Scientific Instruments cannot anticipate every possible circumstance which
may represent a potential hazard. The warnings detailed within this manual refer to
the most likely potential hazards, but by definition cannot be all inclusive. If the user
employs an operating procedure, item of equipment or a method of working which
is not specifically recommended by Peak Scientific, the user must ensure that the
equipment will not be damaged or become hazardous to persons or property.
Symbols
This manual uses the following symbols to highlight specific areas important to the
safe and proper use of the generator.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
personal injury or in the worst case death. Do not proceed beyond a WARNING
notice until the indicated conditions are fully understood or met.
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure,
process or similar, which if not correctly performed or adhered to, could cause
damage to the generator or the application. Do not proceed beyond a CAUTION
notice until the indicated conditions are fully understood or met.
Caution, risk of electric shock. Ensure power to the generator has been removed
before proceeding.
Safety Notice to Users
WARNING
WARNING
WARNING
Page 6
These instructions must be read thoroughly and understood before
installation and operation of your Peak Generator. Use of the
generator in a manner not specified by Peak Scientific MAY impair the
SAFETY provided by the equipment.
When handling, operating or carrying out any maintenance, personnel
must employ safe engineering practices and observe all relevant local
health and safety requirements and regulations. The attention of UK
users is drawn to the Health and Safety at Work Act 1974, and the
Institute of Electrical Engineers regulations.
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment maybe
impaired.
EU Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen Gas Generator
Models: MS Table
To which this declaration relates, is in conformity with the following applicable EU
Directives, harmonized standards, and other normative requirements.
•
Low Voltage Directive 2014/35/EU
EN 61010-1: 2010 Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use.
•
Electromagnetic Compatibility Directive 2014/30/EU
EN 61326-1: 2013 Electrical Equipment for Measurement, Control and Laboratory
Use – EMC Requirements. (Class A)
•
Restriction on the use of certain hazardous substances in electronic equipment
(RoHS) Directive 2011/65/EU as amended by EU 2015/863.
•
FCC 47 CFR Part 15 class A
Unintentional radiators; Conducted and Radiated emissions limits.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
UK Declaration of Conformity
We
Peak Scientific Instruments Ltd.
Of
Fountain Crescent, Inchinnan, Renfrewshire, PA4 9RE
Hereby declare that, this declaration of conformity is issued under the sole
responsibility of the manufacturer.
Equipment:
Nitrogen Gas Generator
Models: MS Table
To which this declaration relates, is in conformity with the following applicable UK
Statutory Instruments, Standards and other normative requirements.
•
The Electrical Equipment (Safety) Regulations 2016 (SI 2016 / 1101) as amended.
BS61010-1:2010 Safety Requirements for Electrical Equipment for Measurement
Control and Laboratory Use.
•
The Electromagnetic Compatibility Regulations 2016 (SI 2016 / 1091) as amended.
BS61326-1:2013 Electrical Equipment for Measurement , Control and Laboratory Use
– EMC Requirements.
•
The Restriction of the Use of Certain Hazardous Substances in Electrical and
Electronic Equipment Regulations 2012 (SI 2012 / 3032) as amended.
Signed for and on behalf of Peak Scientific by
Signed:
Name:
Fraser Dunn
Position:
Design Engineering Manager
Peak Scientific Instruments ltd,
Inchinnan, Renfrew, Scotland, PA4 9RE, UK.
Date:
10th August 2021
WEEE Compliance Statement
The Waste Electrical and Electronic Equipment (WEEE) Regulations SI 2013 No 3113
and or the Waste Electrical and Electronic Equipment (WEEE) Directive 2012/19/EU
apply to all electrical and electronic equipment placed on the market in the UK and EU
covered by the scope of regulations which can be found in the Government Guidance
Notes (PDF) produced by the Department for Business Innovation and skills for the UK
and here for Europe.
All PEAK products that are subject to the WEEE directive are compliant with the
WEEE marking requirement. Such products are marked with the “crossed-out wheelie
bin” symbol (shown below) in accordance with European standard EN50419. All
old electrical equipment can be recycled. Please do not dispose of any electrical
equipment (including those marked with this symbol) in general rubbish bins. Please
contact your dealer or distributor for clarity.
