Extracted text
2024
Rev 10 - 11/04/2024
3
3
4
5
5
5
6
7
7
7
7
7
8
8
9
9
11
12
12
12
13
13
13
14
Warranty Entitlement
14
Supported Maintenance Guide
15
Service Parts List
17
10
SMG Guide Added
Cleo Denholm
24/06/2024
Introduction
The following is a guide for maintenance of the Infinity 103x Nitrogen Generator as provided for by
the Peak Service Plan.
This procedure is to be performed on an annual basis.
1. Check that the LC/MS is not in use and that it is OK to disconnect the Gas Supply.
Check the Gas Demand is turned OFF.
2. Shut the Air Supply Valve.
3. Remove the Front Access Panel.
4. Carefully slacken the Gas Outlet Lines from the right side of the unit and allow it to
depressurize. Confirm that the gauges return to Zero. Any pressure gauge that
does not return to zero may be damaged and must be replaced.
5. Replace the 2-off RAC Filter Elements.
Page 15
6. Unscrew the top and bottom covers from the breathing air filter.
7. Replace the Breathing Air Filter Elements and re-assemble breathing air filter.
8. Carefully open the Air Supply Valve.
9. Re-tighten the Nitrogen Outlet Lines and re-fit the Front Panel.
10. Switch the LC/MS gas demand ON.
Page 16
Service Parts List
Part Number
Description
Quantity
3302277 (00-0039*)
RAC Filter Element
2
3302280 (00-0045*)
Coalescing Element for BA Filter
1
3302281 (00-0046*)
Active Carbon Element for BA Filter
1
Element part numbers with (*) are obsolete, the 330XXXX part number should be ordered.
Page 17
Page 18
Page 19
2024