Technical Specification
Environment
1B
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
1N
5°C (41°F)
30°C (86°F)
2000 meters
70% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Inlet Conditions
Maximum Inlet Air Pressure
Minimum Air Inlet Flow
N/A
N/A
8.3-10 bar (120-145 psi)
115 L/min (4.06 cfm)
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
4.50 bar (60 psi)
19 L/min (0.67 cfm)
7.60 bar (110 psi)
26 L/min (0.91 cfm)
4.1 bar (60 psi)
25 L/min (0.88 cfm)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP
3
30 minutes
Generator Outlets
Curtain Maximum Output Pressure
Curtain Maximum Outlet Flow Rate
Source Maximum Output Pressure
Source Maximum Outlet Flow Rate
Exhaust Maximum Output Pressure
Exhaust Maximum Outlet Flow Rate
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlet
Pressure Gauges
Start-Up Time
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord Type
Pollution Degree
Installation Category
110-240 VAC
50/60 Hz
1 Amp
C14
C13
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 10
100.0 (39.4) x 83.0 (32.6) x 80.4 (31.6)
100.0 (220.5)
103.5 (229.0)
132.5 (292.0)
136.0 (300.0)
Technical Specification
Environment
2N Hi-Flow
Minimum Operating Ambient Temperature
Maximum Operating Ambient Temperature
Maximum Altitude
Maximum Relative Humidity
Minimum Storage Temperature*
Maximum Storage Temperature*
5°C (41°F)
30°C (86°F)
2000 meters
70% Non-Condensing
-20°C (-4°F)
60°C (140°F)
Inlet Conditions
Maximum Inlet Air Pressure
Minimum Air Inlet Flow
8.3-10 bar (120-145 psi)
115 L/min (4.06 cfm)
Generator Outlets
Curtain Maximum Output Pressure
Curtain Maximum Outlet Flow Rate
Source Maximum Output Pressure
Source Maximum Outlet Flow Rate
Exhaust Maximum Output Pressure
Exhaust Maximum Outlet Flow Rate
Particles
Phthalates
Suspended Liquids
Gas Outlets
Drain Outlet
Pressure Gauges
Start-Up Time
4.50 bar (65 psi)
10 L/min (0.35 cfm)
7.60 bar (110 psi)
22 L/min (0.78 cfm)
4.1 bar (60 psi)
10 L/min (0.35 cfm)
<0.01µm
NONE
NONE
3 x ¼” BSPP
1 x ¼” BSPP
3
30 minutes
Electrical Requirements
Voltage
Frequency
Current
Input Connection
Power Cord Type
Pollution Degree
Installation Category
110-240 VAC
50/60 Hz
1 Amp
C14
C13
2
II
General
Dimensions cm (inches) WxDxH
Generator Weight Kg (lbs)
Shipping Weight Kg (lbs)
Page 11
100.0 (39.4) x 83.0 (32.6) x 80.4 (31.6)
108.5 (239.2)
136.0 (300.0)
Unpacking
Although Peak Scientific takes every precaution with safe transit and packaging, it is
advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ and ‘TIP-N-TELL’ labels for signs of rough handling prior to
unpacking.
Any damage should be reported immediately to the carrier and Peak Scientific or the
Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require
two people to remove the unit from the shipping crate and to manoeuvre the air
compressor onto the bench.
Please save the product packaging for storage or future shipment of the generator.
Note: Included with the generator is a “Fittings Kit” containing mains power leads
for UK, EU & US and also all the required fittings and warranty registration card. Be
careful not to discard these with the packaging.
Page 12
Fittings Kit Contents
Supplied in the Fittings Kit are all the fittings required to connect the generator to the
application. The contents of the Fittings Kit are as follows:
1. ¼” Teflon Tubing* x 12m
2. 6mm PE Tubing* x 3m
3. 8mm Hex Key x 1
4. UK Mains Power Cable x 1
5. EU Mains Power Cable x 1
6. US Mains Power Cable 110v
7. US Mains Power Cable 230v
x1
x1
8. M5 Plain Washer x 6
9. M5 Spring Washer x 6
10. M5 Screw x 6
11. 6mm Push-In Plug x 1
* included with 1N & 2N Table Variants only
All of the generators output ports are located on the output panel at the rear of the
unit.
Page 13
Installation
Generator Environment
The generator is designed for indoor use only. It should be installed adjacent to
the application(s) it is supplying. If this is not convenient then the unit can be sited
elsewhere, however, consideration should be made of the lengths of pipe runs as
pressure drops can result from extended runs of pipe.
Performance of the generator (like all sophisticated equipment) is affected by
ambient conditions. Note should also be taken to the proximity of Air Conditioning
outlets. These can sometimes give rise to “pockets” of air with high relative humidity.
Operation of the unit within such a pocket could adversely affect its performance.
Consideration should also be given to the air flow around the unit. It is recommended
that an air gap of 75mm (3”) should be maintained down both sides and at the rear of
the unit. Please refer to the drawing below for the general dimensions of the unit.
Minimum Operating Ambient Temperature:
5 °C (41 °F)
Maximum Operating Ambient Temperature:
30 °C (86 °F)
Page 14
Generator Overview
General Dimensions
920.00 mm
800.00 mm
780.00 mm
The generator must always be placed on a flat, level surface. Failure to
do so will affect the performance of the generator.
WARNING
Page 15
Rear Connections
Supplied in the Fittings Kit are all the fittings required to connect the MS Table
Generator to the application. The contents of the Fittings Kit are as follows:
Exhaust Output
Source Output
Curtain Output
Air Input
Drain
Power Inlet
Mains Power Cable
Drain Connection
For users of the 1N and 2N Hi-Flow, use the ¼” Teflon tubing to connect the drain
outlet to a suitable drain connection. It should be noted that the Generator can expel a
considerable amount of water from this (dependant on ambient humidity).
CAUTION
Page 16
If a container is used it should be emptied at regular intervals.
The container must NOT have an air tight seal as water and air are
expelled at pressure.
Electrical Connection
Connect the Generator to an appropriate 110 or 230 volt single-phase supply, refer to
the generator serial plate for input specification and ensure your supply matches the
requirements.
If the appropriate power cord is not supplied; a new plug, rated to at least 13 amps,
can be fitted by a qualified electrician.
WARNING
This unit is classified as SAFETY CLASS 1. THIS UNIT MUST BE
EARTHED. Before connecting the unit to the mains supply, please
check the information on the serial plate. The mains supply must be of
the stated AC voltage and frequency.
EARTH/GROUND (E):-
Green & Yellow
or
Green
LIVE (L):-
Brown
or
Black
Neutral (N):-
Blue
or
White
Electrical requirements are 110 or 230VAC nominal +/- 10% depending on chosen
model. However, running continuously at voltages outwith this is not recommended.
Extended periods at extremes can have a detrimental effect on the operation and life
of the Generator.
WARNING
Page 17
If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment maybe
impaired.
Air Connection
The Table–1N/2N Hi Flow Generator should be connected to a clean, dry, OIL – FREE
source of compressed air. A minimum inlet pressure of 120 psig (8.3 barg) is required.
Any doubts as to the suitability of your compressed air supply should be referred to
Peak Scientific or any of their authorised partners.
The Compressed Air supply should be connected to the inlet located at the rear of the
unit as shown in the Rear Connections section of this manual.
Air Purity
The MS Table generator should be connected to an air supply that, as a minimum,
meets ISO8753-1:2010 Class 1.4.1
Class 1 Particulate
In each cubic metre of compressed air, the particulate count should not exceed
20,000 particles in the 0.1 - 0.5 micron size range, 400 particles in the 0.5 - 1 micron
range and 10 particles in the 1 - 5 micron size range.
Class 4 Water
A pressure dew point (PDP) of +3°C or better is required and no liquid water is allowed.
Class 1 Oil
In each cubic metre of compressed air, not more than 0.01mg of oil is allowed. This is a
total level for liquid oil, oil aerosol and oil vapour.
Page 18
Connecting to the application
Once the initial purge run of 30 minutes has completed the generator is now ready to
be connected to the application(s).
The pressure in the internal storage tanks must be allowed to
dissipate before connecting the generator to the application(s)
CAUTION
Attach the ¼” compression fitting to the outlet of the generator. Using the ¼” tubing
supplied, connect the outlet of the generator to the inlet on the application.
If you require more tubing than is supplied please refer to the Tubing Lengths section.
CAUTION
Once the tubing is connected to the application, please ensure that it
is thoroughly checked for being leak-tight. Even the slightest leak in
the gas supply between the generator and the application can lead to
a reduction in efficiency.
Tubing Lengths
CAUTION
The diameter of the tubing which will be connected to the gas outlet
is important and is determined by the length of tubing required.
Failure to follow these recommendations could lead to pressure
between generator and application.
< 10 meters:
Use ¼”/3⁄16” (¼” O/D, 3⁄16” I/D) P.T.F.E. tubing.
> 10 - 40 meters: Use 3⁄8”/ 5⁄16” (3⁄8” O/D, 5⁄16” I/D). Tubing and fittings not supplied
in the fittings kit.
> 40 metres:
Please contact Peak Scientific with the relevant distance and
we will calculate the flow resistance and the tubing size required.
A combination of ¼”/3⁄16” and 3⁄8”/5⁄16” tubing may be used to ensure that there is no
large diameter tubing within the lab (i.e. for the first 20 meters from the generator
use 3⁄8”/5⁄16” and the final 10 meters to the application use ¼”/3⁄16” tubing). Keep the
connections and bends to a minimum.
Page 19
Normal Operation
The MS Table Gas Generator is designed specifically to minimize operator involvement. Given that the system is installed as described in earlier sections and is serviced
in accordance with the specified maintenance recommendations (see Service Requirements), then it should simply be a matter of turning the Generator on when it is required.
The Generator will automatically produce the factory set flow and pressure as detailed
in the Technical Specifications.
Unusual Operation
If at any time the generator begins to emit excessive noise or vibration, then it should
be switched off and you should contact Peak Scientific or the Peak Partner from which
the generator has been purchased.
Relocation of MS Table
1. Turn off and disconnect power to the generator.
2. Disconnect the Air Connection to the generator.
3. Remove the MS Instrument from on top of the table
4. Disengage the leveling feet.
Once the MS Table has been safely moved to its new location, the MS Table can then
be installed as per the Installation Guide.
Fan Power LED
The MS Table 2N Hi-Flow is fitted with a fan power LED on the front panel.
The LED should be illuminated when the Generator is supplied with power. This is
indication that the fans attached to the front panel are being supplied with power, and
are running correctly.
The front panel should be connected to the power supply as shown below.
Page 20
Safe Operation of Pump Tray
For safe installation or removal of roughing pumps, we recommend use of the sliding
function of the pump tray, as shown below.
This should only be attempted once the Mass Spec has been
positioned on top of the table.
WARNING
Remove the front panel, ensuring that the fan power and earth cables are
disconnected.
Remove earth straps by pulling it off the spade connector on the front panel.
Depress both yellow tabs simultaneously and pull tray outwards until it clicks into
position.
Once pumps have been installed, ensure correct reconnection of earth straps and fan
power lead.
Page 21
Infinity 1031 Installation
The MS Table 1N & 2N Hi-Flow specifically contains an integrated Nitrogen Generator
providing a source of High Purity Nitrogen with 2 independent additional outlets of
clean dry Zero Grade air.
The MS Table 1B does not have an integrated Nitrogen Generator, but does have the
facility to bolt-on to the side of the table a Peak Scientific Instruments Wall Mounted
Infinity 1031 Nitrogen Generator.
Using the remaining M5 fittings from the fittings kit, mount the Infinity 1031 to the side
of the MS Table 1B. Note the front panel of the Infinity 1031 will need to be removed in
order to attach the fittings.
Page 22
Service Requirements
Service Schedule
Purchase Interval
Component
Visit
Genius MS Table 1N Annual Maintenance
Kit
12 Months
Genius MS Table 2N Hi-Flow Annual
Maintenance Kit
Infinity 1031 Annual Maintenance Kit
Page 23
www.peakscientific.com/ordering
Peak Protected
With Peak Scientific you invest in not only a product but peace of mind. With a network of
certified Peak engineers stationed throughout the globe, Peak’s rapid response team are
never far away and our commitment is to keep your generator running day in, day out,
protecting your laboratory workflow.
[Peak Protected] can provide…
Installation
A dedicated Peak engineer will visit
your lab to install and setup your
generator
IQ/OQ
Certified assurance for applications
requiring documented qualification
Complete plan
Swift response by a Peak Service
Engineer within 72 hours & planned
preventative maintenance
Spares
Genuine Peak parts with express
delivery, ensuring optimal
performance and lifetime
Premium Protected
Guaranteed rapid on-site response
within 24 hours & planned
preventative maintenance
Technical Support Hotline
Around the clock support by phone
or online with our global technical
helpdesk
To find out more about protecting your investment visit: www.peakscientific.com/protected
Page 24
Cleaning
Clean the outside of the generator only using warm soapy water and a clean damp
cloth. Ensure all excess fluid is thoroughly removed from the cloth prior to use.
Cleaning should only be undertaken with the power switched off and
the power cord removed from the rear of the Generator.
CAUTION
Under no circumstances should any solvents or abrasive cleaning
solutions be used as these can contain fumes that could be harmful to
the generator.
Care should be taken with Leak Detections Liquids.
CAUTION
Page 25
Troubleshooting
Problem
Possible Solution
The green LED has not illuminated but the
power switch is illuminated.
•
Contact your service provider.
•
Check for leaks between the generator
and instrument.
•
Contact your service provider.
•
Contact your service provider.
The instrument is reporting low pressure.
The unit is switched on but the fans are not
turning and the LED is not lit.
Page 26
Go Online or Complete and Return
We know that registering any of your recently purchased products is not
the first thing on your mind- but it is very important to both of us. Not all
warranties are alike and Peak Scientific stand out against other gas suppliers as
we offer a comprehensive, quick response, on-site warranty. This means that
in the very unlikely case that your gas generator develops a fault we have rapid
support teams on-hand around the world who are able to come to your lab and
get you back up and running in no time.
Register for your comprehensive 12 month on-site warranty with ease online at
www.peakscientific.com/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Go Online or Complete and Return
You can register for your FREE 12 month Warranty with ease online at
www.peakscientific/protected.
Alternatively, you can send the completed form to Peak Scientific by post or
email at warranty@peakscientific.com.
Product Warranty Registration
Contact name
Email address
Company
Address
City/town
Postcode
Country
Telephone
Generator serial #
Model type
Installation date
Do you still use an
alternative gas solution i.e.
cylinders or bulk liquid?
Yes
What gas requirements do
you have in your lab?
Hydrogen
No
Nitrogen
Zero Air
Extend your cover with
Peak Scientific offer comprehensive gas generator after sales support
packages. Peak [Protected] aftercare support can guarantee an on-site
response within 72 hours*, genuine parts from our ISO9001 approved factory
and a 95% first-time fix rate. See our enclosed Peak [Protected] leaflet for
further information.
Important!
You have 1 month to register your Peak Scientific product from the date of installation. Once
registered the warranty will be honoured for a period of 12 months. If you wish to defer the installation
of your generator, you must notify Peak Scientific immediately by emailing warranty@peakscientific.
com. For generators that remain unregistered after 1 month from the shipment date, the warranty will
be considered active from the date of factory dispatch.
* Complete Plan only
Important!
You have 1 month to register your Peak Scientific product from the date of
installation. Once registered the warranty will be honoured for a period of 12
months. If you wish to defer the installation of your generator, you must notify
Peak Scientific immediately by emailing warranty@peakscientific.com. For
generators that remain unregistered after 1 month from the shipment date, the
warranty will be considered active from the date of factory dispatch.
Page 27
Peak Scientific has highly trained, fully certified Field Service Engineers located in over 20 countries
across every continent around the world. This allows us to provide an industry-leading rapid
response service to our customers. With [Peak Protected], your laboratory’s productivity becomes
our top priority.
To discuss Peak Protected generator cover and payment options speak to your local Peak
Representative or for further information contact: protected@peakscientific.com
Peak Scientific
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Tel: +44 141 812 8100
Fax: +44 141 812 8200
For further information on any of our
generator products please contact
marketing@peakscientific.com
Copyright © Peak Scientific Instruments Ltd 2021 — MS Table User Manual | Rev 3 | PN UM-0012| RSID 1927 | EN —06/10/2021